VITROS 3600 Immunodiagnostic System Product Code 6802783 and Refurbished Product Code 6802914 Running Software Version 3.8.0
Class I - DangerousWhat Should You Do?
- Check if you have this product: Product Code: 6802783, UDI-DI: 10758750002979; Product Code: 6802914, UDI-DI: 10758750007103; Serial Numbers: J36000109 J36000294 J36001012 J36001489 J36000292 J36000340 J36001024 J36001495 J36000484 J36000348 J36001027 J36001513 J36000489 J36000446 J36001051 J36001515 J36000955 J36000511 J36001088 J36001519 J36001151 J36000517 J36001108 J36001528 J36001167 J36000580 J36001111 J36001535 J36001222 J36000688 J36001112 J36001570 J36001283 J36000753 J36001117 J36001573 J36001330 J36000861 J36001237 J36001575 J36001348 J36000877 J36001245 J36001579 J36000920 J36001246 J36001581 J36000923 J36001401 J36001582 J36000929 J36001403 J36001583 J36000933 J36001427 J36001600 J36000941 J36001451 J36001602 J36000942 J36001488 J36001618 J36001000
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Ortho-Clinical Diagnostics, Inc.
- Reason for Recall:
- During planned monitoring of data post-release of Software Version 3.8.0 an increased occurrence of software issues relating to unresponsive subsystems, unresponsive user interface (sometimes referred to as a screen freeze), and/or printer disconnections was identified. The consequence of these issues would be a delay in reporting results due to the required system shutdowns to recover the system. This delay in reporting results could occur during any assay, including which the speed of the result is deemed critical or STAT and could result in a delay of treatment in a patient. Due to the release strategy of this software version, the user could not update the system but a field engineer had to manually install each affected system.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
VITROS 3600 Immunodiagnostic System Product Code 6802783 and Refurbished Product Code 6802914 Running Software Version 3.8.0
Product Codes/Lot Numbers:
Product Code: 6802783, UDI-DI: 10758750002979; Product Code: 6802914, UDI-DI: 10758750007103; Serial Numbers: J36000109 J36000294 J36001012 J36001489 J36000292 J36000340 J36001024 J36001495 J36000484 J36000348 J36001027 J36001513 J36000489 J36000446 J36001051 J36001515 J36000955 J36000511 J36001088 J36001519 J36001151 J36000517 J36001108 J36001528 J36001167 J36000580 J36001111 J36001535 J36001222 J36000688 J36001112 J36001570 J36001283 J36000753 J36001117 J36001573 J36001330 J36000861 J36001237 J36001575 J36001348 J36000877 J36001245 J36001579 J36000920 J36001246 J36001581 J36000923 J36001401 J36001582 J36000929 J36001403 J36001583 J36000933 J36001427 J36001600 J36000941 J36001451 J36001602 J36000942 J36001488 J36001618 J36001000
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0543-2024
Related Recalls
VITROS Chemistry Products LAC Slides. Catalog Number: 8150112. 1 Sales Unit = contains 5 cartridges of 18 slides, for a total of 90 tests.
Ortho-Clinical Diagnostics
VITROS Chemistry Products LAC Slides on VITROS XT 3400 and VITROS XT 7600 Systems are experiencing an increased occurrence of calibration failure related to lots from coating 0130 and above with Condition Code TH4-63J during calibration, causing the delay of lac test results and potentially impact patient management. The issue was identified by QuidelOrtho through an increase in customer complaints compared to the historical baseline with average failure rate of 1% for coating 0130 in 2025 while average of failure rate for other coating was 0.06% in 2025 and average of failure rate for all coatings is 0.03% for 2024.
Chemistry Products LAC Slides. Catalog Number: 8433880. 1 Sales Unit = contains 5 cartridges of 60 slides, for a total of 300 tests.
Ortho-Clinical Diagnostics
VITROS Chemistry Products LAC Slides on VITROS XT 3400 and VITROS XT 7600 Systems are experiencing an increased occurrence of calibration failure related to lots from coating 0130 and above with Condition Code TH4-63J during calibration, causing the delay of lac test results and potentially impact patient management. The issue was identified by QuidelOrtho through an increase in customer complaints compared to the historical baseline with average failure rate of 1% for coating 0130 in 2025 while average of failure rate for other coating was 0.06% in 2025 and average of failure rate for all coatings is 0.03% for 2024.
VITROS Chemistry Products Ca Slides; Catalog Number: 145 0261;
Ortho-Clinical Diagnostics
Generations 67 and above of VITROS Ca Slides are experiencing an increased occurrence of Condition Code TH4-63J during calibration on VITROS XT 3400 and VITROS XT 7600 Systems, causing calibration failures that delay calcium test results and potentially impact patient management. The imaging reflectometer algorithm detects excessive "spikes" (imperfections) in slide images and generates the TH4-63J condition code when the threshold of 600 spikes is exceeded. Updated analysis revealed a 19-fold increase in failure rates from 0.0043% (2024 baseline) to 0.082% for Generation 67+ slides in 2025, with 100% of calibrator failures occurring at the lowest concentration level. This can result in delayed diagnosis and treatment of conditions such as hypercalcemic crisis, parathyroid disorders, and other calcium-related emergencies requiring urgent intervention.