Copeland HA Resurfacing Head, MB/HA EAS Head; Item Nos. 11-114641 11-114642 11-114643 11-114644 11-114644S 11-114645 11-114646 11-114647 11-114648 11-114661 11-114662 11-114663 11-114664 11-114664S 11-114665 11-114666 11-114667 11-114668 11-114632 11-114633 Product Usage: Partial Shoulder Replacement.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    All lots manufactured between January 2008 - May 2019: 197000 838570 047410 869160 047400 047390 203030 203040 224470 482850 052950 076580 917420 326610 362350 565380 374150 420170 374160 882210 882090 746150 691040 969190 942350 726180 048880 385080 111310 013150 868010 638680 745730 780270 780240 013140 834290 623340 817490 028380 186120 409420 847030 679990 755600 409330 706270 169490 936000 400990 343410 235880 917410 872790 715870 179470 313330 324140 362650 305750 326590 920210 564070 305740 632120 684640 536380 564770 014640R 745910 629320 629350 014640 807570 946770 111280 946800 735450 108390 370610 385070 884110 013160 884080 621660 919220 836010 645380 706600 899340 712500 838730 963630 021010 896300 673910 669190 146510 613110 678830 335900 684340 106100 105780 140000 447810 154700 911280 363670 102500 931980 519820 114830 762310 076570 296510 917390 096130 059170 362810 552990 684260 587040 565220 484880 634740 485340 414520 906620 808780 678310 969290 013240 013250 014670 014850 384890 108310 808790 561410 808770 336710 018410 617860 313520 378380 579070 566190 927200 297800 673900 348320 702090 939400 736230 268220 055290 347210 347230 342820 725450 817310 153990 725430 698880 698890 823840 188130 037810 705530 602470 301330 463540 940980 028450 304010 491810 596490 684830 771630 731090 384570 343400 049790 057690 766800 513670 114820 915280 408980 715850 917370 917380 296500 076560 326600 510940 568130 363140 420030R 420030 808760 099520 917350 414510 629400 808720 014610 813330 044040 484870 621620 694280 694250 739520 991720 563690 671840 983580 250250 190920 231850 390870 757150 799950 517900 204080 764010 114810 249030 915250 061900 917340 076550 917360 131600 320520 348900 326030 565370 788320 610590 200890 462920 540650R 584540 540650 540610 691160 634350 485330 808710 085020 108330 906800 042230 367330 013270 100260 584530 759100 621970 808700 781920 834590 645360 654830 307730 580280 789060 295500 110720 296810 218340 971510 290820 984540 715090 130360 280430 896280 701290 322680 582620 912810 911710 178270 591690 191690 367880 867300 112070 266870 798940 568280 613100 028440 620410 620420 784160 046210 712950 443870 642180 503770 420330 134180 294890 628730 763910 251590 096000 381290 086080 208040 304890 314410 498860 498870 546180 831840 419740 828320 767110 917310 991930 771690 296490 260290 235790 917330 076520 076540 427590 326690 326700 484860 674000 540570 584510 584520 808670 553580 484850 746060 683370 808660 946820 049430 125700 385030 677300 917320 645350 371650 371910 919370 808680 089930 089920 485520 606500 208850 502490 298670 656410 045750 085450 313510 745020 721210 144450 378370 712490 431600 433430 297790 760970 842280 174340 250930 297880 934950 206630 912780 434850 201690 915720 182160 627310 411050 450540 644770 191740 763920 944360 613090 168190 716270 824710 565110 759500 279840 751390 313950 297070 102460 399850 728220 741280 062670 599590 070530 958220 079180 917290 917300 052930 132080 132090 261190 193570 227680 378250 326580 363150 305470 553570 787630 484840 634340 606650 227710 674890 654520 866510 399620 485320 808640 055710 942310 795900 969300 808650 055720 744940 561400 056140 809740 477760 354200 896230 322660 483840 930010 135000 329910 991820 449680 192060 479170 738870 626460 715840 917270 320620 881600 540140 808620 553610 808630 485310 606660 553611 808510 055700 758650 143790 055690 552750 422710 751380 973120 763930 233200 503760 305760 211440 715830 791670 035100 457730 209330 965510 324130 052920 911290 339120 586290 540090 364180 586300 276570 162030 162040 553940 785680 852640 054840 382030 374030 805470 134580 790650 302140 547610R 297990 547610 896180 356960 771130 372160 747300 787100 688710 343770 823820 352910 611740 475630 768550 768550R 465490 919980 485950 139990 284880 342280 266180 728880 875320 541510 652040 747580 331370 327140 169040 866470 715820 261020 245080 275200 308520 275190 296480 250150 245050 374480 563990 563970 558620 558630 586280 561200 561210 558610 786920 865980 865990 385950 196500 044100 973720 373980 301080 301090 373990 374000 772460 697320 021410 748760 987390 165460 986940 383960 245320 429590 358790 095990 018850 293160 293170 886190 855020 941050 965160 855030 815540 992450 171270 172200 934030 934020 299320 337530 971820 179730 299340 482040 499120 701660 374100 646370 373540 374110 684180 705520 646320 944070 944380 906950 802610 802590 057210 969620 144420 296660 144410 144430 429290 712460 183950 680430 760910 715080 917470 760980 242270 968580 132950 297960 297810 297850 399380 297820 673960 673920 896130 936160 939390 364060 770560 415910 322670 471660 740640 687610 947490 025190 574890 851130 627290 787090 787080 531340 135140 309650 191670 191680 547500 367870 450150 725280 688670 992130 787730 093270 797850 602450 602460 134990 686500 768520 824690 809340 991800 336150 946080 335880 201960 684430 505740 490180 759490 117200 024480 252230 406320 132610 143430 409410 925400 682980 435670 561500 342270 271940 827340 139770 161980 453360 542830 522760 875310 039180 473210 781530 610870 294890 947310 070400 069880 026340 372930 430830 819690 399840 496920 886220 980410 416120 363660 164590 821940 115710 503750 212310 462060 120690 881240 881180 079230 169030 079220 531660 327130 090650 211420 845510 866440 090660 045460 819530 700180 972280 500720 209350 081430 304580 304620 304590 304610 232410 304600 304630 304640 209360 304650 308090 304670 209340 304680 363390 304660 439320 308270 552440 439330 586260 395930 564370 395960 586270 865920 865910 081420 865900 014180 276590 188720 188710 188700 289960 308150 305600 305610 314160 742650 874490 822830 990810 484530 933900 043010 336660 366540 624460 292740 426590 021300 608740 757630 125560 125550 125570 256220 256230 256250 256260 256270 256240 301800 301810 256280 346750 398940 533890 533900 322450 463080 322460 606490 671690 398950 671680 016951 859320 242750 031320 031310 136950 601010 295710 580240 580260 580270 483470 580250 656390 603690 603710 603700 912390 912440 912410 912360 912460 108920 044790 045110 912350 912370 912400 912430 971030 170520 117340 117380 254630 316930 330330 117230 117240 179720 117250 218290 491020 490960 381410 381160 898760 144330 144320 302070 997400 997020 712510 406860 874810 547650 117620 712480 227950 258090 442140 441380 535900 651460 578700 487770 650360 715530 680500 995880 650300 968570 968560 715070 229180 760730 964980 399420 297940 297950 399410 570680 399430 297930 673950 954480 896270 939380 033680 983850 985750 109670 251080 251090 322640 519470 687600 693640 749050 915700 955480 142250 021050 419750 273850 531290 604870 627280 787070 627270 706110 240900 084560 367830 367860 367820 367840 669170 725250 367850 669140 688700 669130 669160 698670 688690 669180 458280 669120 883600 787720 125880 622120 602440 602430 134980 910760 205920 416030 329900 611720 686490 898590 037300 037310 063740 946070 965570 824680 991790 069560 809330 946060 113230 201950 335870 671830 006130 085250 160010 359270 422910 423070 670240 970390 057550 906250 231030 455030 455360 266250 406310 162310 649440 649430 828560 766990 870250 945370 143410 143420 410710 421170 880810 506030 682970 919970 785240 551280 119550 751370 561490 485930 003950 409830 243880 649340 271930 139980 522000 453340 542810 542820 668240 453350 126380 589770 313940 313930 555550 712940 875300 039170 214050 154780 673520 685010 935690 294870 935680 411780 452970 580230 940620 299340 584000 819680 053300 819670 289370 496800 496770 496780 399830 095080 164510 095090 744250 652030 685720 095070 115700 115670 171340 115690 503740 115680 893580 747570 212300 436640 841670 705920 866330 866320 169020 247060 547110 595380 595410 233730 531650 169000 595390 845490 090630 845500 090620 881170 866410 683040 090640 626430 211410 866400 626390 560760 035110 709380 052900 457800 132070 540070 539990 361570 586240 586250 605810 052880 886710 967150 886550 013320 013370 278090 668560 834550 656110 790070 656100 174700 933990 992440 701930 685180 374040 749140 237030 144310 277060 277070 555150 302060 594510 962630 442090 058150 297830 896150 673940 297870 487570 566520 566540 191660 658790 688660 893550 372670 824670 335860 024470 682960 421160 485920 729180 925510 485940 424440 353450 000130 522010 542800 927390 268300 555540 039150 552140 164570 685710 989050 533590 531640 866390 700170 915670 762330 971710 917650 052850 052860 132050 276630 132060 361560 416000 439280 416010 439310 439300 439260 361540 439270 439290 609990 876830 832940 876840 876850 013400 013380 144040 771280 935470 935480 339050 373950 301070 373970 483060 435000 813570 715820 715810 199740 199750 199760 898100 919050 391980 383940 545820 383950 018840 870250 095970 846240 431610 148760 431620 693580 341560 886180 698720 932880 012030 616130 616150 729610 398630 162020 162000 189050 279040 162010 289360 392050 600250 392060 580320 631730 710790 891500 553930 852630 934010 010230 972020 155640 226070 117570 179670 179680 505250 570330 373510 373500 374050 373520 373490 639310 566150 786760 786190 786780 841980 898030 898050 898040 898020 134530 282140 898750 452970 431680 277050 302050 549460 302040 408370 406850 710750 927560 968540 227940 258110 113160 673420 441370 552900 680450 650350 715060 014620 916220 650310 940490 048900 968550 283510 089290 715160 148040 148050 229070 121320 534520 986580 986570 033670 251070 322610 394060 322540 332100 394080 325680 553970 554000 687580 867010 603250 711220 762890 820860 819490 553980 693630 749040 687590 746390 714720 831850 021040 178260 296580 695730 574880 787060 627260 123210 191650 084550 453810 191620 191630 276180 453790 367810 453800 688680 823800 911030 892820 093190 498410 498370 498400 498420 622110 317640 910770 236680 329880 686480 611700 686470 766080 037290 037280 686240 824660 824640 965550 946040 321670 965560 991780 334820 476440 490200 113220 511220 696970 671820 847280 024460 013420 175540 429810 454110 649410 649420 649390 852280 933990 132600 945360 143400 503920 745440 374190 506020 435660 682950 421150 919950 919960 254820 843980 379540 396960 368060 550900 082420 975920 927580 927590 214760 271920 347570 284940 214750 522170 563850 594840 403940 342260 681280 542790 453330 542780 815540 982470 927350 268150 927320 927570 498820 039140 297060 151700 294860 452960 845700 940610 583990 894920 050620 247840 981160 496810 825210 399810 416110 164490 741270 095060 769680 164500 652010 060010 548410 523030 115660 961870 212280 479190 705910 120680 881220 168980 079190 866360 531630 533580 845450 090570 595340 845480 090580 090590 090610 045440 045450 626360 626380 687850 866380 907610 251580 095980 191250 406800 042590 179690 314420 374080 374090 478790 478800 646330 304880 208030 188710 715810 700140 500710 431800 097420 279660 304530 304520 304560 097410 719660 097390 431790 097400 304550 324120 510880 304570 439250 469810 586310 469830 539930 469820 586340 539830 539960 586230 439240 467930 586320 746210 415990 438920 583490 608790 583410 876820 876810 876800 583430 876790 967110 967130 012890 013470 917640 044110 279910 512330 044090 013010 669230 143910 199730 384890 390050 391950 742820 391960 877280 545810 985610 018830 148750 251570 039120 552940 341550 431590 431600 677340 159200 837890 295670 295680 295650 295660 161990 289370 289350 642680 852620 580300 790080 710780 972010 140520 170480 187670 179710 334720 179660 992430 570190 373460 237080 632890 373480 639230 373470 568300 646360 802580 898010 898740 802570 277030 277040 478520 348600 431580 452890 302030 408360 302020 547630 406830 258100 183940 680530 482740 650320 715150 015690 996510 967270 968540 148030 182470 148020 968530 148010 597820 597690 399390 399400 297920 897640 897650 033660 058130 079670 079730 266440 355010 433340 434840 519460 501080 372120 569900 693620 195970 985090 985100 108480 500790 232350 400180 531280 948140 695720 574870 986890 191600 191610 084540 367790 367800 241000 496350 688640 457400 698580 688650 725220 944350 944340 037790 266860 355420 413900 413890 797840 236650 236660 329870 686460 766070 611690 318020 063750 686100 686230 437910 967220 768530 824630 965540 991760 809320 201940 824620 505710 505720 759480 006110 037620 006120 095360 160000 214620 558800 742590 861710 906240 024440 455010 406290 455110 718590 007560 945350 421140 785230 339690 506000 079870 919940 109370 471380 427230 561470 427260 561450 196120 139970 927560 769300 196020 214740 342250 196070 608160 214730 904540 423120 091560 927550 542760 542770 590970 498570 313900 555530 589820 370320 105040 875290 039100 297050 685020 673230 454490 642190 935670 763920 102140 949660 349900 368180 666420 903230 231800 164470 163410 343380 744060 080070 164480 651990 961850 115640 115650 661790 114790 436620 436630 479210 989100 841660 705900 881200 866310 168970 531620 584940 233720 558520 533560 584920 845430 249150 845440 560730 211390 045410 626330 420300 715800 754520 431750 431770 965500 307790 965490 052870 132040 306940 754580 539680 539780 539720 536370 886130 865870 819090 865850 886140 783740 013410 044060 044080 865880 512410 645530 553990 554000 603680 789330 971500 313570 805630 897890 969610 163440 302010 549540 984530 715140 650330 940300 968520 399370 125560 297840 673930 556720 880330 880530 121110 110400 332090 332070 434820 486740 372110 915690 749030 153450 531270 191730 367780 507100 669100 669110 911020 797830 910780 063760 686160 768540 965530 113210 490190 906220 458120 766980 421130 682940 682940R 109270 243870 550890 207590 973990 271910 542750 927290 927310 496940 343370 562390 401280 115630 212270 705890 866350 233700 276660 090750 090740 560720 045390 085450 719650 715780 027290 469790R 469780 364140 469770 469790 469800 585650 608230 586220 633280 632390 013600 044070 013490 044050 130140 780230 847450 561630 685700 120670 744620 533550 090730 019770 563730 589160 722600 084620 626500 420370 777480 420350 420380 211500 211510 746160 045470 084610 308510 626490 159240 211490 362640 405790 405800 544750 544770 681260 684450 626480 420340 845550 103840 211470 342030 362680 572910 196650 084590 845540 188020 229970 247610 305730 405780 540670 553000 808740 808750 918980 521370 018560 103810 103820 131610 209320 211460 323210 437960 556500 766160 959910 085900 086150 873360 736580 560790 029780 050050 076530 103800 119850 285340 315810 405770 430800 455630 683360 683380 719730 726350 043990 044780 049110 833870 833880 880280 880290 880310 880320 052940 159200 211450 317740 327530 450430 540540 944420 596950 845530 866480 084580 084560 308500 159120 308900 365800 405730 405740 405750 405760 408990 906680 917280 590200 159090 293200 456970 469870 489230 564010 571400 587700 616190 683800 779990 866000 885370 078470 895090 005880 086490 093240 093250 103770 103780 103790 117030 191480 232180 293180 293190 296860 354660 365420 366260 385910 444820 456950 489210 489220 571370 571380 571390 616160 616170 616180 683710 683750 683760 712950 729630 729640 779980 825530 855010 865930 969900 969910 087870 092050 308490 103730 103740 103750 103760 157950 208840 295700 295720 296130 297580 298700 301310 301790 303540 312420 318000 318090 318100 323200 335010 339120 354150 374140 383100 392330 392340 420280 420290 420300 420310 420320 420330 422320 426710 431130 431140 433080 445070 450980 457340 519680 556480 556490 560640 587690 648490 683020 683030 689740 726080 766140 766150 815620 825520 887120 898820 938830 945810 103720 159080 260750 295730 301090 329000 417070 418320 444810 583580 570080 514830 308480 560750 012020 044340 066810 103690 103700 103710 148070 177660 260740 293150 306340 307530 317970 323870 328990 329890 366250 397850 427610 430820 432520 444800 451570 489190 489200 571330 571340 571350 571360 587660 587670 587680 588540 616120 682960 766130 779960 779970 825510 870120 885360 898090 941040 761210 420310 308470 044330 103670 103680 107340 156880 312970 323860 328980 353340 384950 385030 385070 385140 385170 385180 385190 385210 385230 385240 385260 385270 385290 385300 385310 385320 385330 427600 439290 444790 456940 469850 469860 479160 563940 587650 588530 736720 766120 781840 825500 962870 103650 103660 147990 298690 323190 417300 431760 433070 683900 766110 825490 005870 159030 159040 159060 683890 UDI: (01)00880304439962(17)180628(10)882090 (01)00880304439962(17)180728(10)942350 (01)00880304439962(17)180728(10)726180 (01)00880304439962(17)180828(10)111310 (01)00880304439962(17)180828(10)013150 (01)00880304439962(17)180828(10)868010 (01)00880304439962(17)190128(10)623340 (01)00880304439962(17)230128(10)028380 (01)00880304439962(17)230830(10)186120 (01)00880304439962(17)240204(10)409420 (01)00880304439962(17)240207(10)847030 (01)00880304439962(17)240221(10)679990 (01)00880304439962(17)240301(10)755600 (01)00880304439962(17)240505(10)409330 (01)00880304439962(17)240612(10)706270 (01)00880304439962(17)240708(10)169490 (01)00880304439962(17)251222(10)936000 (01)00880304439962(17)251229(10)400990 (01)00880304439962(17)260425(10)343410 (01)00880304439962(17)261121(10)235880 (01)00880304439917(17)180328(10)313330 (01)00880304439917(17)180328(10)324140 (01)00880304439917(17)180701(10)014640R (01)00880304439917(17)180828(10)884080 (01)00880304439917(17)201228(10)838730 (01)00880304439917(17)230324(10)335900 (01)00880304439917(17)230325(10)684340 (01)00880304439917(17)230418(10)106100 (01)00880304439917(17)230709(10)105780 (01)00880304439917(17)240825(10)140000 (01)00880304439917(17)241126(10)447810 (01)00880304439917(17)250126(10)154700 (01)00880304439917(17)250219(10)911280 (01)00880304439917(17)260207(10)363670 (01)00880304439917(17)260317(10)102500 (01)00880304439917(17)260426(10)931980 (01)00880304439917(17)260518(10)519820 (01)00880304439917(17)260624(10)114830 (01)00880304439955(17)180628(10)634740 (01)00880304439955(17)180701(10)485340 (01)00880304439955(17)180728(10)969290 (01)00880304439955(17)180828(10)108310 (01)00880304439955(17)181028(10)808770 (01)00880304439955(17)200328(10)313520 (01)00880304439955(17)220228(10)698890 (01)00880304439955(17)220528(10)037810 (01)00880304439955(17)230116(10)028450 (01)00880304439955(17)230204(10)304010 (01)00880304439955(17)230213(10)491810 (01)00880304439955(17)230221(10)596490 (01)00880304439955(17)230319(10)684830 (01)00880304439955(17)251221(10)771630 (01)00880304439955(17)251222(10)731090 (01)00880304439955(17)251223(10)384570 (01)00880304439955(17)260208(10)343400 (01)00880304439955(17)260210(10)049790 (01)00880304439955(17)260318(10)057690 (01)00880304439955(17)260322(10)766800 (01)00880304439955(17)260427(10)513670 (01)00880304439955(17)270803(10)114820 (01)00880304439894(17)180601(10)420030R (01)00880304439894(17)180628(10)808760 (01)00880304439894(17)180728(10)917350 (01)00880304439894(17)180728(10)629400 (01)00880304439894(17)180728(10)808720 (01)00880304439894(17)180728(10)014610 (01)00880304439894(17)230319(10)563690 (01)00880304439894(17)230327(10)671840 (01)00880304439894(17)230415(10)983580 (01)00880304439894(17)230520(10)250250 (01)00880304439894(17)250611(10)190920 (01)00880304439894(17)251016(10)231850 (01)00880304439894(17)251021(10)390870 (01)00880304439894(17)251222(10)757150 (01)00880304439894(17)251228(10)799950 (01)00880304439894(17)260509(10)517900 (01)00880304439894(17)260620(10)204080 (01)00880304439894(17)270305(10)764010 (01)00880304439894(17)270921(10)114810 (01)00880304439894(17)271016(10)249030 (01)00880304439924(17)180228(10)076550 (01)00880304439924(17)180601(10)540650R (01)00880304439924(17)180828(10)584530 (01)00880304439924(17)191228(10)295500 (01)00880304439924(17)200128(10)296810 (01)00880304439924(17)220228(10)367880 (01)00880304439924(17)230326(10)028440 (01)00880304439924(17)241029(10)620410 (01)00880304439924(17)241117(10)620420 (01)00880304439924(17)241201(10)784160 (01)00880304439924(17)241204(10)046210 (01)00880304439924(17)241208(10)712950 (01)00880304439924(17)250112(10)443870 (01)00880304439924(17)250126(10)642180 (01)00880304439924(17)260523(10)503770 (01)00880304439924(17)280320(10)420330 (01)00880304439900(17)180201(10)917310 (01)00880304439900(17)180728(10)808660 (01)00880304439900(17)201028(10)712490 (01)00880304439900(17)210328(10)842280 (01)00880304439900(17)210428(10)912780 (01)00880304439900(17)220228(10)191740 (01)00880304439900(17)230121(10)824710 (01)00880304439900(17)230220(10)565110 (01)00880304439900(17)230416(10)759500 (01)00880304439900(17)230607(10)279840 (01)00880304439900(17)240715(10)751390 (01)00880304439900(17)241023(10)313950 (01)00880304439900(17)250130(10)297070 (01)00880304439900(17)250313(10)102460 (01)00880304439900(17)260114(10)399850 (01)00880304439900(17)251223(10)728220 (01)00880304439900(17)251229(10)741280 (01)00880304439900(17)260126(10)062670 (01)00880304439900(17)260216(10)599590 (01)00880304439900(17)260315(10)070530 (01)00880304439931(17)180328(10)193570 (01)00880304439931(17)180328(10)227680 (01)00880304439931(17)180701(10)674890 (01)00880304439931(17)180728(10)399620 (01)00880304439931(17)180728(10)485320 (01)00880304439931(17)180728(10)808640 (01)00880304439931(17)180728(10)055710 (01)00880304439931(17)180828(10)055720 (01)00880304439931(17)230215(10)991820 (01)00880304439931(17)230218(10)449680 (01)00880304439931(17)230225(10)192060 (01)00880304439931(17)260726(10)479170 (01)00880304439931(17)260729(10)738870 (01)00880304439931(17)280302(10)626460 (01)00880304439948(17)180628(10)553610 (01)00880304439948(17)240121(10)422710 (01)00880304439948(17)240421(10)751380 (01)00880304439948(17)240825(10)973120 (01)00880304439948(17)250210(10)763930 (01)00880304439948(17)250326(10)233200 (01)00880304439948(17)260524(10)503760 (01)00880304439948(17)260628(10)305760 (01)00880304439948(17)280301(10)211440 (01)00880304204829(17)180228(10)965510 (01)00880304204829(17)180528(10)586290 (01)00880304204829(17)191228(10)785680 (01)00880304204829(17)191228(10)852640 (01)00880304204829(17)200428(10)805470 (01)00880304204829(17)200528(10)134580 (01)00880304204829(17)200628(10)790650 (01)00880304204829(17)200628(10)302140 (01)00880304204829(17)210201(10)547610R (01)00880304204829(17)210528(10)771130 (01)00880304204829(17)210528(10)372160 (01)00880304204829(17)220528(10)352910 (01)00880304204829(17)230218(10)475630 (01)00880304204829(17)230219(10)768550 (01)00880304204829(17)230219(10)768550R (01)00880304204829(17)240209(10)465490 (01)00880304204829(17)240319(10)919980 (01)00880304204829(17)240518(10)485950 (01)00880304204829(17)240706(10)139990 (01)00880304204829(17)240721(10)284880 (01)00880304204829(17)240810(10)342280 (01)00880304204829(17)240916(10)266180 (01)00880304204829(17)241030(10)728880 (01)00880304204829(17)241201(10)875320 (01)00880304204829(17)250105(10)541510 (01)00880304204829(17)260216(10)652040 (01)00880304204829(17)260526(10)747580 (01)00880304204829(17)260630(10)331370 (01)00880304204829(17)270305(10)327140 (01)00880304204829(17)270419(10)169040 (01)00880304204829(17)280123(10)866470 (01)00880304204812(17)181201(10)373980 (01)00880304204812(17)190928(10)965160 (01)00880304204812(17)200501(10)906950 (01)00880304204812(17)210528(10)322670 (01)00880304204812(17)210528(10)471660 (01)00880304204812(17)210628(10)687610 (01)00880304204812(17)211128(10)531340 (01)00880304204812(17)230123(10)336150 (01)00880304204812(17)230128(10)946080 (01)00880304204812(17)230204(10)335880 (01)00880304204812(17)230204(10)201960 (01)00880304204812(17)230225(10)684430 (01)00880304204812(17)230228(10)505740 (01)00880304204812(17)230319(10)490180 (01)00880304204812(17)230327(10)759490 (01)00880304204812(17)230822(10)117200 (01)00880304204812(17)230826(10)024480 (01)00880304204812(17)230916(10)252230 (01)00880304204812(17)231011(10)406320 (01)00880304204812(17)240120(10)132610 (01)00880304204812(17)240120(10)143430 (01)00880304204812(17)240201(10)409410 (01)00880304204812(17)240221(10)925400 (01)00880304204812(17)240226(10)682980 (01)00880304204812(17)240227(10)435670 (01)00880304204812(17)240507(10)561500 (01)00880304204812(17)240706(10)342270 (01)00880304204812(17)240717(10)271940 (01)00880304204812(17)240806(10)827340 (01)00880304204812(17)240825(10)139770 (01)00880304204812(17)240828(10)161980 (01)00880304204812(17)240904(10)453360 (01)00880304204812(17)240905(10)542830 (01)00880304204812(17)241013(10)522760 (01)00880304204812(17)241203(10)875310 (01)00880304204812(17)241209(10)039180 (01)00880304204812(17)250415(10)473210 (01)00880304204812(17)250426(10)781530 (01)00880304204812(17)250430(10)610870 (01)00880304204812(17)250430(10)294890 (01)00880304204812(17)250519(10)947310 (01)00880304204812(17)250526(10)070400 (01)00880304204812(17)250527(10)069880 (01)00880304204812(17)250603(10)026340 (01)00880304204812(17)250618(10)372930 (01)00880304204812(17)250717(10)430830 (01)00880304204812(17)251006(10)819690 (01)00880304204812(17)251029(10)399840 (01)00880304204812(17)251106(10)496920 (01)00880304204812(17)251222(10)886220 (01)00880304204812(17)260108(10)980410 (01)00880304204812(17)260120(10)416120 (01)00880304204812(17)260125(10)363660 (01)00880304204812(17)260223(10)164590 (01)00880304204812(17)260324(10)821940 (01)00880304204812(17)260421(10)115710 (01)00880304204812(17)260425(10)503750 (01)00880304204812(17)260620(10)212310 (01)00880304204812(17)261107(10)462060 (01)00880304204812(17)261115(10)120690 (01)00880304204812(17)261215(10)881240 (01)00880304204812(17)270114(10)881180 (01)00880304204812(17)270305(10)079230 (01)00880304204812(17)270315(10)169030 (01)00880304204812(17)270319(10)079220 (01)00880304204812(17)270422(10)531660 (01)00880304204812(17)270825(10)327130 (01)00880304204812(17)271010(10)090650 (01)00880304204812(17)280108(10)211420 (01)00880304204812(17)280112(10)845510 (01)00880304204812(17)280117(10)866440 (01)00880304204812(17)280220(10)090660 (01)00880304204812(17)280305(10)045460 (01)00880304204812(17)280327(10)819530 (01)00880304204805(17)201228(10)680500 (01)00880304204805(17)220101(10)367830 (01)00880304204805(17)220828(10)329900 (01)00880304204805(17)221228(10)824680 (01)00880304204805(17)230101(10)991790 (01)00880304204805(17)230117(10)113230 (01)00880304204805(17)230128(10)201950 (01)00880304204805(17)230220(10)335870 (01)00880304204805(17)230320(10)671830 (01)00880304204805(17)230415(10)006130 (01)00880304204805(17)230508(10)085250 (01)00880304204805(17)230508(10)160010 (01)00880304204805(17)230606(10)359270 (01)00880304204805(17)230607(10)422910 (01)00880304204805(17)230607(10)423070 (01)00880304204805(17)230710(10)670240 (01)00880304204805(17)230824(10)970390 (01)00880304204805(17)230829(10)057550 (01)00880304204805(17)230829(10)906250 (01)00880304204805(17)230916(10)231030 (01)00880304204805(17)230930(10)455030 (01)00880304204805(17)230930(10)455360 (01)00880304204805(17)231011(10)266250 (01)00880304204805(17)231011(10)406310 (01)00880304204805(17)231018(10)162310 (01)00880304204805(17)231023(10)649440 (01)00880304204805(17)231109(10)649430 (01)00880304204805(17)231113(10)828560 (01)00880304204805(17)231124(10)766990 (01)00880304204805(17)231209(10)870250 (01)00880304204805(17)240109(10)945370 (01)00880304204805(17)240120(10)143410 (01)00880304204805(17)240121(10)143420 (01)00880304204805(17)240204(10)410710 (01)00880304204805(17)240206(10)421170 (01)00880304204805(17)240222(10)880810 (01)00880304204805(17)240226(10)506030 (01)00880304204805(17)240305(10)682970 (01)00880304204805(17)240316(10)919970 (01)00880304204805(17)240326(10)785240 (01)00880304204805(17)240413(10)551280 (01)00880304204805(17)240414(10)119550 (01)00880304204805(17)240422(10)751370 (01)00880304204805(17)240520(10)561490 (01)00880304204805(17)240522(10)485930 (01)00880304204805(17)240528(10)003950 (01)00880304204805(17)240604(10)409830 (01)00880304204805(17)240613(10)243880 (01)00880304204805(17)240618(10)649340 (01)00880304204805(17)240702(10)271930 (01)00880304204805(17)240706(10)139980 (01)00880304204805(17)240714(10)522000 (01)00880304204805(17)240806(10)453340 (01)00880304204805(17)240806(10)542810 (01)00880304204805(17)240813(10)542820 (01)00880304204805(17)240825(10)668240 (01)00880304204805(17)240825(10)453350 (01)00880304204805(17)240925(10)126380 (01)00880304204805(17)241008(10)589770 (01)00880304204805(17)241013(10)313940 (01)00880304204805(17)241023(10)313930 (01)00880304204805(17)241028(10)555550 (01)00880304204805(17)241029(10)712940 (01)00880304204805(17)241121(10)875300 (01)00880304204805(17)241211(10)039170 (01)00880304204805(17)241229(10)214050 (01)00880304204805(17)250124(10)154780 (01)00880304204805(17)250304(10)673520 (01)00880304204805(17)250306(10)685010 (01)00880304204805(17)250310(10)935690 (01)00880304204805(17)250331(10)294870 (01)00880304204805(17)250402(10)935680 (01)00880304204805(17)250403(10)411780 (01)00880304204805(17)250408(10)452970 (01)00880304204805(17)250421(10)580230 (01)00880304204805(17)250526(10)940620 (01)00880304204805(17)250625(10)299340 (01)00880304204805(17)250810(10)584000 (01)00880304204805(17)250904(10)819680 (01)00880304204805(17)250916(10)053300 (01)00880304204805(17)250916(10)819670 (01)00880304204805(17)251014(10)289370 (01)00880304204805(17)251103(10)496800 (01)00880304204805(17)251104(10)496770 (01)00880304204805(17)251218(10)496780 (01)00880304204805(17)251229(10)399830 (01)00880304204805(17)260115(10)095080 (01)00880304204805(17)260125(10)164510 (01)00880304204805(17)260126(10)095090 (01)00880304204805(17)260208(10)744250 (01)00880304204805(17)260208(10)652030 (01)00880304204805(17)260218(10)685720 (01)00880304204805(17)260322(10)095070 (01)00880304204805(17)260323(10)115700 (01)00880304204805(17)260328(10)115670 (01)00880304204805(17)260329(10)171340 (01)00880304204805(17)260419(10)115690 (01)00880304204805(17)260420(10)503740 (01)00880304204805(17)260421(10)115680 (01)00880304204805(17)260509(10)893580 (01)00880304204805(17)260518(10)747570 (01)00880304204805(17)260620(10)212300 (01)00880304204805(17)260715(10)436640 (01)00880304204805(17)260818(10)841670 (01)00880304204805(17)261125(10)705920 (01)00880304204805(17)
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Zimmer Biomet, Inc.
Reason for Recall:
Elevated levels of bacterial endotoxin and residual debris remain on the devices due to cleaning issue.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Copeland HA Resurfacing Head, MB/HA EAS Head; Item Nos. 11-114641 11-114642 11-114643 11-114644 11-114644S 11-114645 11-114646 11-114647 11-114648 11-114661 11-114662 11-114663 11-114664 11-114664S 11-114665 11-114666 11-114667 11-114668 11-114632 11-114633 Product Usage: Partial Shoulder Replacement.

Product Codes/Lot Numbers:

All lots manufactured between January 2008 - May 2019: 197000 838570 047410 869160 047400 047390 203030 203040 224470 482850 052950 076580 917420 326610 362350 565380 374150 420170 374160 882210 882090 746150 691040 969190 942350 726180 048880 385080 111310 013150 868010 638680 745730 780270 780240 013140 834290 623340 817490 028380 186120 409420 847030 679990 755600 409330 706270 169490 936000 400990 343410 235880 917410 872790 715870 179470 313330 324140 362650 305750 326590 920210 564070 305740 632120 684640 536380 564770 014640R 745910 629320 629350 014640 807570 946770 111280 946800 735450 108390 370610 385070 884110 013160 884080 621660 919220 836010 645380 706600 899340 712500 838730 963630 021010 896300 673910 669190 146510 613110 678830 335900 684340 106100 105780 140000 447810 154700 911280 363670 102500 931980 519820 114830 762310 076570 296510 917390 096130 059170 362810 552990 684260 587040 565220 484880 634740 485340 414520 906620 808780 678310 969290 013240 013250 014670 014850 384890 108310 808790 561410 808770 336710 018410 617860 313520 378380 579070 566190 927200 297800 673900 348320 702090 939400 736230 268220 055290 347210 347230 342820 725450 817310 153990 725430 698880 698890 823840 188130 037810 705530 602470 301330 463540 940980 028450 304010 491810 596490 684830 771630 731090 384570 343400 049790 057690 766800 513670 114820 915280 408980 715850 917370 917380 296500 076560 326600 510940 568130 363140 420030R 420030 808760 099520 917350 414510 629400 808720 014610 813330 044040 484870 621620 694280 694250 739520 991720 563690 671840 983580 250250 190920 231850 390870 757150 799950 517900 204080 764010 114810 249030 915250 061900 917340 076550 917360 131600 320520 348900 326030 565370 788320 610590 200890 462920 540650R 584540 540650 540610 691160 634350 485330 808710 085020 108330 906800 042230 367330 013270 100260 584530 759100 621970 808700 781920 834590 645360 654830 307730 580280 789060 295500 110720 296810 218340 971510 290820 984540 715090 130360 280430 896280 701290 322680 582620 912810 911710 178270 591690 191690 367880 867300 112070 266870 798940 568280 613100 028440 620410 620420 784160 046210 712950 443870 642180 503770 420330 134180 294890 628730 763910 251590 096000 381290 086080 208040 304890 314410 498860 498870 546180 831840 419740 828320 767110 917310 991930 771690 296490 260290 235790 917330 076520 076540 427590 326690 326700 484860 674000 540570 584510 584520 808670 553580 484850 746060 683370 808660 946820 049430 125700 385030 677300 917320 645350 371650 371910 919370 808680 089930 089920 485520 606500 208850 502490 298670 656410 045750 085450 313510 745020 721210 144450 378370 712490 431600 433430 297790 760970 842280 174340 250930 297880 934950 206630 912780 434850 201690 915720 182160 627310 411050 450540 644770 191740 763920 944360 613090 168190 716270 824710 565110 759500 279840 751390 313950 297070 102460 399850 728220 741280 062670 599590 070530 958220 079180 917290 917300 052930 132080 132090 261190 193570 227680 378250 326580 363150 305470 553570 787630 484840 634340 606650 227710 674890 654520 866510 399620 485320 808640 055710 942310 795900 969300 808650 055720 744940 561400 056140 809740 477760 354200 896230 322660 483840 930010 135000 329910 991820 449680 192060 479170 738870 626460 715840 917270 320620 881600 540140 808620 553610 808630 485310 606660 553611 808510 055700 758650 143790 055690 552750 422710 751380 973120 763930 233200 503760 305760 211440 715830 791670 035100 457730 209330 965510 324130 052920 911290 339120 586290 540090 364180 586300 276570 162030 162040 553940 785680 852640 054840 382030 374030 805470 134580 790650 302140 547610R 297990 547610 896180 356960 771130 372160 747300 787100 688710 343770 823820 352910 611740 475630 768550 768550R 465490 919980 485950 139990 284880 342280 266180 728880 875320 541510 652040 747580 331370 327140 169040 866470 715820 261020 245080 275200 308520 275190 296480 250150 245050 374480 563990 563970 558620 558630 586280 561200 561210 558610 786920 865980 865990 385950 196500 044100 973720 373980 301080 301090 373990 374000 772460 697320 021410 748760 987390 165460 986940 383960 245320 429590 358790 095990 018850 293160 293170 886190 855020 941050 965160 855030 815540 992450 171270 172200 934030 934020 299320 337530 971820 179730 299340 482040 499120 701660 374100 646370 373540 374110 684180 705520 646320 944070 944380 906950 802610 802590 057210 969620 144420 296660 144410 144430 429290 712460 183950 680430 760910 715080 917470 760980 242270 968580 132950 297960 297810 297850 399380 297820 673960 673920 896130 936160 939390 364060 770560 415910 322670 471660 740640 687610 947490 025190 574890 851130 627290 787090 787080 531340 135140 309650 191670 191680 547500 367870 450150 725280 688670 992130 787730 093270 797850 602450 602460 134990 686500 768520 824690 809340 991800 336150 946080 335880 201960 684430 505740 490180 759490 117200 024480 252230 406320 132610 143430 409410 925400 682980 435670 561500 342270 271940 827340 139770 161980 453360 542830 522760 875310 039180 473210 781530 610870 294890 947310 070400 069880 026340 372930 430830 819690 399840 496920 886220 980410 416120 363660 164590 821940 115710 503750 212310 462060 120690 881240 881180 079230 169030 079220 531660 327130 090650 211420 845510 866440 090660 045460 819530 700180 972280 500720 209350 081430 304580 304620 304590 304610 232410 304600 304630 304640 209360 304650 308090 304670 209340 304680 363390 304660 439320 308270 552440 439330 586260 395930 564370 395960 586270 865920 865910 081420 865900 014180 276590 188720 188710 188700 289960 308150 305600 305610 314160 742650 874490 822830 990810 484530 933900 043010 336660 366540 624460 292740 426590 021300 608740 757630 125560 125550 125570 256220 256230 256250 256260 256270 256240 301800 301810 256280 346750 398940 533890 533900 322450 463080 322460 606490 671690 398950 671680 016951 859320 242750 031320 031310 136950 601010 295710 580240 580260 580270 483470 580250 656390 603690 603710 603700 912390 912440 912410 912360 912460 108920 044790 045110 912350 912370 912400 912430 971030 170520 117340 117380 254630 316930 330330 117230 117240 179720 117250 218290 491020 490960 381410 381160 898760 144330 144320 302070 997400 997020 712510 406860 874810 547650 117620 712480 227950 258090 442140 441380 535900 651460 578700 487770 650360 715530 680500 995880 650300 968570 968560 715070 229180 760730 964980 399420 297940 297950 399410 570680 399430 297930 673950 954480 896270 939380 033680 983850 985750 109670 251080 251090 322640 519470 687600 693640 749050 915700 955480 142250 021050 419750 273850 531290 604870 627280 787070 627270 706110 240900 084560 367830 367860 367820 367840 669170 725250 367850 669140 688700 669130 669160 698670 688690 669180 458280 669120 883600 787720 125880 622120 602440 602430 134980 910760 205920 416030 329900 611720 686490 898590 037300 037310 063740 946070 965570 824680 991790 069560 809330 946060 113230 201950 335870 671830 006130 085250 160010 359270 422910 423070 670240 970390 057550 906250 231030 455030 455360 266250 406310 162310 649440 649430 828560 766990 870250 945370 143410 143420 410710 421170 880810 506030 682970 919970 785240 551280 119550 751370 561490 485930 003950 409830 243880 649340 271930 139980 522000 453340 542810 542820 668240 453350 126380 589770 313940 313930 555550 712940 875300 039170 214050 154780 673520 685010 935690 294870 935680 411780 452970 580230 940620 299340 584000 819680 053300 819670 289370 496800 496770 496780 399830 095080 164510 095090 744250 652030 685720 095070 115700 115670 171340 115690 503740 115680 893580 747570 212300 436640 841670 705920 866330 866320 169020 247060 547110 595380 595410 233730 531650 169000 595390 845490 090630 845500 090620 881170 866410 683040 090640 626430 211410 866400 626390 560760 035110 709380 052900 457800 132070 540070 539990 361570 586240 586250 605810 052880 886710 967150 886550 013320 013370 278090 668560 834550 656110 790070 656100 174700 933990 992440 701930 685180 374040 749140 237030 144310 277060 277070 555150 302060 594510 962630 442090 058150 297830 896150 673940 297870 487570 566520 566540 191660 658790 688660 893550 372670 824670 335860 024470 682960 421160 485920 729180 925510 485940 424440 353450 000130 522010 542800 927390 268300 555540 039150 552140 164570 685710 989050 533590 531640 866390 700170 915670 762330 971710 917650 052850 052860 132050 276630 132060 361560 416000 439280 416010 439310 439300 439260 361540 439270 439290 609990 876830 832940 876840 876850 013400 013380 144040 771280 935470 935480 339050 373950 301070 373970 483060 435000 813570 715820 715810 199740 199750 199760 898100 919050 391980 383940 545820 383950 018840 870250 095970 846240 431610 148760 431620 693580 341560 886180 698720 932880 012030 616130 616150 729610 398630 162020 162000 189050 279040 162010 289360 392050 600250 392060 580320 631730 710790 891500 553930 852630 934010 010230 972020 155640 226070 117570 179670 179680 505250 570330 373510 373500 374050 373520 373490 639310 566150 786760 786190 786780 841980 898030 898050 898040 898020 134530 282140 898750 452970 431680 277050 302050 549460 302040 408370 406850 710750 927560 968540 227940 258110 113160 673420 441370 552900 680450 650350 715060 014620 916220 650310 940490 048900 968550 283510 089290 715160 148040 148050 229070 121320 534520 986580 986570 033670 251070 322610 394060 322540 332100 394080 325680 553970 554000 687580 867010 603250 711220 762890 820860 819490 553980 693630 749040 687590 746390 714720 831850 021040 178260 296580 695730 574880 787060 627260 123210 191650 084550 453810 191620 191630 276180 453790 367810 453800 688680 823800 911030 892820 093190 498410 498370 498400 498420 622110 317640 910770 236680 329880 686480 611700 686470 766080 037290 037280 686240 824660 824640 965550 946040 321670 965560 991780 334820 476440 490200 113220 511220 696970 671820 847280 024460 013420 175540 429810 454110 649410 649420 649390 852280 933990 132600 945360 143400 503920 745440 374190 506020 435660 682950 421150 919950 919960 254820 843980 379540 396960 368060 550900 082420 975920 927580 927590 214760 271920 347570 284940 214750 522170 563850 594840 403940 342260 681280 542790 453330 542780 815540 982470 927350 268150 927320 927570 498820 039140 297060 151700 294860 452960 845700 940610 583990 894920 050620 247840 981160 496810 825210 399810 416110 164490 741270 095060 769680 164500 652010 060010 548410 523030 115660 961870 212280 479190 705910 120680 881220 168980 079190 866360 531630 533580 845450 090570 595340 845480 090580 090590 090610 045440 045450 626360 626380 687850 866380 907610 251580 095980 191250 406800 042590 179690 314420 374080 374090 478790 478800 646330 304880 208030 188710 715810 700140 500710 431800 097420 279660 304530 304520 304560 097410 719660 097390 431790 097400 304550 324120 510880 304570 439250 469810 586310 469830 539930 469820 586340 539830 539960 586230 439240 467930 586320 746210 415990 438920 583490 608790 583410 876820 876810 876800 583430 876790 967110 967130 012890 013470 917640 044110 279910 512330 044090 013010 669230 143910 199730 384890 390050 391950 742820 391960 877280 545810 985610 018830 148750 251570 039120 552940 341550 431590 431600 677340 159200 837890 295670 295680 295650 295660 161990 289370 289350 642680 852620 580300 790080 710780 972010 140520 170480 187670 179710 334720 179660 992430 570190 373460 237080 632890 373480 639230 373470 568300 646360 802580 898010 898740 802570 277030 277040 478520 348600 431580 452890 302030 408360 302020 547630 406830 258100 183940 680530 482740 650320 715150 015690 996510 967270 968540 148030 182470 148020 968530 148010 597820 597690 399390 399400 297920 897640 897650 033660 058130 079670 079730 266440 355010 433340 434840 519460 501080 372120 569900 693620 195970 985090 985100 108480 500790 232350 400180 531280 948140 695720 574870 986890 191600 191610 084540 367790 367800 241000 496350 688640 457400 698580 688650 725220 944350 944340 037790 266860 355420 413900 413890 797840 236650 236660 329870 686460 766070 611690 318020 063750 686100 686230 437910 967220 768530 824630 965540 991760 809320 201940 824620 505710 505720 759480 006110 037620 006120 095360 160000 214620 558800 742590 861710 906240 024440 455010 406290 455110 718590 007560 945350 421140 785230 339690 506000 079870 919940 109370 471380 427230 561470 427260 561450 196120 139970 927560 769300 196020 214740 342250 196070 608160 214730 904540 423120 091560 927550 542760 542770 590970 498570 313900 555530 589820 370320 105040 875290 039100 297050 685020 673230 454490 642190 935670 763920 102140 949660 349900 368180 666420 903230 231800 164470 163410 343380 744060 080070 164480 651990 961850 115640 115650 661790 114790 436620 436630 479210 989100 841660 705900 881200 866310 168970 531620 584940 233720 558520 533560 584920 845430 249150 845440 560730 211390 045410 626330 420300 715800 754520 431750 431770 965500 307790 965490 052870 132040 306940 754580 539680 539780 539720 536370 886130 865870 819090 865850 886140 783740 013410 044060 044080 865880 512410 645530 553990 554000 603680 789330 971500 313570 805630 897890 969610 163440 302010 549540 984530 715140 650330 940300 968520 399370 125560 297840 673930 556720 880330 880530 121110 110400 332090 332070 434820 486740 372110 915690 749030 153450 531270 191730 367780 507100 669100 669110 911020 797830 910780 063760 686160 768540 965530 113210 490190 906220 458120 766980 421130 682940 682940R 109270 243870 550890 207590 973990 271910 542750 927290 927310 496940 343370 562390 401280 115630 212270 705890 866350 233700 276660 090750 090740 560720 045390 085450 719650 715780 027290 469790R 469780 364140 469770 469790 469800 585650 608230 586220 633280 632390 013600 044070 013490 044050 130140 780230 847450 561630 685700 120670 744620 533550 090730 019770 563730 589160 722600 084620 626500 420370 777480 420350 420380 211500 211510 746160 045470 084610 308510 626490 159240 211490 362640 405790 405800 544750 544770 681260 684450 626480 420340 845550 103840 211470 342030 362680 572910 196650 084590 845540 188020 229970 247610 305730 405780 540670 553000 808740 808750 918980 521370 018560 103810 103820 131610 209320 211460 323210 437960 556500 766160 959910 085900 086150 873360 736580 560790 029780 050050 076530 103800 119850 285340 315810 405770 430800 455630 683360 683380 719730 726350 043990 044780 049110 833870 833880 880280 880290 880310 880320 052940 159200 211450 317740 327530 450430 540540 944420 596950 845530 866480 084580 084560 308500 159120 308900 365800 405730 405740 405750 405760 408990 906680 917280 590200 159090 293200 456970 469870 489230 564010 571400 587700 616190 683800 779990 866000 885370 078470 895090 005880 086490 093240 093250 103770 103780 103790 117030 191480 232180 293180 293190 296860 354660 365420 366260 385910 444820 456950 489210 489220 571370 571380 571390 616160 616170 616180 683710 683750 683760 712950 729630 729640 779980 825530 855010 865930 969900 969910 087870 092050 308490 103730 103740 103750 103760 157950 208840 295700 295720 296130 297580 298700 301310 301790 303540 312420 318000 318090 318100 323200 335010 339120 354150 374140 383100 392330 392340 420280 420290 420300 420310 420320 420330 422320 426710 431130 431140 433080 445070 450980 457340 519680 556480 556490 560640 587690 648490 683020 683030 689740 726080 766140 766150 815620 825520 887120 898820 938830 945810 103720 159080 260750 295730 301090 329000 417070 418320 444810 583580 570080 514830 308480 560750 012020 044340 066810 103690 103700 103710 148070 177660 260740 293150 306340 307530 317970 323870 328990 329890 366250 397850 427610 430820 432520 444800 451570 489190 489200 571330 571340 571350 571360 587660 587670 587680 588540 616120 682960 766130 779960 779970 825510 870120 885360 898090 941040 761210 420310 308470 044330 103670 103680 107340 156880 312970 323860 328980 353340 384950 385030 385070 385140 385170 385180 385190 385210 385230 385240 385260 385270 385290 385300 385310 385320 385330 427600 439290 444790 456940 469850 469860 479160 563940 587650 588530 736720 766120 781840 825500 962870 103650 103660 147990 298690 323190 417300 431760 433070 683900 766110 825490 005870 159030 159040 159060 683890 UDI: (01)00880304439962(17)180628(10)882090 (01)00880304439962(17)180728(10)942350 (01)00880304439962(17)180728(10)726180 (01)00880304439962(17)180828(10)111310 (01)00880304439962(17)180828(10)013150 (01)00880304439962(17)180828(10)868010 (01)00880304439962(17)190128(10)623340 (01)00880304439962(17)230128(10)028380 (01)00880304439962(17)230830(10)186120 (01)00880304439962(17)240204(10)409420 (01)00880304439962(17)240207(10)847030 (01)00880304439962(17)240221(10)679990 (01)00880304439962(17)240301(10)755600 (01)00880304439962(17)240505(10)409330 (01)00880304439962(17)240612(10)706270 (01)00880304439962(17)240708(10)169490 (01)00880304439962(17)251222(10)936000 (01)00880304439962(17)251229(10)400990 (01)00880304439962(17)260425(10)343410 (01)00880304439962(17)261121(10)235880 (01)00880304439917(17)180328(10)313330 (01)00880304439917(17)180328(10)324140 (01)00880304439917(17)180701(10)014640R (01)00880304439917(17)180828(10)884080 (01)00880304439917(17)201228(10)838730 (01)00880304439917(17)230324(10)335900 (01)00880304439917(17)230325(10)684340 (01)00880304439917(17)230418(10)106100 (01)00880304439917(17)230709(10)105780 (01)00880304439917(17)240825(10)140000 (01)00880304439917(17)241126(10)447810 (01)00880304439917(17)250126(10)154700 (01)00880304439917(17)250219(10)911280 (01)00880304439917(17)260207(10)363670 (01)00880304439917(17)260317(10)102500 (01)00880304439917(17)260426(10)931980 (01)00880304439917(17)260518(10)519820 (01)00880304439917(17)260624(10)114830 (01)00880304439955(17)180628(10)634740 (01)00880304439955(17)180701(10)485340 (01)00880304439955(17)180728(10)969290 (01)00880304439955(17)180828(10)108310 (01)00880304439955(17)181028(10)808770 (01)00880304439955(17)200328(10)313520 (01)00880304439955(17)220228(10)698890 (01)00880304439955(17)220528(10)037810 (01)00880304439955(17)230116(10)028450 (01)00880304439955(17)230204(10)304010 (01)00880304439955(17)230213(10)491810 (01)00880304439955(17)230221(10)596490 (01)00880304439955(17)230319(10)684830 (01)00880304439955(17)251221(10)771630 (01)00880304439955(17)251222(10)731090 (01)00880304439955(17)251223(10)384570 (01)00880304439955(17)260208(10)343400 (01)00880304439955(17)260210(10)049790 (01)00880304439955(17)260318(10)057690 (01)00880304439955(17)260322(10)766800 (01)00880304439955(17)260427(10)513670 (01)00880304439955(17)270803(10)114820 (01)00880304439894(17)180601(10)420030R (01)00880304439894(17)180628(10)808760 (01)00880304439894(17)180728(10)917350 (01)00880304439894(17)180728(10)629400 (01)00880304439894(17)180728(10)808720 (01)00880304439894(17)180728(10)014610 (01)00880304439894(17)230319(10)563690 (01)00880304439894(17)230327(10)671840 (01)00880304439894(17)230415(10)983580 (01)00880304439894(17)230520(10)250250 (01)00880304439894(17)250611(10)190920 (01)00880304439894(17)251016(10)231850 (01)00880304439894(17)251021(10)390870 (01)00880304439894(17)251222(10)757150 (01)00880304439894(17)251228(10)799950 (01)00880304439894(17)260509(10)517900 (01)00880304439894(17)260620(10)204080 (01)00880304439894(17)270305(10)764010 (01)00880304439894(17)270921(10)114810 (01)00880304439894(17)271016(10)249030 (01)00880304439924(17)180228(10)076550 (01)00880304439924(17)180601(10)540650R (01)00880304439924(17)180828(10)584530 (01)00880304439924(17)191228(10)295500 (01)00880304439924(17)200128(10)296810 (01)00880304439924(17)220228(10)367880 (01)00880304439924(17)230326(10)028440 (01)00880304439924(17)241029(10)620410 (01)00880304439924(17)241117(10)620420 (01)00880304439924(17)241201(10)784160 (01)00880304439924(17)241204(10)046210 (01)00880304439924(17)241208(10)712950 (01)00880304439924(17)250112(10)443870 (01)00880304439924(17)250126(10)642180 (01)00880304439924(17)260523(10)503770 (01)00880304439924(17)280320(10)420330 (01)00880304439900(17)180201(10)917310 (01)00880304439900(17)180728(10)808660 (01)00880304439900(17)201028(10)712490 (01)00880304439900(17)210328(10)842280 (01)00880304439900(17)210428(10)912780 (01)00880304439900(17)220228(10)191740 (01)00880304439900(17)230121(10)824710 (01)00880304439900(17)230220(10)565110 (01)00880304439900(17)230416(10)759500 (01)00880304439900(17)230607(10)279840 (01)00880304439900(17)240715(10)751390 (01)00880304439900(17)241023(10)313950 (01)00880304439900(17)250130(10)297070 (01)00880304439900(17)250313(10)102460 (01)00880304439900(17)260114(10)399850 (01)00880304439900(17)251223(10)728220 (01)00880304439900(17)251229(10)741280 (01)00880304439900(17)260126(10)062670 (01)00880304439900(17)260216(10)599590 (01)00880304439900(17)260315(10)070530 (01)00880304439931(17)180328(10)193570 (01)00880304439931(17)180328(10)227680 (01)00880304439931(17)180701(10)674890 (01)00880304439931(17)180728(10)399620 (01)00880304439931(17)180728(10)485320 (01)00880304439931(17)180728(10)808640 (01)00880304439931(17)180728(10)055710 (01)00880304439931(17)180828(10)055720 (01)00880304439931(17)230215(10)991820 (01)00880304439931(17)230218(10)449680 (01)00880304439931(17)230225(10)192060 (01)00880304439931(17)260726(10)479170 (01)00880304439931(17)260729(10)738870 (01)00880304439931(17)280302(10)626460 (01)00880304439948(17)180628(10)553610 (01)00880304439948(17)240121(10)422710 (01)00880304439948(17)240421(10)751380 (01)00880304439948(17)240825(10)973120 (01)00880304439948(17)250210(10)763930 (01)00880304439948(17)250326(10)233200 (01)00880304439948(17)260524(10)503760 (01)00880304439948(17)260628(10)305760 (01)00880304439948(17)280301(10)211440 (01)00880304204829(17)180228(10)965510 (01)00880304204829(17)180528(10)586290 (01)00880304204829(17)191228(10)785680 (01)00880304204829(17)191228(10)852640 (01)00880304204829(17)200428(10)805470 (01)00880304204829(17)200528(10)134580 (01)00880304204829(17)200628(10)790650 (01)00880304204829(17)200628(10)302140 (01)00880304204829(17)210201(10)547610R (01)00880304204829(17)210528(10)771130 (01)00880304204829(17)210528(10)372160 (01)00880304204829(17)220528(10)352910 (01)00880304204829(17)230218(10)475630 (01)00880304204829(17)230219(10)768550 (01)00880304204829(17)230219(10)768550R (01)00880304204829(17)240209(10)465490 (01)00880304204829(17)240319(10)919980 (01)00880304204829(17)240518(10)485950 (01)00880304204829(17)240706(10)139990 (01)00880304204829(17)240721(10)284880 (01)00880304204829(17)240810(10)342280 (01)00880304204829(17)240916(10)266180 (01)00880304204829(17)241030(10)728880 (01)00880304204829(17)241201(10)875320 (01)00880304204829(17)250105(10)541510 (01)00880304204829(17)260216(10)652040 (01)00880304204829(17)260526(10)747580 (01)00880304204829(17)260630(10)331370 (01)00880304204829(17)270305(10)327140 (01)00880304204829(17)270419(10)169040 (01)00880304204829(17)280123(10)866470 (01)00880304204812(17)181201(10)373980 (01)00880304204812(17)190928(10)965160 (01)00880304204812(17)200501(10)906950 (01)00880304204812(17)210528(10)322670 (01)00880304204812(17)210528(10)471660 (01)00880304204812(17)210628(10)687610 (01)00880304204812(17)211128(10)531340 (01)00880304204812(17)230123(10)336150 (01)00880304204812(17)230128(10)946080 (01)00880304204812(17)230204(10)335880 (01)00880304204812(17)230204(10)201960 (01)00880304204812(17)230225(10)684430 (01)00880304204812(17)230228(10)505740 (01)00880304204812(17)230319(10)490180 (01)00880304204812(17)230327(10)759490 (01)00880304204812(17)230822(10)117200 (01)00880304204812(17)230826(10)024480 (01)00880304204812(17)230916(10)252230 (01)00880304204812(17)231011(10)406320 (01)00880304204812(17)240120(10)132610 (01)00880304204812(17)240120(10)143430 (01)00880304204812(17)240201(10)409410 (01)00880304204812(17)240221(10)925400 (01)00880304204812(17)240226(10)682980 (01)00880304204812(17)240227(10)435670 (01)00880304204812(17)240507(10)561500 (01)00880304204812(17)240706(10)342270 (01)00880304204812(17)240717(10)271940 (01)00880304204812(17)240806(10)827340 (01)00880304204812(17)240825(10)139770 (01)00880304204812(17)240828(10)161980 (01)00880304204812(17)240904(10)453360 (01)00880304204812(17)240905(10)542830 (01)00880304204812(17)241013(10)522760 (01)00880304204812(17)241203(10)875310 (01)00880304204812(17)241209(10)039180 (01)00880304204812(17)250415(10)473210 (01)00880304204812(17)250426(10)781530 (01)00880304204812(17)250430(10)610870 (01)00880304204812(17)250430(10)294890 (01)00880304204812(17)250519(10)947310 (01)00880304204812(17)250526(10)070400 (01)00880304204812(17)250527(10)069880 (01)00880304204812(17)250603(10)026340 (01)00880304204812(17)250618(10)372930 (01)00880304204812(17)250717(10)430830 (01)00880304204812(17)251006(10)819690 (01)00880304204812(17)251029(10)399840 (01)00880304204812(17)251106(10)496920 (01)00880304204812(17)251222(10)886220 (01)00880304204812(17)260108(10)980410 (01)00880304204812(17)260120(10)416120 (01)00880304204812(17)260125(10)363660 (01)00880304204812(17)260223(10)164590 (01)00880304204812(17)260324(10)821940 (01)00880304204812(17)260421(10)115710 (01)00880304204812(17)260425(10)503750 (01)00880304204812(17)260620(10)212310 (01)00880304204812(17)261107(10)462060 (01)00880304204812(17)261115(10)120690 (01)00880304204812(17)261215(10)881240 (01)00880304204812(17)270114(10)881180 (01)00880304204812(17)270305(10)079230 (01)00880304204812(17)270315(10)169030 (01)00880304204812(17)270319(10)079220 (01)00880304204812(17)270422(10)531660 (01)00880304204812(17)270825(10)327130 (01)00880304204812(17)271010(10)090650 (01)00880304204812(17)280108(10)211420 (01)00880304204812(17)280112(10)845510 (01)00880304204812(17)280117(10)866440 (01)00880304204812(17)280220(10)090660 (01)00880304204812(17)280305(10)045460 (01)00880304204812(17)280327(10)819530 (01)00880304204805(17)201228(10)680500 (01)00880304204805(17)220101(10)367830 (01)00880304204805(17)220828(10)329900 (01)00880304204805(17)221228(10)824680 (01)00880304204805(17)230101(10)991790 (01)00880304204805(17)230117(10)113230 (01)00880304204805(17)230128(10)201950 (01)00880304204805(17)230220(10)335870 (01)00880304204805(17)230320(10)671830 (01)00880304204805(17)230415(10)006130 (01)00880304204805(17)230508(10)085250 (01)00880304204805(17)230508(10)160010 (01)00880304204805(17)230606(10)359270 (01)00880304204805(17)230607(10)422910 (01)00880304204805(17)230607(10)423070 (01)00880304204805(17)230710(10)670240 (01)00880304204805(17)230824(10)970390 (01)00880304204805(17)230829(10)057550 (01)00880304204805(17)230829(10)906250 (01)00880304204805(17)230916(10)231030 (01)00880304204805(17)230930(10)455030 (01)00880304204805(17)230930(10)455360 (01)00880304204805(17)231011(10)266250 (01)00880304204805(17)231011(10)406310 (01)00880304204805(17)231018(10)162310 (01)00880304204805(17)231023(10)649440 (01)00880304204805(17)231109(10)649430 (01)00880304204805(17)231113(10)828560 (01)00880304204805(17)231124(10)766990 (01)00880304204805(17)231209(10)870250 (01)00880304204805(17)240109(10)945370 (01)00880304204805(17)240120(10)143410 (01)00880304204805(17)240121(10)143420 (01)00880304204805(17)240204(10)410710 (01)00880304204805(17)240206(10)421170 (01)00880304204805(17)240222(10)880810 (01)00880304204805(17)240226(10)506030 (01)00880304204805(17)240305(10)682970 (01)00880304204805(17)240316(10)919970 (01)00880304204805(17)240326(10)785240 (01)00880304204805(17)240413(10)551280 (01)00880304204805(17)240414(10)119550 (01)00880304204805(17)240422(10)751370 (01)00880304204805(17)240520(10)561490 (01)00880304204805(17)240522(10)485930 (01)00880304204805(17)240528(10)003950 (01)00880304204805(17)240604(10)409830 (01)00880304204805(17)240613(10)243880 (01)00880304204805(17)240618(10)649340 (01)00880304204805(17)240702(10)271930 (01)00880304204805(17)240706(10)139980 (01)00880304204805(17)240714(10)522000 (01)00880304204805(17)240806(10)453340 (01)00880304204805(17)240806(10)542810 (01)00880304204805(17)240813(10)542820 (01)00880304204805(17)240825(10)668240 (01)00880304204805(17)240825(10)453350 (01)00880304204805(17)240925(10)126380 (01)00880304204805(17)241008(10)589770 (01)00880304204805(17)241013(10)313940 (01)00880304204805(17)241023(10)313930 (01)00880304204805(17)241028(10)555550 (01)00880304204805(17)241029(10)712940 (01)00880304204805(17)241121(10)875300 (01)00880304204805(17)241211(10)039170 (01)00880304204805(17)241229(10)214050 (01)00880304204805(17)250124(10)154780 (01)00880304204805(17)250304(10)673520 (01)00880304204805(17)250306(10)685010 (01)00880304204805(17)250310(10)935690 (01)00880304204805(17)250331(10)294870 (01)00880304204805(17)250402(10)935680 (01)00880304204805(17)250403(10)411780 (01)00880304204805(17)250408(10)452970 (01)00880304204805(17)250421(10)580230 (01)00880304204805(17)250526(10)940620 (01)00880304204805(17)250625(10)299340 (01)00880304204805(17)250810(10)584000 (01)00880304204805(17)250904(10)819680 (01)00880304204805(17)250916(10)053300 (01)00880304204805(17)250916(10)819670 (01)00880304204805(17)251014(10)289370 (01)00880304204805(17)251103(10)496800 (01)00880304204805(17)251104(10)496770 (01)00880304204805(17)251218(10)496780 (01)00880304204805(17)251229(10)399830 (01)00880304204805(17)260115(10)095080 (01)00880304204805(17)260125(10)164510 (01)00880304204805(17)260126(10)095090 (01)00880304204805(17)260208(10)744250 (01)00880304204805(17)260208(10)652030 (01)00880304204805(17)260218(10)685720 (01)00880304204805(17)260322(10)095070 (01)00880304204805(17)260323(10)115700 (01)00880304204805(17)260328(10)115670 (01)00880304204805(17)260329(10)171340 (01)00880304204805(17)260419(10)115690 (01)00880304204805(17)260420(10)503740 (01)00880304204805(17)260421(10)115680 (01)00880304204805(17)260509(10)893580 (01)00880304204805(17)260518(10)747570 (01)00880304204805(17)260620(10)212300 (01)00880304204805(17)260715(10)436640 (01)00880304204805(17)260818(10)841670 (01)00880304204805(17)261125(10)705920 (01)00880304204805(17)

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0527-2020

Related Recalls

Zimmer Biomet is conducting a medical device correction to update the Instructions for Use (IFU) and the Surgical Techniques for the Polaris Spinal System Translation Screw and Cypher MIS Screw System. Users should be aware that changes were made to the contraindications section indicating that the Translation Screws should not be used with a direct current stimulation, as a high probability of construct failure exists when using Translation Screws containing CoCr with direct current simulation system.

Nov 17, 2020 Other Medical Devices Nationwide View Details →