Varian On-Board Imager Advanced Imaging System. The On-Board Imager device is used for verification of correct patient position in relation to isocenter and verification of the treatment fields in relation to anatomical and/or fiducial land¿¿ marks.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Part Number P1003290001
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Varian Medical Systems, Inc.
Reason for Recall:
Varian MedicalSystems' X-ray generator Supplier (CPI) has identified that 32 VMS150 X-ray Generators were shipped that are missing a ground wire between the EMIbraid on the induc­ tion assembly cable and the ground stud on the lid of the High Voltage Module.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Varian On-Board Imager Advanced Imaging System. The On-Board Imager device is used for verification of correct patient position in relation to isocenter and verification of the treatment fields in relation to anatomical and/or fiducial land¿¿ marks.

Product Codes/Lot Numbers:

Part Number P1003290001

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0526-2015

Related Recalls

After a recent upgrade to the collimator software version 8.5, the firm became aware that the multi-leaf collimator leaves did not move during an arc treatment. The issue is related to a transient carriage primary- secondary interlock due to carriage fault on the MLC that immediately preceded the initiation of the arc treatment. The issue will occur only with MLC software version 8.5 and only affects conformal arc treatments, including VMAT and RapidArc. There have been no reports of adverse health consequences due to this issue.

Aug 12, 2019 Infusion Pumps Nationwide View Details →

There is an anomaly with the Eclipse(TM) Treatment Planning System [TPS] Proton Convolution Superposition [PCS] dose calculation algorithm. The PCS algo- rithm calculates the water equivalent range incorrectly for non-square 3D CT images (either different number of pixels in X and Y, or non-square pixels). Before the dose is calculated, Eclipse resamples the CT images to create a calculation image with a maximum resolution of 256 x 256. The PCS algorithm assumes uniform resolution in X and Y directions, or that X = Y for all images, and erroneously sets the Y = X for dose calcu- lation. For images that are not square, where either X<>Y, or the length of X does not equal the length of Y, the computed water equivalent range R is erroneous compared to the correct range R: 1) X=Y; R =R (R is correct); 2) X<Y: R <R (R is too small) and 3) X>Y R >R (R is too big).

Dec 26, 2018 Other Medical Devices Nationwide View Details →

Reports have been received of an anomaly that can result in a treatment without intended gating (respiratory tracking/monitoring). This issue occurs when a patient planned with gating is treated on more than one system.

Aug 20, 2018 Infusion Pumps Nationwide View Details →