Kreatech PapV-16, Flu labeled ISH probe, to detect presence of HPV-16 DNA inserts in human tissue, Model Number 40V016V495
Class I - DangerousWhat Should You Do?
- Check if you have this product: UDI-DI: (01)08718858480241 Lot Number: 00088454
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Kreatech Biotechnology Bv
- Reason for Recall:
- The DNA fragments for this specific lot number are too large, resulting in potential background around the cells due to suboptimal hybridization. An absence of signal could also be observed as a result of the suboptimal hybridization of the larger DNA fragments. The user may interpret potential background as a false positive result and an absence of signal as a false negative result, if proper positive and negative controls are not included.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Kreatech PapV-16, Flu labeled ISH probe, to detect presence of HPV-16 DNA inserts in human tissue, Model Number 40V016V495
Product Codes/Lot Numbers:
UDI-DI: (01)08718858480241 Lot Number: 00088454
Distribution:
Distributed in: US, FL, NY
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0522-2025