WANG Transbronchial Aspiration Needles Cat. No. MW-222, WANG Cytology Needles, Central or Peripheral Regions, 22 g x 13 mm long needle. --- Common Name: Bronchoscope (flexible or rigid) and accessories. The ConMed WANG Transbronchial Aspiration Needles are designed for use with a bronchoscope. They are used to puncture the tracheobronchial wall and to aspirate sufficient tissue and/or cell specimens to stage bronchogenic carcinoma. These are disposable sterile, single-use devices.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Begining Lot 1207251, Ending Lot Code 1207251; Beginning Lot Code 1209171, Ending Lot Code 1210151. Lot codes on boxes and packaging contain a lot code in the following form 12 09 17 1 where 12 represents the year (2012), 09 represents the month (September), 17 is the day of the month, and 1 is the Manufacturing Shift Code (1, 2, 3, 4, 5, 6 or 9).
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- ConMed Corporation
- Reason for Recall:
- ConMed received complaints from a user facility reporting that the WANG Transbronchial Aspiration Needles would not insert properly through the instrument channel of a 2.0 mm bronchoscope. If the device will not fit through the instrument channel of a 2.0 mm bronchoscope, it must be replaced with an alternate device and could result in minor inconvenience to the user.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
WANG Transbronchial Aspiration Needles Cat. No. MW-222, WANG Cytology Needles, Central or Peripheral Regions, 22 g x 13 mm long needle. --- Common Name: Bronchoscope (flexible or rigid) and accessories. The ConMed WANG Transbronchial Aspiration Needles are designed for use with a bronchoscope. They are used to puncture the tracheobronchial wall and to aspirate sufficient tissue and/or cell specimens to stage bronchogenic carcinoma. These are disposable sterile, single-use devices.
Product Codes/Lot Numbers:
Begining Lot 1207251, Ending Lot Code 1207251; Beginning Lot Code 1209171, Ending Lot Code 1210151. Lot codes on boxes and packaging contain a lot code in the following form 12 09 17 1 where 12 represents the year (2012), 09 represents the month (September), 17 is the day of the month, and 1 is the Manufacturing Shift Code (1, 2, 3, 4, 5, 6 or 9).
Distribution:
Distributed in: CA, FL, IA, IN, MA, MD, ME, MO, MS, NC, NM, OH, PA, TX
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0502-2013
Related Recalls
Suction ports may potentially be occluded on the affected devices. If this occurs, the affected devices will not be able to perform the suction function during use, which could cause a delay in procedure and could require another device to be used.
Tip Guides are potentially misaligned laterally which could affect the accuracy of the guide system and could cause a delay in procedure and may require another device or alternate surgical method to be used
Tip Guides are potentially misaligned laterally which could affect the accuracy of the guide system and could cause a delay in procedure and may require another device or alternate surgical method to be used