Helion S Exam Light or H300 The device is intended to locally illuminate an examination area of the patient's body with high intensity light in a clinic or doctor's practice.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Unknown
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Trumpf Medical Systems, Inc.
Reason for Recall:
Possible fatigue failure of the plastic joint may occur after an average use of seven years.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Helion S Exam Light or H300 The device is intended to locally illuminate an examination area of the patient's body with high intensity light in a clinic or doctor's practice.

Product Codes/Lot Numbers:

Unknown

Distribution:

Distributed in: US, AR, CA, CO, KS, MA, MI, MO, NJ, NY, NC, OH, OK, PA, SD, TX, WA

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0484-2014

Related Recalls

Trumpf has identified six incidents of pinched cables.

Oct 25, 2016 Other Medical Devices Nationwide View Details β†’

Three complaints describing TruSystem tables unexpectedly descending approximately 20 centimeters.

Jul 10, 2016 Other Medical Devices View Details β†’