Allegedly counterfeit batteries manufactured by CSB batteries being used with Plum Infusion Systems: Plum 360" Infusion System, Pump List Number 30010; Plum A+ & Plum A+3" Infusion Systems, Pump List Numbers: 11005, 11971, 12391, 12618, 20678, 20679, 20792, 60529, 12348, 1197
Class I - DangerousWhat Should You Do?
- Check if you have this product: All batteries WITHOUT the following information on the label: 1) ICU Medlcal Test Label, 2) CE Mark, and WITH a Date code (yellow label found on side of battery): W2401xxxx - W2406xxx
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- ICU Medical, Inc.
- Reason for Recall:
- ICU Medical has received reports of allegedly counterfeit CSB batteries being used with Plum Infusion Systems. While these non-OEM batteries are visually similar to the Plum batteries supplied by ICU Medical, they are in fact not the same batteries and have not been tested or validated for use with Plum Infusion Systems. Preliminary reports suggest that these non-OEM batteries fail to hold their charge and the pump may display messages to replace batteries earlier than expected. OSI Batteries and their customers are distributing these non-OEM CSB batteries without authorization from ICU Medical.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Allegedly counterfeit batteries manufactured by CSB batteries being used with Plum Infusion Systems: Plum 360" Infusion System, Pump List Number 30010; Plum A+ & Plum A+3" Infusion Systems, Pump List Numbers: 11005, 11971, 12391, 12618, 20678, 20679, 20792, 60529, 12348, 1197
Product Codes/Lot Numbers:
All batteries WITHOUT the following information on the label: 1) ICU Medlcal Test Label, 2) CE Mark, and WITH a Date code (yellow label found on side of battery): W2401xxxx - W2406xxx
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0482-2025
Related Recalls
Plum Duo Precision IV Pump, 40002-0403
ICU Medical
Under certain conditions, the pump may not display a soft limit alert, and will not notify the clinician of a soft limit violation before confirming and starting delivery. This may potentially lead to under- or over-delivery.
ICU Medical Plum Duo Infusion Pump, List Numbers: 400020401 (US), 400021001 (Philippines)
ICU Medical
Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is delivered from the primary line container at the piggyback therapy rate after the piggyback therapy is complete. In certain cases, as described below, the clinician may receive an Upper Hard Limit (UHL) or Lower Hard Limit (LHL) Violation message when programming a piggyback flush, which prevents the flush from being programmed.
Tego Connector: REF: 011-D1000, 01C-D1000, 055-D1000, D1000; Conector Tego: REF: LAT-D1000
ICU Medical
Needle free access device intended as accessory to vascular access used in hemodialysis or accessory to intravascular administration sets has a silicone seal which may bulge at the top, separate from the body, tear; may result in fluid leaks, occluded fluid flow, inability to inject/withdraw blood product, delayed/interrupted therapy, exposure to biological contaminants, air infused into the body