Total Toe Instrument Kit, Labeled in part: "Instrument Set, 15mm Toe, Dorsal Flange Combi" Part Number: 9000-3002 The component in question is the hex driver (part number 2009-4001).
Class I - DangerousWhat Should You Do?
- Check if you have this product: SN 000242 SN 000244 SN 000246 SN 000247
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Arthrosurface, Inc.
- Reason for Recall:
- An oversight in the implementation of the design change record resulted in outdated components being shipped to customers. The component did not pass a half-cycle sterilization.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Total Toe Instrument Kit, Labeled in part: "Instrument Set, 15mm Toe, Dorsal Flange Combi" Part Number: 9000-3002 The component in question is the hex driver (part number 2009-4001).
Product Codes/Lot Numbers:
SN 000242 SN 000244 SN 000246 SN 000247
Distribution:
Distributed in: US, PA, WV, CA
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0475-2016
Related Recalls
Fully threaded Taper Post Fixation components not cleared for marketing in the US with current indications.
Fully threaded Taper Post Fixation components not cleared for marketing in the US with current indications.