Atellica IM 1600 Analyzer - Product Usage: automated, immunoassay analyzers designed to perform in vitro diagnostic tests on clinical specimens. Siemens Material Number (SMN): 11066000
Class I - DangerousWhat Should You Do?
- Check if you have this product: Software version: V1.23.1 or lower UDI: 00630414002026
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Siemens Healthcare Diagnostics, Inc.
- Reason for Recall:
- Test Definition scanning may reset custom settings to defaults. When a 2D Master Curve and TDef barcode for a new kit lot of reagent is scanned and the TDef version is a newer version than the version that is currently on the system, some of the customer defined settings for that assay may reset to default values
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Atellica IM 1600 Analyzer - Product Usage: automated, immunoassay analyzers designed to perform in vitro diagnostic tests on clinical specimens. Siemens Material Number (SMN): 11066000
Product Codes/Lot Numbers:
Software version: V1.23.1 or lower UDI: 00630414002026
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0474-2021
Related Recalls
Food Panel 3 Specific Allergen; Siemens Material Number (SMN): 10385891; Catalog Number: FP3L4;
Siemens Healthcare Diagnostics
The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.
3gAllergy Specific IgE Universal Kit; Kit Component: 3gAllergy Specific IgE Adjustor Antibody (L2UNS1), 3gAllergy Specific IgE Control Antibody (L2UNS2); Siemens Material Number (SMN): 10380875; Catalog Number: L2KUN6;
Siemens Healthcare Diagnostics
The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.
IMMULITE 2000 Oak Mix Specific Allergen; Siemens Material Number (SMN): 10386105; Catalog Number: L2KUN6;
Siemens Healthcare Diagnostics
The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.