COULTER EPICS XL Flow Cytometer COULTER EPICS XL-MCL Flow Cytometer SYSTEM II Software

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    EPICS Catalog Number: EPICS CL 3 COLOR W/MCL 220V 6604985 XL 4CLR FLOW CENTRE 220V AC 6605465 PKG; 4 COLOR XL M/M (220V) 6605465R XL-MCL 4 CLR FLOW CENTRE 220V A 6605437 XL-MCL 4 CLR FLOW CENTRE 240V AC 6605438 XL-MCL 4 CLR FLOW CENTRE 220V A 6605437R EPICS XL 4 COLORS 115V 6604714 XL-MCL 4CLR FLOW CENTRE 100VAC 6605435 TN, XL Flow Cytometer 100-240V, 50/60Hz 626551 EPICS XL 3 COLOR 230V 6604726 EPICS XL 4 CLR/ MCL 115V 6604988 EPICS XL 4 CLR/ MCL 220V 6604989 XL 4CLR FLOW CENTRE 240VAC 6605466 R-EPICS XL 3 COLOR 115V 6604713 XL 4 CLR W/FLOW 1 YR WARR 6605464 PKG;4 COLOR XL M/M (100V) 6605463 EPICS XL 4 COLOR 220V 6604723 TN, XL/MCL Flow Cytometer 100-240V, 50/60Hz 626552 EPICS XL 3 COLOR 240V 6604725 EPICS XL 4 CLR/MCL 240V 6604990 4 CLR,MCL,MM(120V)REFURB 6605436R XL-MCL 4 CLR FLOWCENTRE 240VAC 6605438R 4 COLOR XL M/M (120V)REFURB 6605464R XL 4CLR FLOW CENTRE 240VAC 6605466R 4 CLR, MCL, MM PKG (240V) 6605438U XL-MCL 4 CLR W/FLOW (RECOND) 6605436U All products manufactured affected for the time period: FC-500 From 01/22/2002 to 01/31/2018 EPICS XL/XL-MCL From 11/01/1992 to 03/15/2013
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Beckman Coulter Inc.
Reason for Recall:
A manufacturing defect on the amplifier board of the Flow Cytometers my cause signal loss and/or signal drifting resulting in absence of data or a population shift in the data plots. This can impact patient results for any application.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

COULTER EPICS XL Flow Cytometer COULTER EPICS XL-MCL Flow Cytometer SYSTEM II Software

Product Codes/Lot Numbers:

EPICS Catalog Number: EPICS CL 3 COLOR W/MCL 220V 6604985 XL 4CLR FLOW CENTRE 220V AC 6605465 PKG; 4 COLOR XL M/M (220V) 6605465R XL-MCL 4 CLR FLOW CENTRE 220V A 6605437 XL-MCL 4 CLR FLOW CENTRE 240V AC 6605438 XL-MCL 4 CLR FLOW CENTRE 220V A 6605437R EPICS XL 4 COLORS 115V 6604714 XL-MCL 4CLR FLOW CENTRE 100VAC 6605435 TN, XL Flow Cytometer 100-240V, 50/60Hz 626551 EPICS XL 3 COLOR 230V 6604726 EPICS XL 4 CLR/ MCL 115V 6604988 EPICS XL 4 CLR/ MCL 220V 6604989 XL 4CLR FLOW CENTRE 240VAC 6605466 R-EPICS XL 3 COLOR 115V 6604713 XL 4 CLR W/FLOW 1 YR WARR 6605464 PKG;4 COLOR XL M/M (100V) 6605463 EPICS XL 4 COLOR 220V 6604723 TN, XL/MCL Flow Cytometer 100-240V, 50/60Hz 626552 EPICS XL 3 COLOR 240V 6604725 EPICS XL 4 CLR/MCL 240V 6604990 4 CLR,MCL,MM(120V)REFURB 6605436R XL-MCL 4 CLR FLOWCENTRE 240VAC 6605438R 4 COLOR XL M/M (120V)REFURB 6605464R XL 4CLR FLOW CENTRE 240VAC 6605466R 4 CLR, MCL, MM PKG (240V) 6605438U XL-MCL 4 CLR W/FLOW (RECOND) 6605436U All products manufactured affected for the time period: FC-500 From 01/22/2002 to 01/31/2018 EPICS XL/XL-MCL From 11/01/1992 to 03/15/2013

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0472-2019

Related Recalls

It has been determined that certain lots of Access 2 Reaction Vessels may contain manufacturing deformities. The bottom (round end) of the Reaction Vessel may have a small protrusion or hair like deformity. This could cause causing instrument errors (e.g., pipetting, rake, wash carousel errors) that could delay patient results.

Nov 7, 2025 Infusion Pumps Nationwide View Details →

Due to a software error code that crashes during acquisition on patient sample when running customer defined acquisition protocol. This may lead to delayed patient results.

Sep 22, 2025 Diagnostic Equipment Nationwide View Details →