1) Cytomics FC 500 With CXP Software BECKMAN COULTER 2) Cytomics FC 500 MPL With MXP Software BECKMAN COULTER flow cytometer for in vitro diagnostics
Class I - DangerousWhat Should You Do?
- Check if you have this product: FC 500 Catalog Num FC 500" Flow Cytometer MCL with UPS 626553 FC 500" Flow Cytometer MPL 626554 FC 500" MCL Flow Cytometer 5 Color (100V) 6605627 FC 500" MCL Flow Cytometer 5 Color (120V) 6605628 FC 500" MCL Flow Cytometer 5 Color (220V) 6605629 FC 500" MCL Flow Cytometer 5 Color (240V) 6605630 FC 500 Flow Cytometer, Remfg Assy, 5 Clr (220V) A89264 FC 500" Flow Cytometer 100-240V, 50-60HZ, Remfg Assy A89151 FC 500 Recond MCL UPS Flow Cytometer 100-240V, 50-60HZ A88262 FC 500 Flow Cytometer 100-240V A88263 All products manufactured affected for the time period: FC-500 From 01/22/2002 to 01/31/2018 EPICS XL/XL-MCL From 11/01/1992 to 03/15/2013
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Beckman Coulter Inc.
- Reason for Recall:
- A manufacturing defect on the amplifier board of the Flow Cytometers my cause signal loss and/or signal drifting resulting in absence of data or a population shift in the data plots. This can impact patient results for any application.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
1) Cytomics FC 500 With CXP Software BECKMAN COULTER 2) Cytomics FC 500 MPL With MXP Software BECKMAN COULTER flow cytometer for in vitro diagnostics
Product Codes/Lot Numbers:
FC 500 Catalog Num FC 500" Flow Cytometer MCL with UPS 626553 FC 500" Flow Cytometer MPL 626554 FC 500" MCL Flow Cytometer 5 Color (100V) 6605627 FC 500" MCL Flow Cytometer 5 Color (120V) 6605628 FC 500" MCL Flow Cytometer 5 Color (220V) 6605629 FC 500" MCL Flow Cytometer 5 Color (240V) 6605630 FC 500 Flow Cytometer, Remfg Assy, 5 Clr (220V) A89264 FC 500" Flow Cytometer 100-240V, 50-60HZ, Remfg Assy A89151 FC 500 Recond MCL UPS Flow Cytometer 100-240V, 50-60HZ A88262 FC 500 Flow Cytometer 100-240V A88263 All products manufactured affected for the time period: FC-500 From 01/22/2002 to 01/31/2018 EPICS XL/XL-MCL From 11/01/1992 to 03/15/2013
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0471-2019
Related Recalls
It has been determined that certain lots of Access 2 Reaction Vessels may contain manufacturing deformities. The bottom (round end) of the Reaction Vessel may have a small protrusion or hair like deformity. This could cause causing instrument errors (e.g., pipetting, rake, wash carousel errors) that could delay patient results.
Due to a software error code that crashes during acquisition on patient sample when running customer defined acquisition protocol. This may lead to delayed patient results.
Beckman Coulter UniCel DxI 600, Part Numbers: A71460, A30260; Photometric Chemistry Analyzer
Beckman Coulter
An internal investigation has determined that due to defective sub-components within the gantry assembly, a z-axis CAD movement motion error may occur and cause an analyzer to enter a Not Ready or Paused state, leading to a delay in obtaining results.