Affixus Hip Fracture Nail, RH 130 DEG 11MM X 400MM

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Item No. 814511400; Lot No. (Exp. Date) UDI: 507300 (05/21/2027) UDI (01)00887868039188 (17)270521 (10)507300
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Zimmer Biomet, Inc.
Reason for Recall:
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Affixus Hip Fracture Nail, RH 130 DEG 11MM X 400MM

Product Codes/Lot Numbers:

Item No. 814511400; Lot No. (Exp. Date) UDI: 507300 (05/21/2027) UDI (01)00887868039188 (17)270521 (10)507300

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0469-2018

Related Recalls

Zimmer Biomet is conducting a medical device correction to update the Instructions for Use (IFU) and the Surgical Techniques for the Polaris Spinal System Translation Screw and Cypher MIS Screw System. Users should be aware that changes were made to the contraindications section indicating that the Translation Screws should not be used with a direct current stimulation, as a high probability of construct failure exists when using Translation Screws containing CoCr with direct current simulation system.