Recall Details
- Company:
- Siemens Medical Solutions USA, Inc
- Reason for Recall:
- Possible installation error of the quench lines of superconducting magnets, used to vent helium gas in the rare event of a quench. Any restriction, obstruction, or improper installation of the quench line may cause the helium gas to be blown directly into the magnet room or other areas. This could lead to displacement of oxygen, cause cold burns, and/or pose a risk to users, patient or others.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
MAGNETOM systems, magnetic resonance diagnostic devices (MRDD).
Product Codes/Lot Numbers:
all units distributed since 1991
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0467-2016
Related Recalls
MAMMOMAT Inspiration;
Siemens Medical Solutions USA
There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.
MAMMOMAT Fusion;
Siemens Medical Solutions USA
There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.
MAMMOMAT Revelation;
Siemens Medical Solutions USA
There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.