Simplexa Direct Amplification Disc Kit, Product Kit Code MOL1455
Class I - DangerousWhat Should You Do?
- Check if you have this product: Kit Product Code MOL1455 (UDI 01-30816101025092) contains 3 discs with Disc Product Code MOL1452. Below lists the MOL1455 Kit Lot numbers (and MOL1452 Disc Lot Numbers): 12410N (12152N), 12411N (12147N), 12413N (12149N), 12477N (12150N), 12602N (11548NA), 12603N (12144N), 12721N (12148N), 12889N (12151N), 13369N (12817N), 13370N (13151N), 13434N (12382N), 13652N (12835N), and 13242N (12154N)*. *Kit Lot #13242N (Disc Lot #12154N) was shipped to Europe and not within the US.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- DiaSorin Molecular LLC
- Reason for Recall:
- A subset of affected discs are at risk of leakage when it is used more than one time with Simplexa COVID-19 Direct and/or Simplexa Flu A/B & RSV Direct Gen II. Leakage from a well that had a positive sample may cause further cross-contamination in areas of the lab if not properly cleaned.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Simplexa Direct Amplification Disc Kit, Product Kit Code MOL1455
Product Codes/Lot Numbers:
Kit Product Code MOL1455 (UDI 01-30816101025092) contains 3 discs with Disc Product Code MOL1452. Below lists the MOL1455 Kit Lot numbers (and MOL1452 Disc Lot Numbers): 12410N (12152N), 12411N (12147N), 12413N (12149N), 12477N (12150N), 12602N (11548NA), 12603N (12144N), 12721N (12148N), 12889N (12151N), 13369N (12817N), 13370N (13151N), 13434N (12382N), 13652N (12835N), and 13242N (12154N)*. *Kit Lot #13242N (Disc Lot #12154N) was shipped to Europe and not within the US.
Distribution:
Nationwide distribution
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0457-2022
Related Recalls
There is a potential for the direct amplification disc to malfunction which may result in spillage of liquid from the disc.
Simplexa COVID-19 Direct, REF: MOL4150; including Reaction Mix, REF: MOL4150; with previous assay definition v2
DiaSorin Molecular
Direct amplification Discs , used with COVID-19 and Flu A/B & RSV assays, may leak and cause contamination, cross-contamination, invalid results, and error codes.
Simplexa Flu A/B & RSV Direct Gen II, REF: MOL2655; including Reaction Mix REF: MOL2656; with previous assay definition v5.
DiaSorin Molecular
Direct amplification Discs , used with COVID-19 and Flu A/B & RSV assays, may leak and cause contamination, cross-contamination, invalid results, and error codes.