-Outer Kit Box (P/N RFIT-LBL-0140) - FilmArray¿ Meningitis/Encephalitis (ME) Panel REF RFIT-ASY-0118 30 For FilmArray systems LOT KIT UDI (01)00815381020123 CE 0086 IVD -Fitment Label (P/N RFIT-LBL-0138) which identifies the pouch label - FilmArray¿ ME Panel v1.4 LOT SN CE 0086 IVD Product Usage: The FilmArray Meningitis/Encephalitis (ME) Panel is a qualitative multiplexed nucleic acid-based in vitro diagnostic test intended for use with FilmArray systems.
Class I - DangerousWhat Should You Do?
- Check if you have this product: ME Panel Kit Lot No. ME Pouch Lot. No. Kit Expiration Date 580818 892018 4/2/2019 584418 896018 4/5/2019 Note - Kits contain a Primer Pill Lot 210318. The Primer Pill is a freeze dried/lyophilized "pellet" that contains all of the primers and buffer used in the PCR1 reaction.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- BioFire Diagnostics, LLC
- Reason for Recall:
- Firm identified an issue in the manufacturing process which might contribute to elevated rates of false positive results for HSV-1 while using the ME Panel. Only specific kits are affected.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
-Outer Kit Box (P/N RFIT-LBL-0140) - FilmArray¿ Meningitis/Encephalitis (ME) Panel REF RFIT-ASY-0118 30 For FilmArray systems LOT KIT UDI (01)00815381020123 CE 0086 IVD -Fitment Label (P/N RFIT-LBL-0138) which identifies the pouch label - FilmArray¿ ME Panel v1.4 LOT SN CE 0086 IVD Product Usage: The FilmArray Meningitis/Encephalitis (ME) Panel is a qualitative multiplexed nucleic acid-based in vitro diagnostic test intended for use with FilmArray systems.
Product Codes/Lot Numbers:
ME Panel Kit Lot No. ME Pouch Lot. No. Kit Expiration Date 580818 892018 4/2/2019 584418 896018 4/5/2019 Note - Kits contain a Primer Pill Lot 210318. The Primer Pill is a freeze dried/lyophilized "pellet" that contains all of the primers and buffer used in the PCR1 reaction.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0446-2019
Related Recalls
BioFire Joint Infection (JI) Panel REF: RFIT-ASY-0138, 30 test Kit.
BioFire Diagnostics
Contamination to in-vitro diagnostic test may result in false positives.
BioFire Respiratory Panel 2.1, IVD, REF: 423742 (30 test kit)
BioFire Diagnostics
Due to unintended movement of pouch chemistry, nucleic acid test may result in elevated risk of false negative results and risk of control failures.
BioFire Respiratory Panel 2.1 plus Kit, REF: 423740, For Filmarray systems
BioFire Diagnostics
Due to manufacturing error, respiratory panel may result in elevated control failures and/or false negative results.