Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; BREATHING KIT-LF DYNJ10550
Class I - DangerousWhat Should You Do?
- Check if you have this product: DYNJ10550 UDI-DI 40889942469704 (case), 10889942469703 (ea) Lot 25EBQ445
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Medline Industries, LP
- Reason for Recall:
- Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; BREATHING KIT-LF DYNJ10550
Product Codes/Lot Numbers:
DYNJ10550 UDI-DI 40889942469704 (case), 10889942469703 (ea) Lot 25EBQ445
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0443-2026
Related Recalls
Quick Strip Fabric Sterile Adhesive Bandages, 1" x 3", Medline Item Number NON25660QS
Medline Industries
Medline has identified that all lots of the Curad Touch-Free and Quick-Strip Bandages may exhibit open seals in the packaging which may compromise the sterility of the product.
Quick Strip Fabric Sterile Adhesive Bandages, 0.75" x 3", Medline Item Number NON25650QS
Medline Industries
Medline has identified that all lots of the Curad Touch-Free and Quick-Strip Bandages may exhibit open seals in the packaging which may compromise the sterility of the product.
Quick Strip Plastic Sterile Adhesive Bandages, 0.75" x 3", Medline Item Number NON25500QS
Medline Industries
Medline has identified that all lots of the Curad Touch-Free and Quick-Strip Bandages may exhibit open seals in the packaging which may compromise the sterility of the product.