Automated Impella Controller (AIC), product number 0042-000-US with software versions V8.4 and V8.4.1.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    UDI-DI (GTIN): 00813502010022, serial numbers IC3772, IC3773, IC5721 and IC4404.
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Abiomed, Inc.
Reason for Recall:
Pump not detected as connected to controller due to software issue.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Automated Impella Controller (AIC), product number 0042-000-US with software versions V8.4 and V8.4.1.

Product Codes/Lot Numbers:

UDI-DI (GTIN): 00813502010022, serial numbers IC3772, IC3773, IC5721 and IC4404.

Distribution:

Distributed in: US, AZ, CA, FL, NY, VA

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0440-2024

Related Recalls

Retrospective submission for following issues identified: 1. Alarm failures, power-path faults, electrical-short risks, and inaccurate motor-current sensing were identified in certain AIC consoles. 2. There is improper alignment between the purge cassette and the motor drive in the AIC. This can result in a piston block event and lead to the inability to complete priming of the purge system, requiring the user to switch to backup AIC console.

Jan 17, 2011 Other Medical Devices Nationwide View Details →

Potential that the "Purge System Blocked" alarm display on the Automated Impella Controller (AIC) could be delayed when using first generation Impella 5.5 (0550-0008*) pumps.

Feb 16, 2026 Other Medical Devices View Details →