Liberty Select Cycler Software Version 2.9.0 Model Number 180343 and RTLR180343
Class I - DangerousWhat Should You Do?
- Check if you have this product: All Liberty Select Cyclers with Software Version 2.9.0 Model Number 180343 and RTLR180343
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
β οΈ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Fresenius Medical Care Holdings, Inc.
- Reason for Recall:
- The device may detect an incorrect Heater Bag volume which may lead to a 'Supply Bag Line Blocked' alarm during treatment. This alarm may cause the Heater Bag to contain more fluid than expected and the cycler does not recognize the extra fluid is present and it will not make it available for treatment. Not having enough solution to complete treatment may lead to inadequate dialysis which could result in the accumulation of toxins and fluid. Symptoms may include shortness of breath, increased weight gain, and increased blood pressure.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- completed
Product Information
Full Description:
Liberty Select Cycler Software Version 2.9.0 Model Number 180343 and RTLR180343
Product Codes/Lot Numbers:
All Liberty Select Cyclers with Software Version 2.9.0 Model Number 180343 and RTLR180343
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0439-2021
Related Recalls
OptifluxΒΏ High Flux E-beam Dialyzer Model Numbers: (1) 0500316E OPTIFLUX F160NRe 12/CS 1.5SA EBEAM, (2) 0500318E OPTIFLUX F180NRe 12/CS 1.7SA EBEAM, (3) 0500320E OPTIFLUX F200NRe 12/CS 1.9SA EBEAM, (4) 0500325E OPTIFLUX F250NRe 12/CS 2.5SA EBEAM
Fresenius Medical Care Holdings
To provide further information to the user to ensure proper use of the medical device. the caps had been changed from a threaded version to a press-on version whereby the user must press the caps firmly onto the dialyzer to securely affix the cap.
0.9% Sodium Chloride Injection, USP, Flexible Plastic Container, Rx Only Fresenius Medical Care North America, Waltham, MA 02451, 1-800-323-5188, NDC 49230-300-10
Fresenius Medical Care Holdings
Lack of Assurance of Sterility
5008X HDF Hemodialysis Blood Tubing Set, 24 units per case, with the following models: a. 5008X Standard HDF Bloodline, Model Number: 03-5100-7. b. 5008X Standard HDF Bloodline w/CLIC, Model Number: 03-5100-7C. c. 5008X Standard HDF Bloodline w/ Twister, Model Number: 03-5150-2.
Fresenius Medical Care Holdings
Potential of blood leak occurring between the lower port of the venous chamber and the tubing include: hemodiafiltration (HDF), hemodialysis (HD), hemofiltration (HF), and isolated ultrafiltration.