1202 Flex Focus Ultrasound System, cart models UA1214 and UA1814, used with: Flex Focus 200, Flex Focus 300, Flex Focus 400, Flex Focus 500, Flex Focus 700, Flex Focus 800; System, Imaging, Pulsed Doppler, Ultrasonic

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    1) GTIN 05704916001056, Serial Numbers: 4005090, 4005710, 4002577, 4001807, 4002211, 4004085, 4006424, 4004772, 4101695, 4006420, 4100894, 4003775, 4004724, 4000321, 4004803, 4101755, 4101757, 4004635, 4004641, 4003082, 4004188, 4004448, 4100213, 4101302, 4000466, 4002982, 4006605, 4004941, 4100713, 4003569, 4003390, 4006608, 4101728, 4101906, 4005490, 4100128, 4001502, 4005922, 4006602, 4101724, 4005913, 4101394, 4001066, 4005230, 4000771, 4001536, 4003417, 4004297, 4102893, 4103104, 4003275, 4004575, 4100395, 4004038, 4004201, 4006350, 4100035, 4006123, 4100092, 4005380, 4005961, 4000460, 4005679, 4001915, 4005334, 4005346, 4001503, 4001694, 4003141, 4003626, 4004412, 4004431, 4005447, 4100060, 4005190, 4002880, 4101872, 4004088, 4004339, 4102033, 4005920, 4101823, 4007095, 4005228, 4004443, 4101990, 4002023, 4002190, 4005993, 4004423, 4101552, 4006362, 4001532, 4100603, 4005243, 4006010, 4006113, 4006114, 4006117, 4006126, 4101988, 4006950, 4006815, 4005778, 4101506, 4102036, 1909816, 4101756, 4002916, 4005906, 4101758, 4003821, 4005258, 4003139, 4101504, 4003884, 4005352, 4005769, 4005771, 4101555, 4003052, 4100463, 4007074, 4100140, 4002947, 4001137, 4006812, 4001266, 4005650, 4101956, 4003731, 4006409, 4004632, 4005337, 4100943, 4102414, 1910262, 4000166, 4101225, 4003164, 4001683, 4005372, 1910167, 4001651, 4005649, 4002186, 4101725, 4101991, 4000074, 4000845, 4001946, 4004994, 4007036, 4007038, 4007039, 4007040, 4007042, 4007053, 4100040, 4100057, 4100068, 4100069, 4100070, 4100127, 4100194, 4100819, 4100822, 4100839, 4100870, 4100873, 4101526, 4101554, 4101632, 4101646, 4101677, 4101803, 4101828, 4101836, 4101863, 4101867, 4101983, 4102012, 4102015, 4102105, 4005729, 4102419, 4100133, 4005972, 4102328, 4102417, 4004344, 1910010, 4004855, 4006810, 4003561, 4005461, 4101634, 4100330, 4006425, 4101099, 4004192, 4005257, 4006943, 1907405, 1907367, 4002579, 4100225, 4101206, 4005868, 4005872, 4102993, 4101253, 4001126, 4005184, 4003763, 4001414, 4001692, 4001702, 4005128, 4101117, 4101958, 4101538, 4005496, 4001534, 4001656, 4101349, 4001996, 4002470, 4005442, 4005928, 4002588, 4005969, 4001421, 4004649, 4101537, 4006418, 4006809, 4002276, 4100887, 4001389, 4100995, 4101180, 4102121, 4003545, 4006998, 4102575, 4001707, 4004885, 4006796, 4006404, 4102416, 4006997, 4006941, 4101826, 4101114, 4004766, 4002422, 4004696, 4007085, 4003532, 4101804, 4005735, 4003770, 4101606, 4003229, 4001132, 4001133, 4001136, 4001448, 4003059, 4003885, 4004343, 4004567, 4005936, 4100744, 4101601, 4003000, 4002272, 4002467, 4005797, 4003682, 4004789, 4001499, 4004006, 4004426, 4005485, 4005487, 4000782, 4003768, 4101249, 4101304, 1907823, 1908864, 4003525, 4003539, 4004671, 4101400, 4101401, 4101403, 4101407, 4003391, 4101228, 4005945, 4005793, 4101722, 4005680, 4002448, 4000145, 4000889, 4000421, 4002995, 4002839, 4004413, 4001680, 4003542, 4005977, 4005341, 4006624, 4006879, 4005169, 4100533, 4100600, 4005504, 4100679, 4005998, 4003772, 4005168, 4005383, 4101404, 4101955, 4005373, 4100096, 4101156, 4005967, 4101245, 4100480, 4004232, 4006924, 4101132, 4005773, 4007097, 4100761, 4101398, 4005351, 4004628, 4004658, 4004999, 4005335, 4101247, 4004889, 4005444, 4004725, 4007049, 4005088, 4005344, 4003056, 4100411, 4004042, 4100062, 4100454, 4003162, 4004226, 4000966, 4100390, 4100210, 4003415, 4003457, 4003879, 4004385, 4005935, 4007007, 4006486, 4004467, 4005384, 4101270, 4004875, 4004010, 4101419, 4102169, 4002921, 4002948, 4004436, 1910149, 4006308, 4000841, 4001531, 4003528, 4003198, 4003625, 4005436, 4003160, 4003684, 4006801, 4100086, 4101396, 4004828, 4006874, 4004422, 4000994, 4006329, 4006987, 4005249, 4100550, 4005369, 4005092, 4005229, 4005642, 4005879, 4101361, 4002277, 4003054, 4004618, 4005724, 4005852, 4005934, 4003225, 4006932, 4006988, 4001381, 4001415, 4005185, 4000136, 4000327, 4000461, 4000464, 4000474, 4004698, 4001686, 4004231, 4001422, 4005507, 4101183, 4101764, 4101961, 4101967, 4002352, 4005751, 4002069, 4002625, 4101701, 4100232, 4005086, 4006333, 4004656, 4005743, 4000871, 4000882, 4004598, 4101687, 4001690, 4001992, 4002020, 4002021, 4002078, 4005374, 4100546, 4004388, 4007072, 4005027, 4001805, 4002729, 4003140, 4001418, 4003017, 4004781, 4001497, 4101769, 4002398, 4007058, 4004338, 4002575, 4001498, 4004393, 4101638, 4005661, 4005942, 4100432, 4000797, 4000798, 4102801, 1910018, 4000324, 4001410, 4003025, 4003881, 4004868, 4001920, 4003946, 4004225, 4100151, 4100145, 4002183, 4002214, 4003282, 4003503, 4004590, 4006330, 4005981, 4007035, 4007052, 4007059, 4101364, 4004830, 4006355, 4100910, 4101248, 4003565, 4006876, 4006621, 4007099, 4100936, 4001409, 4101834, 4003316, 4004002, 4002220, 4003317, 4004565, 4101409, 4002507, 4003512, 4004793, 4002217, 4003081, 4100973, 4001924, 4002209, 4005235, 4005840, 4100216, 4101408, 4004390, 4004403, 4002846, 4005924, 4004729, 4100906, 4100911, 4100930, 4000185, 4000323, 4001135, 1911127, 4000143, 4000141, 4000164, 4001494, 4001687, 4003521, 4101532, 4101729, 4100129, 4103008, 4005338, 4004715, 1910260, 4001809, 4002587, 4101809, 4001971, 4007106, 4004425, 4005957, 4100147, 4100963, 4003452, 4006934, 4102640, 4002585, 4100722, 4005720, 4005727, 4005038, 4003527, 4101005, 4003451, 4005897, 4101023, 4102951, 4005984, 4003226, 4000465, 4002808, 4003318, 4004831, 4004838, 4005345, 4102534, 4000325, 4004083, 4005678, 4002401, 4005651, 4007103, 4100503, 4102497, 4005170, 4006598, 4000877, 4000883, 4006415, 4003079, 4002915, 4002953, 4006000, 4001263, 4001700, 4003138, 4100971, 4003002, 4004432, 4006599, 4101970, 1906178, 4005239, 4001649, 4004633, 4003458, 4003501, 4002514, 4004643, 4004926, 4007062, 4004773, 4005033, 4101169, 4004761, 4005672, 4004988, 4005261, 4004708, 4002815, 4005952, 4004642, 4004826, 4005974, 4006410, 4006858, 4006949, 4100196, 4100904, 4003478, 4004987, 4001412, 4005890, 4007079, 4005386, 4005767, 4000775, 4001802, 4006975, 4005902, 4003553, 4004454, 4005030, 4004592, 4006335, 4002170, 4101226, 4005668, 4100202, 4103007, 4002583, 4002958, 4003547, 4004700, 4005850, 4004886, 4001416, 4100056, 4100061, 4006004, 4004931, 4005665, 4100743, 4102826, 4001387, 4100705, 4002913, 4102052, 4004637, 4100714, 4003114, 4003880, 4001078, 4005503, 4001506, 1910254, 4004400, 4004861, 4006596, 4101417, 4101915, 4006407, 4002491, 4004404, 4004820, 4002540, 1910141, 4004424, 4101630, 4101685, 4101686, 4005831, 4007076, 4006121, 4102276, 4004697, 4002248, 4005791, 4005846, 4102535, 4002275, 4004441, 4101989, 4100440, 4005028, 4102468, 4006011, 4006811, 4004572, 4003766, 4004286, 4100862, 4005907, 1908876, 4004701, 4100939, 4102835, 4101266, 4100212, 4006305, 4005459, 4102519, 4002763, 4101754, 4002949, 4001407, 4001420, 4001508, 4003511, 4003938, 4002806, 4006937, 4100075, 4100076, 4100080, 4000783, 4000416, 4002442, 4003781, 4003277, 4003389, 4005884, 4005889, 4100054, 4005163, 1900917, 1900975, 1901001, 1906172, 1906174, 1906176, 1907401, 1907466, 1907497, 1908878, 1908968, 1910014, 1910139, 1910143, 1910145, 1910173, 1910221, 1910264, 1911129, 1911838, 4000068, 4000417, 4000419, 4000469, 4000773, 4000776, 4000780, 4000840, 4000878, 4000888, 4000967, 4000970, 4001067, 4001071, 4001139, 4001192, 4001383, 4001419, 4001449, 4001451, 4001495, 4001509, 4001512, 4001514, 4001515, 4001533, 4001540, 4001542, 4001550, 4001636, 4001638, 4001642, 4001646, 4001684, 4001685, 4001689, 4001691, 4001695, 4001703, 4001705, 4001839, 4001841, 4001916, 4002024, 4002025, 4002168, 4002169, 4002189, 4002244, 4002246, 4002306, 4002307, 4002399, 4002419, 4002420, 4002421, 4002423, 4002426, 4002444, 4002446, 4002468, 4002484, 4002485, 4002617, 4002618, 4002622, 4002623, 4002627, 4002725, 4002765, 4003045, 4003053, 4003186, 4003349, 4003767, 4003935, 4003943, 4004033, 4004415, 4004438, 4004458, 4004460, 4004471, 4004472, 4004694, 4004709, 4004727, 4004758, 4004867, 4004990, 4004998, 4005034, 4005245, 4005776, 4005801, 4005838, 4005882, 4006614, 4006978, 4100150, 4100199, 4100490, 4100724, 4100892, 4100912, 4100937, 4100942, 4101001, 4101087, 4101105, 4000167, 4005975, 4000186, 4003537, 4005886, 4001537, 4004730, 4002769, 4005914, 4005443, 4004864, 4101131, 4000993, 4005437, 4000996, 4001997, 4003084, 4003554, 4000774, 4000190, 4004940, 4001496, 1909812, 4003020, 4002375, 4002513, 1910012, 4000969, 4002449, 4002985, 4000847, 4002215, 4007073, 4102059, 4001944, 4001706, 4005342, 1907464, 4001643, 4001681, 4005768, 4007051, 4101137, 4002877, 4004768, 4100064, 4100895, 4001966, 4001191, 4004337, 4004340, 4004943, 4000793, 1910020, 4003522, 4004778, 4101531, 4006823, 4001510, 4100422, 4100433, 4100434, 1906625, 1910016, 4000183, 4002586, 4000153, 4005750, 4003497, 4003538, 4005871, 1907399, 4004576, 4004421, 4004429, 4004837, 4005777, 4004427, 4002539, 4004411, 4101533, 4003729, 4004616, 4000876, 4000884, 4000887, 4006983, 4002578, 4003727, 4004473, 4004832, 4005870, 4004228, 4006592, 4004854, 4006918, 4006935, 4101148, 4003723, 4003596, 4005186, 4005222, 4004439, 4005653, 4005343, 4100041, 4100055, 4100059, 4007066, 1900918, 4002998, 4003117, 4003158, 4005445, 4101684, 4002997, 4003817, 4006929, 4004435, 4003044, 4004634, 4005985, 4003877, 4100207, 4003019, 4002167, 4005867, 4002187, 4005044, 4102922, 4101731, 4102016, 4003496, 4003510, 4005237, 4005786, 4000891, 4004196, 4005040, 4004447, 4102996, 4102208, 4100215, 4005669, 4006854, 4100421, 1911806, 4004457, 4003486, 4005733, 4003878, 4005893, 4003146, 4101466, 4007080, 4005189, 4003426, 4100451, 4100913, 4001076, 4003309, 4005452, 4005456, 4002274, 4006871, 4003468, 4004657, 4004942, 4102553, 4005648, 4001417, 4004928, 4000843, 4001998, 4006124, 4101639, 4005997, 4002210, 4100131, 4004937, 4001635, 4003189, 4100444, 4001075, 4001513, 4003592, 4006798, 4005045, 4000890, 4004434, 4003514, 4005746, 4100135, 4005742, 4000880, 4102055, 1910147, 4003057, 4004780, 4002354, 4004332, 4101101, 4102273, 4102627, 4102950, 4005387, 4100420, 4006492, 4006016, 4005811, 4003815, 4102420, 4003479, 4003314, 4005382, 4005491, 4005640, 4102277, 1910314, 4000995, 4002212, 4101711, 4001493, 4003550, 4004112, 4005671, 4003477, 4003494, 4003500, 4003505, 4003507, 4003513, 4003628, 4003508, 4005043, 4000965, 4000971, 4003042, 4004036, 4004626, 4005677, 4005847, 4005851, 4005853, 4005876, 4005895, 4005903, 4005921, 4005980, 4004445, 4004446, 4004874, 4001202, 4000322, 4000471, 4005973, 4101176, 4001388, 4101136, 4001072, 4001134, 4000873, 4000886, 4001803, 4002185, 4003193, 4005259, 4100861, 4003425, 4003232, 4005161, 4005162, 4002402, 4002469, 1908972, 4006309, 4004342, 4004005, 4000184, 4004849, 4101627, 1910256, 4002213, 4002728, 4003732, 4005134, 4005918, 4102994, 4004651, 4004653, 4004189, 4003046, 4004813, 4004927, 4005932, 4000998, 4006483, 4101529, 4102722, 4000781, 4002488, 4003049, 4006411, 4101631, 4102915, 4005670, 4002400, 4101002, 4006795, 4006001, 4002630, 4100832, 4006436, 4102800, 4003816, 4004851, 4003568, 4004190, 4004227, 4004236, 4004239, 1911125, 4001704, 4000336, 4005859, 4000328, 4002954, 4003219, 4101700, 4005233, 4003546, 4003780, 4005856, 4101909, 4005195, 4101106, 4005505, 4001507, 4102948, 4101912, 4001069, 4002245, 4101391, 4003148, 4005250, 4006594, 4005130, 4004865, 4001425, 4003558, 4002994, 4002301, 4004624, 4003474, 1911808, 4102762, 4003524, 4000422, 4001505, 4100450, 4102122, 4004341, 4101692, 4001838, 4005888, 4005908, 4006412, 1911945, 4102829, 4102758, 4002241, 4101705, 4101135, 4002986, 4003041, 4000470, 4000997, 4002758, 4002766, 4101202, 4101000, 4002222, 4100091, 4100711, 4006364, 4102018, 4001068, 4004597, 4101273, 4005784, 4001538, 4001413, 4101762, 4101474, 4003543, 4101205, 4000334, 4004193, 4101760, 4001065, 4004704, 4001535, 4003629, 4101640, 4000188, 4000413, 1893197, 4000842, 4004420, 4001640, 4102058, 4004282, 4101020, 4102013, 1910258, 4000333, 4002490, 4001074, 4005774, 4002184, 4002302, 4002304, 4002305, 4004084, 4004086, 4102314, 4100972, 4002920, 4102546, 4006872, 4005986, 4004629, 4101246, 4000972, 4003534, 4100482, 4002353, 4003383, 4003587, 4101177, 4101643, 4005891, 4001188, 4005644, 4101868, 4001644, 1910223, 4005877, 4003776, 4004862, 1906237, 1910165, 4002219, 4101021, 4005652, 4100387, 4005660, 4101353, 4000418, 4001408, 4003108, 4101062, 4004081, 4004117, 4100762, 4100063, 4102119, 4101219, 4101807, 4101822, 4003536, 4001073, 4006307, 1908872, 1908974, 1908976, 4001122, 4001511, 1910163, 4001639, 4100709, 4003490, 4006944, 4000849, 4001447, 4001450, 4005492, 4003024, 4003633, 4003686, 4002489, 4002767, 4001941, 4101959, 4005381, 4005449, 4005645, 4101024, 4006866, 4003818, 4004078, 4006797, 1907354, 1907356, 4001190, 4002814, 4003562, 1907358, 1907468, 4000875, 4000881, 4005681, 4005711, 4006817, 4002404, 4004295, 4005241, 4005252, 4102760, 4007056, 4102211, 4005880, 4002811, 4006989, 4006931, 4001801, 4004200, 4005929, 4100095, 4100753, 4003043, 4004695, 4005133, 4005135, 4005139, 4005140, 4006110, 4004866, 4001658, 4005770, 4100038, 4007070, 4005196, 4003320, 4007005, 1909814, 4003601, 4002182, 4003717, 4005855, 4005857, 4102206, 4007108, 4100462, 4006800, 4003283, 4005912, 1910215, 4001693, 4002218, 4000879, 4006863, 4003544, 4005901, 4004717, 4000472, 4000457, 4000423, 4006015, 4007105, 4002347, 4101766, 4101768, 4004844, 4005833, 4000872, 4002221, 4004863, 4102053, 4101172, 4102477, 4002999, 4001701, 4005937, 4100493, 4100534, 4100549, 4102467, 4006882, 4101178, 4004589, 4004595, 4005231, 4000152, 4100126, 4102061, 4003557, 4003190, 4100418, 4001647, 4003143, 4102925, 4004936, 4100750, 4101471, 4101707, 4001942, 4001969, 4001994, 4003469, 4100893, 4003556, 4002019, 4001641, 4001917, 4004869, 4005446, 4101362, 4101835, 4101840, 4102478, 4101802, 4005979, 4102281, 4005647, 4005739, 4005869, 4100583, 4101251, 4101986, 4000160, 4006857, 4006859, 4006922, 4007045, 4004335, 4100708, 4101704, 4101881, 4003506, 4006849, 4100415, 4101808, 4101224, 4003111, 4003594, 4004197, 4004938, 4101904, 4100924, 4003947, 4005834, 4005841, 4005842, 4006792, 4100882, 4002377, 4002068, 4002807, 4001411, 4001099, 4102123, 4102164, 4102165, 4102166, 4102571, 4102572, 4102580, 4103014, 4101602, 4005885, 4005892, 4005909, 4100193, 4102168, 4006313, 4006413, 4007047, 4101535, 4001500, 4001799, 4005740, 4004459, 4100058, 4000885, 4100142, 4100071, 4100097, 4003690, 4101181, 4101726, 4005187, 4004706, 4003351, 4006422, 4004087, 4003107, 4005191, 4002447, 4000473, 4005788, 4006802, 4006005, 4101920, 4101954, 4003720, 4004721, 4004835, 4003145, 4006613, 4005754, 4005506, 4101763, 4006435, 4102619, 1908970, 4000420, 4003942, 4100442, 4004588, 4006012, 4005171, 4004719, 4002803, 4005988, 4002242, 4003689, 4004663, 4004666, 4004842, 4004843, 4005728, 4006917, 4101509, 4101301, 4004659, 4004871, 4005248, 4102836, 4102837, 4101683, 4006818, 4006007, 4002546, 1907462, 4001382, 4100067, 4004007, 4100130, 4001539, 4001682, 4003196, 4006816, 1910171, 4000772, 4100416, 4001386, 4101357, 4100227, 1907763, 1911804, 4002950, 4102056, 4001501, 4003883, 4004428, 4101551, 4002984, 4100034, 4100907, 4100908, 4101702, 4000796, 4002849, 4101917, 4002181, 4002188, 4005732, 4004818, 4101642, 4101678, 4101682, 4101696, 4004710, 4006948, 4000784, 4006936, 4001189, 4000468, 4101174, 4100886, 4100066, 4002022, 4003286, 4003355, 4003472, 4003492, 4004014, 4004989, 4006916, 4007087, 4004564, 4006861, 4002067, 4006426, 4101806, 4102401, 4000462, 4101184, 4001504, 4100678, 4102867, 4005667, 4102621, 4102601, 4001688, 4000964, 4002850, 4002883, 4005982, 4004437, 4100994, 4002465, 4005765, 4101387, 4000844, 4102914, 4005795, 4006125, 4005789, 4004877, 4005333, 4001648, 4001991, 4100484, 4006109, 4005878, 4002216, 4007102, 4001128, 4001652, 4100391, 4003083, 4003455, 4005976, 4102282, 1911835, 4003387, 4005943, 4102628, 4102312, 4005955, 4100214, 4100707, 4003630, 4002576, 4005676, 4100929, 4100932, 4101864, 4000138, 4004883, 4002076, 4002875, 4000139, 4000415, 4100997, 4101019, 4000071, 4004720, 4004722, 4005927, 4101827, 4002547, 4002847, 4004673, 4004850, 4005347, 4006942, 4001970, 4006416, 4004408, 1908874, 4005990, 4006867, 4002247, 4002957, 4003889, 4004290, 4004294, 4004296, 4102514, 4005132, 4002443, 4005499, 4005787, 4100999, 4006311, 4005440, 4101680, 4004723, 4101597, 4001426, 4000777, 4100200, 4003192, 4100696, 4100530, 4002956, 4102213, 4001530, 4000414, 4000424, 4000463, 4004631, 4006814, 4000846, 4000848, 4001000, 4004630, 4005781, 4100884, 4101916, 4101865, 4101905, 4001650, 4101694, 4100230, 4100322, 4006869, 4003106, 4004804, 4100721, 4101227, 4005643, 4006013, 4101691, 4001077, 4004090, 4102971, 4003599, 4007050, 4004334, 4001541, 4100088, 4100094, 4102469, 4004620, 4100960, 4001544, 4005046, 4102828, 4005775, 4102209, 4005849, 4002809, 4003109, 4003713, 4002466, 4001637, 4101059, 4101060, 4003473, 4004416, 4000968, 4102543, 4006366, 4004619, 4102019, 4000182, 4002805, 4000467, 4005992, 4001843, 1908962, 4003222, 4005968, 4000412, 4101100, 4101679, 4005717, 4003022, 4101346, 4006590, 4005253, 4101553, 4004235, 4003447, 4100224, 4003485, 4002486, 4001808, 4103041, 4004079, 4101833, 4005953, 4101170, 4001204, 4003161, 4005026, 4001645, 4003631, 4004430, 4002511, 4001360, 4003730, 4005225, 4006921, 4003523, 4005709, 4006947, 4002887, 4004089, 4000410, 4000779, 4005887, 4005712, 4001840, 4005389, 4006331, 4101351, 4004399, 4100090, 4000778, 4000800, 4003378, 4000999, 4007089, 4003502, 1910022, 4006111, 4004469, 4003003, 4000187, 4004935, 4005032, 4005766, 4001804, 4001844, 4005232, 4002271, 4002308, 4005141, 4003080, 4001424, 4001919, 4102714, 4002756, 4001965, 1906627, 4004665, 4006628, 4005755, 1908870, 4007075, 4002463, 4006414, 4002988, 4004827, 4004417, 4003540, 4003673, 4004410, 4004829, 4001800, 4004888, 4005719, 4000794, 4003163, 4006595, 4100760, 4100763, 4101500, 4101501, 4003552, 4002582, 4003493, 4005348, 4006820, 4101984, 4002841, 4002851, 4003321, 4004622, 4101392, 4004991, 1910219, 4005782, 4001914, 4006847, 4003888, 4000189, 4003448, 4006491, 4006940, 4004409, 4002376, 4100087, 4004782, 4005721, 4003142, 4004716, 4100890, 4102622, 4101810, 4002619, 4006427, 4004578, 4002955, 4005899, 4005495, 4002509, 4101355, 4000458, 4006915, 4006928, 4004770, 1911734, 4000459, 4003424, 4000411, 4002876, 4002842, 1908858, 4000069, 4000073, 4001806, 4001810, 4001913, 4101003, 4005744, 4006597, 4102120, 4101130, 4101144, 4101147, 4101203, 4102866, 4005438, 4006118, 4006850, 4100712, 4100481, 4101155, 4002445, 4100053, 4005388, 4002510, 4005336, 4001131, 4002487, 4005916, 4103017, 4101418, 4005675, 4000623, 4000628, 4001466, 4001585, 4001586, 4004449, 4005138, 4005509, 4005641, 4005654, 4005655, 4005657, 4005658, 4005659, 4005663, 4101271, 1902581, 1908741, 1908745, 1910137, 4000149, 4000209, 4000442, 4000827, 4001088, 4001095, 4001112, 4001130, 4001490, 4002508, 4003279, 4003352, 4003384, 4003517, 4004718, 4005193, 4005484, 4006354, 4007082, 4100204, 4102913, 4004779, 4100560, 4004462, 4100617, 1907495, 4002071, 4101460, 4007096, 4005451, 4004786, 4001618, 4002629, 4004762, 4005779, 4101534, 4101708, 4102057, 4004794, 4006349, 4002843, 4002380, 4003344, 4003388, 4004035, 4101690, 4101698, 4101699, 4001778, 4004796, 4102710, 4101186, 4004731, 4003449, 4004039, 4004115, 4004887, 4006003, 4002329, 4102579, 4003312, 4005803, 4006484, 4100734, 4102577, 5102579, 4003116, 4102285, 4002700, 4003014, 4004032, 4004468, 4004569, 4004879, 4005832, 4006487, 4006926, 4101395, 4102021, 4102620, 4102626, 4102769, 4102772, 4006116, 5006892, 4004878, 4102708, 4100881, 4101057, 4005948, 4101875, 4101902, 4100206, 4006923, 4102768, 4005486, 4006119, 4100079, 4100072, 1908651, 4002541, 4006008, 4101710, 4002952, 4005458, 4000678, 4002126, 4102752, 4100654, 4004037, 4102470, 4005260, 4003627, 4003679, 4004996, 1909090, 4005854, 4001587, 4002551, 4002657, 4004986, 4100329, 4101645, 1910962, 4000144, 4001527, 4101347, 4101918, 4005673, 4101811, 4000575, 4001529, 4003086, 4003721, 4100786, 4102272, 4004674, 4004714, 4005736, 4002131, 4001206, 1907441, 4004636, 4004639, 4100697, 4003563, 4004644, 4100139, 4100198, 4100491, 4101116, 4101129, 4005350, 4101244, 4101104, 4101505, 4101058, 4101061, 4002727, 4001624, 4005845, 4001249, 4100461, 4101689, 4100964, 4101709, 4000162, 4005966, 4100821, 4101413, 4102311, 4102405, 4102639, 1910150, 4003004, 4102325, 4003535, 4006979, 4001380, 4001918, 4006620, 4002303, 4003021, 4003559, 4003598, 4005874, 4100093, 4001423, 4101841, 4005898, 4002584, 4004800, 4002278, 4000799, 4003782, 4000874, 4000785, 4000795, 4006803, 4002378, 4002381, 1911732, 4001993, 4003691, 4102492, 4003564, 4003886, 4100231, 4002382, 4100885, 4002989, 4001193, 4004407, 4004771, 4006981, 4000870, 4004841, 4006853, 4004405, 4002542, 4100602, 4005848, 4006808, 4101604, 4001842, 4006920, 4101557, 4004401, 4102315, 4004040, 4005910, 4005662, 4005875, 4004001, 4101960, 4004699, 4006619, 4000181, 4006403, 4101644, 4100125, 4002243, 4001070, 4001203, 1911039, 4003944, 4004233, 4004930, 4003091, 4003820, 4101525, 4006419, 4002620, 4003147, 4100834, 4006881, 4002512, 4103042, 4100065, 4101969, 4005036, 4002768, 4001543, 4004617, 4101252, 4003227, 4001780, 4003427, 4003462, 4004669, 4004819, 4004934, 4005860, 4005896, 4005958, 4006865, 4100601, 4100664, 4100751, 4101388, 4101636, 4102020, 4102204, 4102555, 4102618, 4103106, 2002566, 2002667, 2002668, 4000330, 4000331, 4000335, 4002330, 4002334, 4002654, 4002658, 4002659, 4002660, 4002661, 4002663, 4002664, 4002688, 4002689, 4002690, 4002692, 4002693, 4004623, 4004662, 4004853, 4004860, 4004870, 4004873, 4005143, 4005183, 4005234, 4005240, 4005790, 4005836, 4005959, 4006607, 4006875, 4007093, 4007094, 4100223, 4100388, 4100389, 4100392, 4100394, 4100419, 4100580, 4100776, 4100842, 4101107, 4101268, 4101354, 4101530, 4101556, 4101869, 4101914, 4101965, 4101987, 4102493, 4102547, 4102624, 4102625, 4102721, 4102784, 4102787, 4102827, 4102830, 4102911, 4102912, 4102921, 4103004, 4103005, 4103028, 4103029, 4103030, 4103031, 4103032, 4103043, 4103044, 4103056, 4103092, 4103093, 4103094, 4103095, 4103096, 4103103, 4103105, 4103130, 4103134, 4103149, 4103151, 4103152, 4000441, 4001491, 1911677, 4000836, 4004792, 4007006, 4101913, 4102317, 1893118, 4006880, 4102600, 4102851, 4102852, 4000433, 4001486, 4001594, 4003112, 4003276, 4005810, 4006310, 4101761, 1901888, 4003159, 1907817, 4003088, 4000838, 1902575, 1907547, 1911572, 1911710, 4000208, 4000435, 4000573, 4002332, 4002762, 4002804, 4003589, 4004013, 4002731, 4001945, 1907761, 4000212, 4000451, 4001119, 4001590, 4001847, 4005656, 4005808, 4100149, 4101953, 1911074, 1908239, 1908241, 1908243, 1908245, 4000443, 4003421, 4101963, 4001118, 4005931, 4006433, 4001922, 4002066, 4004077, 4004383, 4005453, 4005783, 4005796, 4006431, 4006623, 4102785, 4102786, 4103045, 4003228, 1911947, 4001967, 4002166, 4002987, 4005166, 4100083, 4100563, 4102498, 4100528, 4004110, 4004387, 4004783, 4005242, 1904372, 4000213, 4004453, 4101503, 4101507, 4101527, 4102797, 4102967, 4000021, 1908747, 4001264, 4007037, 4005930, 1904324, 4001657, 4003423, 4005129, 4003453, 4006406, 4007098, 4006312, 4005873, 1909803, 1905149, 1907759, 4001525, 4001619, 4001622, 4002624, 4002848, 4002878, 4003055, 4003060, 4003089, 4004463, 4004579, 4004591, 4004596, 4006603, 4100831, 4100840, 4100843, 4004389, 4101399, 4101464, 4101465, 4101629, 4001655, 4002333, 4006609, 1911708, 4100529, 4002162, 4000828, 4000829, 4000832, 4000833, 4000834, 4000835, 4000837, 4003156, 4003187, 4003188, 4003191, 4003197, 4003224, 4004456, 4102578, 4102587, 4102588, 4102596, 4102597, 4102598, 4102599, 4000072, 4001522, 4102279, 4007060, 4001262, 1911730, 4003416, 4006938, 4000577, 4000582, 4001850, 4003381, 4005488, 4102715, 4005753, 4001113, 4001116, 4001100, 4004845, 4004852, 4001105, 4002157, 4002701, 4001106, 4000923, 4002881, 4002885, 4003350, 4100531, 1911142, 4001087, 4003220, 4003675, 4003685, 4005188, 4005798, 4005951, 4006334, 4100962, 4005730, 4000345, 4002074, 4006990, 4101968, 4101971, 4000576, 4000731, 4001248, 4001261, 4001355, 4001626, 4002133, 4002655, 4002734, 4003157, 4003450, 4003567, 4003570, 4004291, 4004293, 4004655, 4004777, 4004798, 4004812, 4004815, 4005737, 4005747, 4005785, 4005991, 4006002, 4006360, 4006793, 4007061, 4100141, 4100967, 4101171, 4101173, 4101599, 4101603, 4101605, 4101878, 4102923, 4100863, 4100925, 4102892, 4101982, 4000156, 4000340, 4000353, 4000446, 4000729, 4000767, 4001008, 4001091, 4001094, 4001125, 4001357, 4002158, 4002163, 4002165, 4002581, 4002879, 4003047, 4003155, 4003221, 4003274, 4003284, 4003518, 4003531, 4003595, 4003774, 4004009, 4004030, 4004109, 4004111, 4004118, 4004707, 4004808, 4004823, 4004846, 4004985, 4005025, 4005039, 4005224, 4005349, 4005455, 4005494, 4005674, 4005682, 4005722, 4005738, 4005983, 4006315, 4006327, 4006352, 4006434, 4006437, 4006604, 4006984, 4007048, 4007081, 4100134, 4101345, 4100935, 4100085, 1900915, 1900922, 4102327, 4102408, 4101502, 4101765, 4101681, 4100417, 4101598, 4102891, 4006428, 4101550, 4003345, 1911840, 4005807, 4101187, 4003483, 4000452, 4102167, 4004834, 4001251, 4102207, 4101469, 4001557, 4003310, 4007055, 4004880, 4001250, 4001104, 4003945, 4001849, 4005664, 4006873, 4100996, 4003420, 4004774, 4101461, 4006353, 4004785, 4001741, 4006593, 4007083, 4100485, 4005858, 4102309, 4000427, 4005244, 4007092, 4000350, 4003287, 4101352, 4001560, 4005946, 1907863, 4001035, 4003603, 4005708, 4004932, 4005087, 1907499, 1907939, 4005096, 4001887, 4100732, 4101831, 4000211, 4002665, 4100733, 4102323, 4102712, 4001114, 4005919, 4005900, 4000036, 4006014, 4004465, 4101838, 4100651, 4100653, 4004229, 4006616, 4001561, 4005167, 4101467, 4100039, 4001563, 4006615, 4006819, 4002812, 4004939, 4003354, 4005221, 4006982, 4006994, 4100073, 4101027, 4003499, 1907568, 4100138, 4102014, 4004280, 4102643, 4000428, 4000578, 4004765, 4102759, 4000921, 4006996, 4005726, 4000348, 4001653, 4004464, 4005037, 4002548, 4102413, 4006627, 4101919, 4102054, 4001103, 4004230, 4004562, 4001554, 4101175, 4102641, 4102949, 4004455, 4102124, 4102163, 4004289, 4100412, 4001385, 4101220, 1907510, 4005262, 4005866, 4100009, 4102017, 4005925, 4002730, 4002874, 4002910, 4002914, 4004857, 4001565, 4005371, 4101877, 4005749, 4100809, 4000432, 4102170, 4102418, 4101688, 4102542, 4006804, 4000922, 4006306, 4001771, 4007008, 4004577, 4101528, 4000682, 4000343, 4005001, 4100810, 4001909, 4003504, 4005031, 4101871, 4005805, 1907873, 4005865, 4100938, 4100941, 4101386, 4101908, 4102713, 4005989, 4100513, 4001846, 4004847, 4003048, 4100998, 4005460, 4001775, 4002549, 4003520, 4004191, 4006860, 4101004, 4101730, 4001857, 4000158, 4000448, 4101727, 4100812, 4005666, 4002125, 4000625, 4100492, 4000455, 4001562, 4100779, 4102115, 4100511, 4005794, 4001555, 4100754, 4003012, 4003634, 4005500, 4003515, 4100735, 4001583, 4100710, 4000352, 4000437, 4100752, 4004807, 4005236, 4004451, 4007063, 4000631, 4001255, 1911712, 4000627, 4003010, 4004806, 4101825, 4001526, 4000728, 4100195, 4002922, 4006856, 4001256, 4005256, 4005378, 4004029, 4100494, 4101829, 4001092, 4001588, 4005095, 4000355, 4000168, 4000429, 4001584, 4100545, 4004627, 4005646, 1907827, 4101085, 4000677, 4003050, 4100787, 4101363, 4001552, 4000159, 4001006, 4102663, 4004621, 4005837, 4002543, 4003113, 1911523, 4002102, 4003418, 4004858, 4100084, 4101697, 4101390, 4102642, 4102286, 4102313, 4003347, 4004764, 4102284, 4000679, 4101414, 4101539, 4004814, 4001551, 4003319, 4102034, 4100833, 4101637, 4101693, 4101874, 4101966, 4102283, 4003488, 4006017, 4101264, 4006852, 4100745, 4100966, 4100500, 4101222, 4100970, 4004625, 4002993, 4101907, 4001123, 4102831, 4003385, 4100205, 4001740, 4100208, 4000912, 4001558, 4002759, 4000732, 4001252, 4003716, 4000730, 4005093, 4003728, 4102511, 4102969, 4003769, 4001115, 4001097, 1911679, 4000681, 4004461, 4006108, 4006115, 4100789, 4002917, 4003110, 4005904, 4003602, 4003941, 4001093, 4001102, 4001257, 4003590, 4100561, 4003475, 4000347, 4000425, 5004302, 4005716, 4004787, 4100453, 4004797, 4000456, 4007003, 4102924, 4002544, 4101406, 4102403, 4102406, 4000676, 4001358, 4006878, 4100841, 4002100, 4004107, 4102518, 4000351, 4002128, 4102409, 4102751, 4002694, 4006611, 4000444, 4005996, 4001742, 4001911, 4100211, 4100222, 4100229, 4100515, 4100526, 4100672, 4101463, 4101830, 4101879, 4101910, 4102278, 4102320, 4001773, 4002101, 4002273, 4004237, 4101753, 4101759, 4102723, 4002355, 4001743, 4003484, 4004795, 4004790, 4002845, 4002405, 4004732, 4005137, 4004433, 4004563, 4001770, 4003692, 4000620, 4002073, 4004840, 4001852, 4002357, 4003676, 4005741, 4003724, 4004573, 4006348, 4005192, 4001556, 4002425, 4003223, 4003144, 4005964, 4000734, 4002918, 4003115, 4000769, 4001774, 4000916, 4006405, 4000579, 4001553, 4005377, 4000622, 4000629, 4003087, 4003948, 4004654, 4100144, 4100148, 4100582, 4100589, 4100590, 4100615, 4100616, 4100618, 4100619, 4100620, 4100621, 4100627, 4100629, 4100630, 4100631, 4100657, 4100666, 4100668, 4100669, 4100671, 4100673, 4100674, 4100675, 4100676, 4100684, 4100699, 1907875, 4001258, 4002379, 4003677, 4003687, 4004444, 4102310, 4102770, 4001361, 4001265, 4004791, 4007071, 4102287, 4002697, 4002103, 4004106, 4003674, 4003688, 4100201, 4100778, 4100965, 4101022, 4003551, 4004570, 4004234, 4100681, 4002695, 4003230, 4100197, 4006794, 4102402, 4101084, 1911868, 4001738, 4100785, 4100811, 4101208, 4100410, 4001253, 1904368, 1909146, 4001564, 4002626, 4004034, 4100548, 4004241, 4101389, 4005165, 1909150, 4100830, 4101952, 4005493, 4005497, 4000725, 4001851, 4100813, 4100431, 4003348, 4100512, 4004466, 4102039, 4100628, 4000733, 1902606, 1904364, 1904366, 1904370, 1906235, 1906264, 1907503, 1907865, 1909152, 1909377, 1909404, 1909410, 1909808, 1910318, 1910322, 1910554, 1911152, 1911159, 1911528, 1911570, 1911681, 4000154, 4000329, 4000357, 4000430, 4000445, 4000447, 4000449, 4000727, 4000831, 4001090, 4001101, 4001260, 4001492, 4001776, 4001889, 4002403, 4002696, 4002722, 4002723, 4002757, 4002761, 4002813, 4002840, 4003090, 4003194, 4003278, 4003281, 4003356, 4003718, 4003822, 4004760, 4004769, 4004776, 4004784, 4005226, 4005941, 4006018, 4006824, 4006851, 4006945, 4100775, 4100968, 4101108, 4102321, 4102709, 4005131, 4002065, 4004450, 4001589, 4003588, 4003597, 4003624, 4003635, 4003637, 4003683, 4100698, 4102404, 4000070, 4100559, 4000147, 4100152, 4006999, 4001853, 4005227, 4005881, 4007046, 4102411, 4101405, 4101416, 4002351, 4003308, 4004574, 4100502, 4005136, 4101360, 4006408, 4001772, 4007104, 4001121, 4002097, 4002667, 4004442, 4000165, 4000572, 4100324, 4100413, 4006822, 4100501, 4100033, 4001781, 4006357, 4005223, 4002161, 4004195, 4100764, 4102509, 4101006, 4101957, 4100514, 1907871, 4005164, 4000735, 4100859, 4102319, 4004822, 4000067, 4002132, 4101063, 4101410, 4001779, 4004848, 4000434, 4005142, 4005220, 4001739, 4003346, 4004881, 1909084, 4004198, 4006933, 4100820, 4002764, 4005376, 4001744, 4100325, 4101028, 1909374, 4102322, 4004199, 4001120, 4000624, 4000155, 4002666, 4003313, 4102412, 4002621, 4002923, 4005905, 4101876, 4006361, 4006589, 4006606, 4006626, 4005809, 4100487, 4002996, 4003151, 4004884, 4101415, 4003015, 4101832, 4003154, 4102771, 1911612, 4001559, 4001592, 4003480, 4003495, 4004440, 4005390, 4005439, 1907512, 4101393, 4004041, 4002628, 4007041, 4007090, 4007101, 4101723, 4001591, 4003678, 4000349, 4102947, 4006821, 4000581, 1895793, 4000151, 4000214, 4002991, 4006973, 1906233, 4001098, 4001117, 4002662, 4002691, 4003526, 4006617, 4100074, 4102041, 1909806, 4004281, 4005883, 4000210, 4101185, 4101628, 4005926, 4000630, 4003530, 4005375, 4005802, 4007078, 4005448, 4000207, 4002838, 4005965, 4006303, 4006601, 4007077, 4100132, 4100441, 4100464, 4100706, 4100731, 4100927, 4100934, 4101133, 4101146, 4101157, 4101350, 4101641, 4101805, 4101824, 4101837, 4101866, 4102171, 4102172, 4102210, 4102212, 4102662, 4103006, 4001362, 4004004, 4004240, 4001359, 4002331, 4003419, 4003422, 4003459, 4003460, 4003541, 4003819, 4004566, 4004856, 4005263, 4005915, 4100077, 4100082, 4100961, 4101911, 4101964, 4102037, 4102038, 4102324, 4102573, 4102574, 4102617, 4102992, 4101839, 4006009, 4002733, 4007001, 4101397, 4102853, 4102968, 4102970, 4100547, 4102032, 4102060, 4101462, 4101188, 4004397, 4101025, 4102865, 4102869, 4102811, 4002912, 4003681, 4102508, 4102280, 4101200, 4103054, 4001489, 4101179, 4101472, 4102720, 4102410, 4102854, 4102407, 4102316, 4000768, 4001972, 4002946, 4003009, 4003018, 4003509, 4003529, 4003560, 4003600, 4004452, 4004571, 4004668, 4005048, 4005339, 4005949, 4005954, 4006927, 4006939, 4007002, 4007067, 4007084, 4100042, 4100078, 4100217, 4100695, 4100777, 4100823, 4100928, 4100931, 4100933, 4101134, 4101182, 4101223, 4101356, 4101468, 4101600, 4101635, 4101767, 4101962, 4102035, 4102117, 4102274, 1908767, 191021
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
B-K Medical A/S
Reason for Recall:
GE HealthCare has become aware that the Instructions for Use within the user guide and service manual do not adequately state the required battery replacement intervals, end-of-life handling procedures, or general battery safety precautions.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

1202 Flex Focus Ultrasound System, cart models UA1214 and UA1814, used with: Flex Focus 200, Flex Focus 300, Flex Focus 400, Flex Focus 500, Flex Focus 700, Flex Focus 800; System, Imaging, Pulsed Doppler, Ultrasonic

Product Codes/Lot Numbers:

1) GTIN 05704916001056, Serial Numbers: 4005090, 4005710, 4002577, 4001807, 4002211, 4004085, 4006424, 4004772, 4101695, 4006420, 4100894, 4003775, 4004724, 4000321, 4004803, 4101755, 4101757, 4004635, 4004641, 4003082, 4004188, 4004448, 4100213, 4101302, 4000466, 4002982, 4006605, 4004941, 4100713, 4003569, 4003390, 4006608, 4101728, 4101906, 4005490, 4100128, 4001502, 4005922, 4006602, 4101724, 4005913, 4101394, 4001066, 4005230, 4000771, 4001536, 4003417, 4004297, 4102893, 4103104, 4003275, 4004575, 4100395, 4004038, 4004201, 4006350, 4100035, 4006123, 4100092, 4005380, 4005961, 4000460, 4005679, 4001915, 4005334, 4005346, 4001503, 4001694, 4003141, 4003626, 4004412, 4004431, 4005447, 4100060, 4005190, 4002880, 4101872, 4004088, 4004339, 4102033, 4005920, 4101823, 4007095, 4005228, 4004443, 4101990, 4002023, 4002190, 4005993, 4004423, 4101552, 4006362, 4001532, 4100603, 4005243, 4006010, 4006113, 4006114, 4006117, 4006126, 4101988, 4006950, 4006815, 4005778, 4101506, 4102036, 1909816, 4101756, 4002916, 4005906, 4101758, 4003821, 4005258, 4003139, 4101504, 4003884, 4005352, 4005769, 4005771, 4101555, 4003052, 4100463, 4007074, 4100140, 4002947, 4001137, 4006812, 4001266, 4005650, 4101956, 4003731, 4006409, 4004632, 4005337, 4100943, 4102414, 1910262, 4000166, 4101225, 4003164, 4001683, 4005372, 1910167, 4001651, 4005649, 4002186, 4101725, 4101991, 4000074, 4000845, 4001946, 4004994, 4007036, 4007038, 4007039, 4007040, 4007042, 4007053, 4100040, 4100057, 4100068, 4100069, 4100070, 4100127, 4100194, 4100819, 4100822, 4100839, 4100870, 4100873, 4101526, 4101554, 4101632, 4101646, 4101677, 4101803, 4101828, 4101836, 4101863, 4101867, 4101983, 4102012, 4102015, 4102105, 4005729, 4102419, 4100133, 4005972, 4102328, 4102417, 4004344, 1910010, 4004855, 4006810, 4003561, 4005461, 4101634, 4100330, 4006425, 4101099, 4004192, 4005257, 4006943, 1907405, 1907367, 4002579, 4100225, 4101206, 4005868, 4005872, 4102993, 4101253, 4001126, 4005184, 4003763, 4001414, 4001692, 4001702, 4005128, 4101117, 4101958, 4101538, 4005496, 4001534, 4001656, 4101349, 4001996, 4002470, 4005442, 4005928, 4002588, 4005969, 4001421, 4004649, 4101537, 4006418, 4006809, 4002276, 4100887, 4001389, 4100995, 4101180, 4102121, 4003545, 4006998, 4102575, 4001707, 4004885, 4006796, 4006404, 4102416, 4006997, 4006941, 4101826, 4101114, 4004766, 4002422, 4004696, 4007085, 4003532, 4101804, 4005735, 4003770, 4101606, 4003229, 4001132, 4001133, 4001136, 4001448, 4003059, 4003885, 4004343, 4004567, 4005936, 4100744, 4101601, 4003000, 4002272, 4002467, 4005797, 4003682, 4004789, 4001499, 4004006, 4004426, 4005485, 4005487, 4000782, 4003768, 4101249, 4101304, 1907823, 1908864, 4003525, 4003539, 4004671, 4101400, 4101401, 4101403, 4101407, 4003391, 4101228, 4005945, 4005793, 4101722, 4005680, 4002448, 4000145, 4000889, 4000421, 4002995, 4002839, 4004413, 4001680, 4003542, 4005977, 4005341, 4006624, 4006879, 4005169, 4100533, 4100600, 4005504, 4100679, 4005998, 4003772, 4005168, 4005383, 4101404, 4101955, 4005373, 4100096, 4101156, 4005967, 4101245, 4100480, 4004232, 4006924, 4101132, 4005773, 4007097, 4100761, 4101398, 4005351, 4004628, 4004658, 4004999, 4005335, 4101247, 4004889, 4005444, 4004725, 4007049, 4005088, 4005344, 4003056, 4100411, 4004042, 4100062, 4100454, 4003162, 4004226, 4000966, 4100390, 4100210, 4003415, 4003457, 4003879, 4004385, 4005935, 4007007, 4006486, 4004467, 4005384, 4101270, 4004875, 4004010, 4101419, 4102169, 4002921, 4002948, 4004436, 1910149, 4006308, 4000841, 4001531, 4003528, 4003198, 4003625, 4005436, 4003160, 4003684, 4006801, 4100086, 4101396, 4004828, 4006874, 4004422, 4000994, 4006329, 4006987, 4005249, 4100550, 4005369, 4005092, 4005229, 4005642, 4005879, 4101361, 4002277, 4003054, 4004618, 4005724, 4005852, 4005934, 4003225, 4006932, 4006988, 4001381, 4001415, 4005185, 4000136, 4000327, 4000461, 4000464, 4000474, 4004698, 4001686, 4004231, 4001422, 4005507, 4101183, 4101764, 4101961, 4101967, 4002352, 4005751, 4002069, 4002625, 4101701, 4100232, 4005086, 4006333, 4004656, 4005743, 4000871, 4000882, 4004598, 4101687, 4001690, 4001992, 4002020, 4002021, 4002078, 4005374, 4100546, 4004388, 4007072, 4005027, 4001805, 4002729, 4003140, 4001418, 4003017, 4004781, 4001497, 4101769, 4002398, 4007058, 4004338, 4002575, 4001498, 4004393, 4101638, 4005661, 4005942, 4100432, 4000797, 4000798, 4102801, 1910018, 4000324, 4001410, 4003025, 4003881, 4004868, 4001920, 4003946, 4004225, 4100151, 4100145, 4002183, 4002214, 4003282, 4003503, 4004590, 4006330, 4005981, 4007035, 4007052, 4007059, 4101364, 4004830, 4006355, 4100910, 4101248, 4003565, 4006876, 4006621, 4007099, 4100936, 4001409, 4101834, 4003316, 4004002, 4002220, 4003317, 4004565, 4101409, 4002507, 4003512, 4004793, 4002217, 4003081, 4100973, 4001924, 4002209, 4005235, 4005840, 4100216, 4101408, 4004390, 4004403, 4002846, 4005924, 4004729, 4100906, 4100911, 4100930, 4000185, 4000323, 4001135, 1911127, 4000143, 4000141, 4000164, 4001494, 4001687, 4003521, 4101532, 4101729, 4100129, 4103008, 4005338, 4004715, 1910260, 4001809, 4002587, 4101809, 4001971, 4007106, 4004425, 4005957, 4100147, 4100963, 4003452, 4006934, 4102640, 4002585, 4100722, 4005720, 4005727, 4005038, 4003527, 4101005, 4003451, 4005897, 4101023, 4102951, 4005984, 4003226, 4000465, 4002808, 4003318, 4004831, 4004838, 4005345, 4102534, 4000325, 4004083, 4005678, 4002401, 4005651, 4007103, 4100503, 4102497, 4005170, 4006598, 4000877, 4000883, 4006415, 4003079, 4002915, 4002953, 4006000, 4001263, 4001700, 4003138, 4100971, 4003002, 4004432, 4006599, 4101970, 1906178, 4005239, 4001649, 4004633, 4003458, 4003501, 4002514, 4004643, 4004926, 4007062, 4004773, 4005033, 4101169, 4004761, 4005672, 4004988, 4005261, 4004708, 4002815, 4005952, 4004642, 4004826, 4005974, 4006410, 4006858, 4006949, 4100196, 4100904, 4003478, 4004987, 4001412, 4005890, 4007079, 4005386, 4005767, 4000775, 4001802, 4006975, 4005902, 4003553, 4004454, 4005030, 4004592, 4006335, 4002170, 4101226, 4005668, 4100202, 4103007, 4002583, 4002958, 4003547, 4004700, 4005850, 4004886, 4001416, 4100056, 4100061, 4006004, 4004931, 4005665, 4100743, 4102826, 4001387, 4100705, 4002913, 4102052, 4004637, 4100714, 4003114, 4003880, 4001078, 4005503, 4001506, 1910254, 4004400, 4004861, 4006596, 4101417, 4101915, 4006407, 4002491, 4004404, 4004820, 4002540, 1910141, 4004424, 4101630, 4101685, 4101686, 4005831, 4007076, 4006121, 4102276, 4004697, 4002248, 4005791, 4005846, 4102535, 4002275, 4004441, 4101989, 4100440, 4005028, 4102468, 4006011, 4006811, 4004572, 4003766, 4004286, 4100862, 4005907, 1908876, 4004701, 4100939, 4102835, 4101266, 4100212, 4006305, 4005459, 4102519, 4002763, 4101754, 4002949, 4001407, 4001420, 4001508, 4003511, 4003938, 4002806, 4006937, 4100075, 4100076, 4100080, 4000783, 4000416, 4002442, 4003781, 4003277, 4003389, 4005884, 4005889, 4100054, 4005163, 1900917, 1900975, 1901001, 1906172, 1906174, 1906176, 1907401, 1907466, 1907497, 1908878, 1908968, 1910014, 1910139, 1910143, 1910145, 1910173, 1910221, 1910264, 1911129, 1911838, 4000068, 4000417, 4000419, 4000469, 4000773, 4000776, 4000780, 4000840, 4000878, 4000888, 4000967, 4000970, 4001067, 4001071, 4001139, 4001192, 4001383, 4001419, 4001449, 4001451, 4001495, 4001509, 4001512, 4001514, 4001515, 4001533, 4001540, 4001542, 4001550, 4001636, 4001638, 4001642, 4001646, 4001684, 4001685, 4001689, 4001691, 4001695, 4001703, 4001705, 4001839, 4001841, 4001916, 4002024, 4002025, 4002168, 4002169, 4002189, 4002244, 4002246, 4002306, 4002307, 4002399, 4002419, 4002420, 4002421, 4002423, 4002426, 4002444, 4002446, 4002468, 4002484, 4002485, 4002617, 4002618, 4002622, 4002623, 4002627, 4002725, 4002765, 4003045, 4003053, 4003186, 4003349, 4003767, 4003935, 4003943, 4004033, 4004415, 4004438, 4004458, 4004460, 4004471, 4004472, 4004694, 4004709, 4004727, 4004758, 4004867, 4004990, 4004998, 4005034, 4005245, 4005776, 4005801, 4005838, 4005882, 4006614, 4006978, 4100150, 4100199, 4100490, 4100724, 4100892, 4100912, 4100937, 4100942, 4101001, 4101087, 4101105, 4000167, 4005975, 4000186, 4003537, 4005886, 4001537, 4004730, 4002769, 4005914, 4005443, 4004864, 4101131, 4000993, 4005437, 4000996, 4001997, 4003084, 4003554, 4000774, 4000190, 4004940, 4001496, 1909812, 4003020, 4002375, 4002513, 1910012, 4000969, 4002449, 4002985, 4000847, 4002215, 4007073, 4102059, 4001944, 4001706, 4005342, 1907464, 4001643, 4001681, 4005768, 4007051, 4101137, 4002877, 4004768, 4100064, 4100895, 4001966, 4001191, 4004337, 4004340, 4004943, 4000793, 1910020, 4003522, 4004778, 4101531, 4006823, 4001510, 4100422, 4100433, 4100434, 1906625, 1910016, 4000183, 4002586, 4000153, 4005750, 4003497, 4003538, 4005871, 1907399, 4004576, 4004421, 4004429, 4004837, 4005777, 4004427, 4002539, 4004411, 4101533, 4003729, 4004616, 4000876, 4000884, 4000887, 4006983, 4002578, 4003727, 4004473, 4004832, 4005870, 4004228, 4006592, 4004854, 4006918, 4006935, 4101148, 4003723, 4003596, 4005186, 4005222, 4004439, 4005653, 4005343, 4100041, 4100055, 4100059, 4007066, 1900918, 4002998, 4003117, 4003158, 4005445, 4101684, 4002997, 4003817, 4006929, 4004435, 4003044, 4004634, 4005985, 4003877, 4100207, 4003019, 4002167, 4005867, 4002187, 4005044, 4102922, 4101731, 4102016, 4003496, 4003510, 4005237, 4005786, 4000891, 4004196, 4005040, 4004447, 4102996, 4102208, 4100215, 4005669, 4006854, 4100421, 1911806, 4004457, 4003486, 4005733, 4003878, 4005893, 4003146, 4101466, 4007080, 4005189, 4003426, 4100451, 4100913, 4001076, 4003309, 4005452, 4005456, 4002274, 4006871, 4003468, 4004657, 4004942, 4102553, 4005648, 4001417, 4004928, 4000843, 4001998, 4006124, 4101639, 4005997, 4002210, 4100131, 4004937, 4001635, 4003189, 4100444, 4001075, 4001513, 4003592, 4006798, 4005045, 4000890, 4004434, 4003514, 4005746, 4100135, 4005742, 4000880, 4102055, 1910147, 4003057, 4004780, 4002354, 4004332, 4101101, 4102273, 4102627, 4102950, 4005387, 4100420, 4006492, 4006016, 4005811, 4003815, 4102420, 4003479, 4003314, 4005382, 4005491, 4005640, 4102277, 1910314, 4000995, 4002212, 4101711, 4001493, 4003550, 4004112, 4005671, 4003477, 4003494, 4003500, 4003505, 4003507, 4003513, 4003628, 4003508, 4005043, 4000965, 4000971, 4003042, 4004036, 4004626, 4005677, 4005847, 4005851, 4005853, 4005876, 4005895, 4005903, 4005921, 4005980, 4004445, 4004446, 4004874, 4001202, 4000322, 4000471, 4005973, 4101176, 4001388, 4101136, 4001072, 4001134, 4000873, 4000886, 4001803, 4002185, 4003193, 4005259, 4100861, 4003425, 4003232, 4005161, 4005162, 4002402, 4002469, 1908972, 4006309, 4004342, 4004005, 4000184, 4004849, 4101627, 1910256, 4002213, 4002728, 4003732, 4005134, 4005918, 4102994, 4004651, 4004653, 4004189, 4003046, 4004813, 4004927, 4005932, 4000998, 4006483, 4101529, 4102722, 4000781, 4002488, 4003049, 4006411, 4101631, 4102915, 4005670, 4002400, 4101002, 4006795, 4006001, 4002630, 4100832, 4006436, 4102800, 4003816, 4004851, 4003568, 4004190, 4004227, 4004236, 4004239, 1911125, 4001704, 4000336, 4005859, 4000328, 4002954, 4003219, 4101700, 4005233, 4003546, 4003780, 4005856, 4101909, 4005195, 4101106, 4005505, 4001507, 4102948, 4101912, 4001069, 4002245, 4101391, 4003148, 4005250, 4006594, 4005130, 4004865, 4001425, 4003558, 4002994, 4002301, 4004624, 4003474, 1911808, 4102762, 4003524, 4000422, 4001505, 4100450, 4102122, 4004341, 4101692, 4001838, 4005888, 4005908, 4006412, 1911945, 4102829, 4102758, 4002241, 4101705, 4101135, 4002986, 4003041, 4000470, 4000997, 4002758, 4002766, 4101202, 4101000, 4002222, 4100091, 4100711, 4006364, 4102018, 4001068, 4004597, 4101273, 4005784, 4001538, 4001413, 4101762, 4101474, 4003543, 4101205, 4000334, 4004193, 4101760, 4001065, 4004704, 4001535, 4003629, 4101640, 4000188, 4000413, 1893197, 4000842, 4004420, 4001640, 4102058, 4004282, 4101020, 4102013, 1910258, 4000333, 4002490, 4001074, 4005774, 4002184, 4002302, 4002304, 4002305, 4004084, 4004086, 4102314, 4100972, 4002920, 4102546, 4006872, 4005986, 4004629, 4101246, 4000972, 4003534, 4100482, 4002353, 4003383, 4003587, 4101177, 4101643, 4005891, 4001188, 4005644, 4101868, 4001644, 1910223, 4005877, 4003776, 4004862, 1906237, 1910165, 4002219, 4101021, 4005652, 4100387, 4005660, 4101353, 4000418, 4001408, 4003108, 4101062, 4004081, 4004117, 4100762, 4100063, 4102119, 4101219, 4101807, 4101822, 4003536, 4001073, 4006307, 1908872, 1908974, 1908976, 4001122, 4001511, 1910163, 4001639, 4100709, 4003490, 4006944, 4000849, 4001447, 4001450, 4005492, 4003024, 4003633, 4003686, 4002489, 4002767, 4001941, 4101959, 4005381, 4005449, 4005645, 4101024, 4006866, 4003818, 4004078, 4006797, 1907354, 1907356, 4001190, 4002814, 4003562, 1907358, 1907468, 4000875, 4000881, 4005681, 4005711, 4006817, 4002404, 4004295, 4005241, 4005252, 4102760, 4007056, 4102211, 4005880, 4002811, 4006989, 4006931, 4001801, 4004200, 4005929, 4100095, 4100753, 4003043, 4004695, 4005133, 4005135, 4005139, 4005140, 4006110, 4004866, 4001658, 4005770, 4100038, 4007070, 4005196, 4003320, 4007005, 1909814, 4003601, 4002182, 4003717, 4005855, 4005857, 4102206, 4007108, 4100462, 4006800, 4003283, 4005912, 1910215, 4001693, 4002218, 4000879, 4006863, 4003544, 4005901, 4004717, 4000472, 4000457, 4000423, 4006015, 4007105, 4002347, 4101766, 4101768, 4004844, 4005833, 4000872, 4002221, 4004863, 4102053, 4101172, 4102477, 4002999, 4001701, 4005937, 4100493, 4100534, 4100549, 4102467, 4006882, 4101178, 4004589, 4004595, 4005231, 4000152, 4100126, 4102061, 4003557, 4003190, 4100418, 4001647, 4003143, 4102925, 4004936, 4100750, 4101471, 4101707, 4001942, 4001969, 4001994, 4003469, 4100893, 4003556, 4002019, 4001641, 4001917, 4004869, 4005446, 4101362, 4101835, 4101840, 4102478, 4101802, 4005979, 4102281, 4005647, 4005739, 4005869, 4100583, 4101251, 4101986, 4000160, 4006857, 4006859, 4006922, 4007045, 4004335, 4100708, 4101704, 4101881, 4003506, 4006849, 4100415, 4101808, 4101224, 4003111, 4003594, 4004197, 4004938, 4101904, 4100924, 4003947, 4005834, 4005841, 4005842, 4006792, 4100882, 4002377, 4002068, 4002807, 4001411, 4001099, 4102123, 4102164, 4102165, 4102166, 4102571, 4102572, 4102580, 4103014, 4101602, 4005885, 4005892, 4005909, 4100193, 4102168, 4006313, 4006413, 4007047, 4101535, 4001500, 4001799, 4005740, 4004459, 4100058, 4000885, 4100142, 4100071, 4100097, 4003690, 4101181, 4101726, 4005187, 4004706, 4003351, 4006422, 4004087, 4003107, 4005191, 4002447, 4000473, 4005788, 4006802, 4006005, 4101920, 4101954, 4003720, 4004721, 4004835, 4003145, 4006613, 4005754, 4005506, 4101763, 4006435, 4102619, 1908970, 4000420, 4003942, 4100442, 4004588, 4006012, 4005171, 4004719, 4002803, 4005988, 4002242, 4003689, 4004663, 4004666, 4004842, 4004843, 4005728, 4006917, 4101509, 4101301, 4004659, 4004871, 4005248, 4102836, 4102837, 4101683, 4006818, 4006007, 4002546, 1907462, 4001382, 4100067, 4004007, 4100130, 4001539, 4001682, 4003196, 4006816, 1910171, 4000772, 4100416, 4001386, 4101357, 4100227, 1907763, 1911804, 4002950, 4102056, 4001501, 4003883, 4004428, 4101551, 4002984, 4100034, 4100907, 4100908, 4101702, 4000796, 4002849, 4101917, 4002181, 4002188, 4005732, 4004818, 4101642, 4101678, 4101682, 4101696, 4004710, 4006948, 4000784, 4006936, 4001189, 4000468, 4101174, 4100886, 4100066, 4002022, 4003286, 4003355, 4003472, 4003492, 4004014, 4004989, 4006916, 4007087, 4004564, 4006861, 4002067, 4006426, 4101806, 4102401, 4000462, 4101184, 4001504, 4100678, 4102867, 4005667, 4102621, 4102601, 4001688, 4000964, 4002850, 4002883, 4005982, 4004437, 4100994, 4002465, 4005765, 4101387, 4000844, 4102914, 4005795, 4006125, 4005789, 4004877, 4005333, 4001648, 4001991, 4100484, 4006109, 4005878, 4002216, 4007102, 4001128, 4001652, 4100391, 4003083, 4003455, 4005976, 4102282, 1911835, 4003387, 4005943, 4102628, 4102312, 4005955, 4100214, 4100707, 4003630, 4002576, 4005676, 4100929, 4100932, 4101864, 4000138, 4004883, 4002076, 4002875, 4000139, 4000415, 4100997, 4101019, 4000071, 4004720, 4004722, 4005927, 4101827, 4002547, 4002847, 4004673, 4004850, 4005347, 4006942, 4001970, 4006416, 4004408, 1908874, 4005990, 4006867, 4002247, 4002957, 4003889, 4004290, 4004294, 4004296, 4102514, 4005132, 4002443, 4005499, 4005787, 4100999, 4006311, 4005440, 4101680, 4004723, 4101597, 4001426, 4000777, 4100200, 4003192, 4100696, 4100530, 4002956, 4102213, 4001530, 4000414, 4000424, 4000463, 4004631, 4006814, 4000846, 4000848, 4001000, 4004630, 4005781, 4100884, 4101916, 4101865, 4101905, 4001650, 4101694, 4100230, 4100322, 4006869, 4003106, 4004804, 4100721, 4101227, 4005643, 4006013, 4101691, 4001077, 4004090, 4102971, 4003599, 4007050, 4004334, 4001541, 4100088, 4100094, 4102469, 4004620, 4100960, 4001544, 4005046, 4102828, 4005775, 4102209, 4005849, 4002809, 4003109, 4003713, 4002466, 4001637, 4101059, 4101060, 4003473, 4004416, 4000968, 4102543, 4006366, 4004619, 4102019, 4000182, 4002805, 4000467, 4005992, 4001843, 1908962, 4003222, 4005968, 4000412, 4101100, 4101679, 4005717, 4003022, 4101346, 4006590, 4005253, 4101553, 4004235, 4003447, 4100224, 4003485, 4002486, 4001808, 4103041, 4004079, 4101833, 4005953, 4101170, 4001204, 4003161, 4005026, 4001645, 4003631, 4004430, 4002511, 4001360, 4003730, 4005225, 4006921, 4003523, 4005709, 4006947, 4002887, 4004089, 4000410, 4000779, 4005887, 4005712, 4001840, 4005389, 4006331, 4101351, 4004399, 4100090, 4000778, 4000800, 4003378, 4000999, 4007089, 4003502, 1910022, 4006111, 4004469, 4003003, 4000187, 4004935, 4005032, 4005766, 4001804, 4001844, 4005232, 4002271, 4002308, 4005141, 4003080, 4001424, 4001919, 4102714, 4002756, 4001965, 1906627, 4004665, 4006628, 4005755, 1908870, 4007075, 4002463, 4006414, 4002988, 4004827, 4004417, 4003540, 4003673, 4004410, 4004829, 4001800, 4004888, 4005719, 4000794, 4003163, 4006595, 4100760, 4100763, 4101500, 4101501, 4003552, 4002582, 4003493, 4005348, 4006820, 4101984, 4002841, 4002851, 4003321, 4004622, 4101392, 4004991, 1910219, 4005782, 4001914, 4006847, 4003888, 4000189, 4003448, 4006491, 4006940, 4004409, 4002376, 4100087, 4004782, 4005721, 4003142, 4004716, 4100890, 4102622, 4101810, 4002619, 4006427, 4004578, 4002955, 4005899, 4005495, 4002509, 4101355, 4000458, 4006915, 4006928, 4004770, 1911734, 4000459, 4003424, 4000411, 4002876, 4002842, 1908858, 4000069, 4000073, 4001806, 4001810, 4001913, 4101003, 4005744, 4006597, 4102120, 4101130, 4101144, 4101147, 4101203, 4102866, 4005438, 4006118, 4006850, 4100712, 4100481, 4101155, 4002445, 4100053, 4005388, 4002510, 4005336, 4001131, 4002487, 4005916, 4103017, 4101418, 4005675, 4000623, 4000628, 4001466, 4001585, 4001586, 4004449, 4005138, 4005509, 4005641, 4005654, 4005655, 4005657, 4005658, 4005659, 4005663, 4101271, 1902581, 1908741, 1908745, 1910137, 4000149, 4000209, 4000442, 4000827, 4001088, 4001095, 4001112, 4001130, 4001490, 4002508, 4003279, 4003352, 4003384, 4003517, 4004718, 4005193, 4005484, 4006354, 4007082, 4100204, 4102913, 4004779, 4100560, 4004462, 4100617, 1907495, 4002071, 4101460, 4007096, 4005451, 4004786, 4001618, 4002629, 4004762, 4005779, 4101534, 4101708, 4102057, 4004794, 4006349, 4002843, 4002380, 4003344, 4003388, 4004035, 4101690, 4101698, 4101699, 4001778, 4004796, 4102710, 4101186, 4004731, 4003449, 4004039, 4004115, 4004887, 4006003, 4002329, 4102579, 4003312, 4005803, 4006484, 4100734, 4102577, 5102579, 4003116, 4102285, 4002700, 4003014, 4004032, 4004468, 4004569, 4004879, 4005832, 4006487, 4006926, 4101395, 4102021, 4102620, 4102626, 4102769, 4102772, 4006116, 5006892, 4004878, 4102708, 4100881, 4101057, 4005948, 4101875, 4101902, 4100206, 4006923, 4102768, 4005486, 4006119, 4100079, 4100072, 1908651, 4002541, 4006008, 4101710, 4002952, 4005458, 4000678, 4002126, 4102752, 4100654, 4004037, 4102470, 4005260, 4003627, 4003679, 4004996, 1909090, 4005854, 4001587, 4002551, 4002657, 4004986, 4100329, 4101645, 1910962, 4000144, 4001527, 4101347, 4101918, 4005673, 4101811, 4000575, 4001529, 4003086, 4003721, 4100786, 4102272, 4004674, 4004714, 4005736, 4002131, 4001206, 1907441, 4004636, 4004639, 4100697, 4003563, 4004644, 4100139, 4100198, 4100491, 4101116, 4101129, 4005350, 4101244, 4101104, 4101505, 4101058, 4101061, 4002727, 4001624, 4005845, 4001249, 4100461, 4101689, 4100964, 4101709, 4000162, 4005966, 4100821, 4101413, 4102311, 4102405, 4102639, 1910150, 4003004, 4102325, 4003535, 4006979, 4001380, 4001918, 4006620, 4002303, 4003021, 4003559, 4003598, 4005874, 4100093, 4001423, 4101841, 4005898, 4002584, 4004800, 4002278, 4000799, 4003782, 4000874, 4000785, 4000795, 4006803, 4002378, 4002381, 1911732, 4001993, 4003691, 4102492, 4003564, 4003886, 4100231, 4002382, 4100885, 4002989, 4001193, 4004407, 4004771, 4006981, 4000870, 4004841, 4006853, 4004405, 4002542, 4100602, 4005848, 4006808, 4101604, 4001842, 4006920, 4101557, 4004401, 4102315, 4004040, 4005910, 4005662, 4005875, 4004001, 4101960, 4004699, 4006619, 4000181, 4006403, 4101644, 4100125, 4002243, 4001070, 4001203, 1911039, 4003944, 4004233, 4004930, 4003091, 4003820, 4101525, 4006419, 4002620, 4003147, 4100834, 4006881, 4002512, 4103042, 4100065, 4101969, 4005036, 4002768, 4001543, 4004617, 4101252, 4003227, 4001780, 4003427, 4003462, 4004669, 4004819, 4004934, 4005860, 4005896, 4005958, 4006865, 4100601, 4100664, 4100751, 4101388, 4101636, 4102020, 4102204, 4102555, 4102618, 4103106, 2002566, 2002667, 2002668, 4000330, 4000331, 4000335, 4002330, 4002334, 4002654, 4002658, 4002659, 4002660, 4002661, 4002663, 4002664, 4002688, 4002689, 4002690, 4002692, 4002693, 4004623, 4004662, 4004853, 4004860, 4004870, 4004873, 4005143, 4005183, 4005234, 4005240, 4005790, 4005836, 4005959, 4006607, 4006875, 4007093, 4007094, 4100223, 4100388, 4100389, 4100392, 4100394, 4100419, 4100580, 4100776, 4100842, 4101107, 4101268, 4101354, 4101530, 4101556, 4101869, 4101914, 4101965, 4101987, 4102493, 4102547, 4102624, 4102625, 4102721, 4102784, 4102787, 4102827, 4102830, 4102911, 4102912, 4102921, 4103004, 4103005, 4103028, 4103029, 4103030, 4103031, 4103032, 4103043, 4103044, 4103056, 4103092, 4103093, 4103094, 4103095, 4103096, 4103103, 4103105, 4103130, 4103134, 4103149, 4103151, 4103152, 4000441, 4001491, 1911677, 4000836, 4004792, 4007006, 4101913, 4102317, 1893118, 4006880, 4102600, 4102851, 4102852, 4000433, 4001486, 4001594, 4003112, 4003276, 4005810, 4006310, 4101761, 1901888, 4003159, 1907817, 4003088, 4000838, 1902575, 1907547, 1911572, 1911710, 4000208, 4000435, 4000573, 4002332, 4002762, 4002804, 4003589, 4004013, 4002731, 4001945, 1907761, 4000212, 4000451, 4001119, 4001590, 4001847, 4005656, 4005808, 4100149, 4101953, 1911074, 1908239, 1908241, 1908243, 1908245, 4000443, 4003421, 4101963, 4001118, 4005931, 4006433, 4001922, 4002066, 4004077, 4004383, 4005453, 4005783, 4005796, 4006431, 4006623, 4102785, 4102786, 4103045, 4003228, 1911947, 4001967, 4002166, 4002987, 4005166, 4100083, 4100563, 4102498, 4100528, 4004110, 4004387, 4004783, 4005242, 1904372, 4000213, 4004453, 4101503, 4101507, 4101527, 4102797, 4102967, 4000021, 1908747, 4001264, 4007037, 4005930, 1904324, 4001657, 4003423, 4005129, 4003453, 4006406, 4007098, 4006312, 4005873, 1909803, 1905149, 1907759, 4001525, 4001619, 4001622, 4002624, 4002848, 4002878, 4003055, 4003060, 4003089, 4004463, 4004579, 4004591, 4004596, 4006603, 4100831, 4100840, 4100843, 4004389, 4101399, 4101464, 4101465, 4101629, 4001655, 4002333, 4006609, 1911708, 4100529, 4002162, 4000828, 4000829, 4000832, 4000833, 4000834, 4000835, 4000837, 4003156, 4003187, 4003188, 4003191, 4003197, 4003224, 4004456, 4102578, 4102587, 4102588, 4102596, 4102597, 4102598, 4102599, 4000072, 4001522, 4102279, 4007060, 4001262, 1911730, 4003416, 4006938, 4000577, 4000582, 4001850, 4003381, 4005488, 4102715, 4005753, 4001113, 4001116, 4001100, 4004845, 4004852, 4001105, 4002157, 4002701, 4001106, 4000923, 4002881, 4002885, 4003350, 4100531, 1911142, 4001087, 4003220, 4003675, 4003685, 4005188, 4005798, 4005951, 4006334, 4100962, 4005730, 4000345, 4002074, 4006990, 4101968, 4101971, 4000576, 4000731, 4001248, 4001261, 4001355, 4001626, 4002133, 4002655, 4002734, 4003157, 4003450, 4003567, 4003570, 4004291, 4004293, 4004655, 4004777, 4004798, 4004812, 4004815, 4005737, 4005747, 4005785, 4005991, 4006002, 4006360, 4006793, 4007061, 4100141, 4100967, 4101171, 4101173, 4101599, 4101603, 4101605, 4101878, 4102923, 4100863, 4100925, 4102892, 4101982, 4000156, 4000340, 4000353, 4000446, 4000729, 4000767, 4001008, 4001091, 4001094, 4001125, 4001357, 4002158, 4002163, 4002165, 4002581, 4002879, 4003047, 4003155, 4003221, 4003274, 4003284, 4003518, 4003531, 4003595, 4003774, 4004009, 4004030, 4004109, 4004111, 4004118, 4004707, 4004808, 4004823, 4004846, 4004985, 4005025, 4005039, 4005224, 4005349, 4005455, 4005494, 4005674, 4005682, 4005722, 4005738, 4005983, 4006315, 4006327, 4006352, 4006434, 4006437, 4006604, 4006984, 4007048, 4007081, 4100134, 4101345, 4100935, 4100085, 1900915, 1900922, 4102327, 4102408, 4101502, 4101765, 4101681, 4100417, 4101598, 4102891, 4006428, 4101550, 4003345, 1911840, 4005807, 4101187, 4003483, 4000452, 4102167, 4004834, 4001251, 4102207, 4101469, 4001557, 4003310, 4007055, 4004880, 4001250, 4001104, 4003945, 4001849, 4005664, 4006873, 4100996, 4003420, 4004774, 4101461, 4006353, 4004785, 4001741, 4006593, 4007083, 4100485, 4005858, 4102309, 4000427, 4005244, 4007092, 4000350, 4003287, 4101352, 4001560, 4005946, 1907863, 4001035, 4003603, 4005708, 4004932, 4005087, 1907499, 1907939, 4005096, 4001887, 4100732, 4101831, 4000211, 4002665, 4100733, 4102323, 4102712, 4001114, 4005919, 4005900, 4000036, 4006014, 4004465, 4101838, 4100651, 4100653, 4004229, 4006616, 4001561, 4005167, 4101467, 4100039, 4001563, 4006615, 4006819, 4002812, 4004939, 4003354, 4005221, 4006982, 4006994, 4100073, 4101027, 4003499, 1907568, 4100138, 4102014, 4004280, 4102643, 4000428, 4000578, 4004765, 4102759, 4000921, 4006996, 4005726, 4000348, 4001653, 4004464, 4005037, 4002548, 4102413, 4006627, 4101919, 4102054, 4001103, 4004230, 4004562, 4001554, 4101175, 4102641, 4102949, 4004455, 4102124, 4102163, 4004289, 4100412, 4001385, 4101220, 1907510, 4005262, 4005866, 4100009, 4102017, 4005925, 4002730, 4002874, 4002910, 4002914, 4004857, 4001565, 4005371, 4101877, 4005749, 4100809, 4000432, 4102170, 4102418, 4101688, 4102542, 4006804, 4000922, 4006306, 4001771, 4007008, 4004577, 4101528, 4000682, 4000343, 4005001, 4100810, 4001909, 4003504, 4005031, 4101871, 4005805, 1907873, 4005865, 4100938, 4100941, 4101386, 4101908, 4102713, 4005989, 4100513, 4001846, 4004847, 4003048, 4100998, 4005460, 4001775, 4002549, 4003520, 4004191, 4006860, 4101004, 4101730, 4001857, 4000158, 4000448, 4101727, 4100812, 4005666, 4002125, 4000625, 4100492, 4000455, 4001562, 4100779, 4102115, 4100511, 4005794, 4001555, 4100754, 4003012, 4003634, 4005500, 4003515, 4100735, 4001583, 4100710, 4000352, 4000437, 4100752, 4004807, 4005236, 4004451, 4007063, 4000631, 4001255, 1911712, 4000627, 4003010, 4004806, 4101825, 4001526, 4000728, 4100195, 4002922, 4006856, 4001256, 4005256, 4005378, 4004029, 4100494, 4101829, 4001092, 4001588, 4005095, 4000355, 4000168, 4000429, 4001584, 4100545, 4004627, 4005646, 1907827, 4101085, 4000677, 4003050, 4100787, 4101363, 4001552, 4000159, 4001006, 4102663, 4004621, 4005837, 4002543, 4003113, 1911523, 4002102, 4003418, 4004858, 4100084, 4101697, 4101390, 4102642, 4102286, 4102313, 4003347, 4004764, 4102284, 4000679, 4101414, 4101539, 4004814, 4001551, 4003319, 4102034, 4100833, 4101637, 4101693, 4101874, 4101966, 4102283, 4003488, 4006017, 4101264, 4006852, 4100745, 4100966, 4100500, 4101222, 4100970, 4004625, 4002993, 4101907, 4001123, 4102831, 4003385, 4100205, 4001740, 4100208, 4000912, 4001558, 4002759, 4000732, 4001252, 4003716, 4000730, 4005093, 4003728, 4102511, 4102969, 4003769, 4001115, 4001097, 1911679, 4000681, 4004461, 4006108, 4006115, 4100789, 4002917, 4003110, 4005904, 4003602, 4003941, 4001093, 4001102, 4001257, 4003590, 4100561, 4003475, 4000347, 4000425, 5004302, 4005716, 4004787, 4100453, 4004797, 4000456, 4007003, 4102924, 4002544, 4101406, 4102403, 4102406, 4000676, 4001358, 4006878, 4100841, 4002100, 4004107, 4102518, 4000351, 4002128, 4102409, 4102751, 4002694, 4006611, 4000444, 4005996, 4001742, 4001911, 4100211, 4100222, 4100229, 4100515, 4100526, 4100672, 4101463, 4101830, 4101879, 4101910, 4102278, 4102320, 4001773, 4002101, 4002273, 4004237, 4101753, 4101759, 4102723, 4002355, 4001743, 4003484, 4004795, 4004790, 4002845, 4002405, 4004732, 4005137, 4004433, 4004563, 4001770, 4003692, 4000620, 4002073, 4004840, 4001852, 4002357, 4003676, 4005741, 4003724, 4004573, 4006348, 4005192, 4001556, 4002425, 4003223, 4003144, 4005964, 4000734, 4002918, 4003115, 4000769, 4001774, 4000916, 4006405, 4000579, 4001553, 4005377, 4000622, 4000629, 4003087, 4003948, 4004654, 4100144, 4100148, 4100582, 4100589, 4100590, 4100615, 4100616, 4100618, 4100619, 4100620, 4100621, 4100627, 4100629, 4100630, 4100631, 4100657, 4100666, 4100668, 4100669, 4100671, 4100673, 4100674, 4100675, 4100676, 4100684, 4100699, 1907875, 4001258, 4002379, 4003677, 4003687, 4004444, 4102310, 4102770, 4001361, 4001265, 4004791, 4007071, 4102287, 4002697, 4002103, 4004106, 4003674, 4003688, 4100201, 4100778, 4100965, 4101022, 4003551, 4004570, 4004234, 4100681, 4002695, 4003230, 4100197, 4006794, 4102402, 4101084, 1911868, 4001738, 4100785, 4100811, 4101208, 4100410, 4001253, 1904368, 1909146, 4001564, 4002626, 4004034, 4100548, 4004241, 4101389, 4005165, 1909150, 4100830, 4101952, 4005493, 4005497, 4000725, 4001851, 4100813, 4100431, 4003348, 4100512, 4004466, 4102039, 4100628, 4000733, 1902606, 1904364, 1904366, 1904370, 1906235, 1906264, 1907503, 1907865, 1909152, 1909377, 1909404, 1909410, 1909808, 1910318, 1910322, 1910554, 1911152, 1911159, 1911528, 1911570, 1911681, 4000154, 4000329, 4000357, 4000430, 4000445, 4000447, 4000449, 4000727, 4000831, 4001090, 4001101, 4001260, 4001492, 4001776, 4001889, 4002403, 4002696, 4002722, 4002723, 4002757, 4002761, 4002813, 4002840, 4003090, 4003194, 4003278, 4003281, 4003356, 4003718, 4003822, 4004760, 4004769, 4004776, 4004784, 4005226, 4005941, 4006018, 4006824, 4006851, 4006945, 4100775, 4100968, 4101108, 4102321, 4102709, 4005131, 4002065, 4004450, 4001589, 4003588, 4003597, 4003624, 4003635, 4003637, 4003683, 4100698, 4102404, 4000070, 4100559, 4000147, 4100152, 4006999, 4001853, 4005227, 4005881, 4007046, 4102411, 4101405, 4101416, 4002351, 4003308, 4004574, 4100502, 4005136, 4101360, 4006408, 4001772, 4007104, 4001121, 4002097, 4002667, 4004442, 4000165, 4000572, 4100324, 4100413, 4006822, 4100501, 4100033, 4001781, 4006357, 4005223, 4002161, 4004195, 4100764, 4102509, 4101006, 4101957, 4100514, 1907871, 4005164, 4000735, 4100859, 4102319, 4004822, 4000067, 4002132, 4101063, 4101410, 4001779, 4004848, 4000434, 4005142, 4005220, 4001739, 4003346, 4004881, 1909084, 4004198, 4006933, 4100820, 4002764, 4005376, 4001744, 4100325, 4101028, 1909374, 4102322, 4004199, 4001120, 4000624, 4000155, 4002666, 4003313, 4102412, 4002621, 4002923, 4005905, 4101876, 4006361, 4006589, 4006606, 4006626, 4005809, 4100487, 4002996, 4003151, 4004884, 4101415, 4003015, 4101832, 4003154, 4102771, 1911612, 4001559, 4001592, 4003480, 4003495, 4004440, 4005390, 4005439, 1907512, 4101393, 4004041, 4002628, 4007041, 4007090, 4007101, 4101723, 4001591, 4003678, 4000349, 4102947, 4006821, 4000581, 1895793, 4000151, 4000214, 4002991, 4006973, 1906233, 4001098, 4001117, 4002662, 4002691, 4003526, 4006617, 4100074, 4102041, 1909806, 4004281, 4005883, 4000210, 4101185, 4101628, 4005926, 4000630, 4003530, 4005375, 4005802, 4007078, 4005448, 4000207, 4002838, 4005965, 4006303, 4006601, 4007077, 4100132, 4100441, 4100464, 4100706, 4100731, 4100927, 4100934, 4101133, 4101146, 4101157, 4101350, 4101641, 4101805, 4101824, 4101837, 4101866, 4102171, 4102172, 4102210, 4102212, 4102662, 4103006, 4001362, 4004004, 4004240, 4001359, 4002331, 4003419, 4003422, 4003459, 4003460, 4003541, 4003819, 4004566, 4004856, 4005263, 4005915, 4100077, 4100082, 4100961, 4101911, 4101964, 4102037, 4102038, 4102324, 4102573, 4102574, 4102617, 4102992, 4101839, 4006009, 4002733, 4007001, 4101397, 4102853, 4102968, 4102970, 4100547, 4102032, 4102060, 4101462, 4101188, 4004397, 4101025, 4102865, 4102869, 4102811, 4002912, 4003681, 4102508, 4102280, 4101200, 4103054, 4001489, 4101179, 4101472, 4102720, 4102410, 4102854, 4102407, 4102316, 4000768, 4001972, 4002946, 4003009, 4003018, 4003509, 4003529, 4003560, 4003600, 4004452, 4004571, 4004668, 4005048, 4005339, 4005949, 4005954, 4006927, 4006939, 4007002, 4007067, 4007084, 4100042, 4100078, 4100217, 4100695, 4100777, 4100823, 4100928, 4100931, 4100933, 4101134, 4101182, 4101223, 4101356, 4101468, 4101600, 4101635, 4101767, 4101962, 4102035, 4102117, 4102274, 1908767, 191021

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0435-2026

Related Recalls