Aptio Automation System with the Aliquoter Module (Inpeco P/N FLX-212) The FlexLab (aka Accelerator a3600 and Aptio Automation) is a modular system designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. The systems consolidate multiple Analytical instruments into a unified workstation.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    The impacted modules are the Aliquoter Modules (lnpeco Part Number FLX-212) with the firmware versions listed below or higher: - AQMb_3-3-0.H86 - AQMa_3-1-1-8.H86 and AQMb_3-1-1-8.H86 - xAQMb_1-1-0.elf; System Serial Numbers: AP2.0005 AP2.0008 AP2.0010 AP2.0014 AP2.0017 AP2.0018 AP2.0022 AP2.0023 AP2.0026 AP2.0027 AP2.0033 AP2.0036 AP2.0039 AP2.0043 AP2.0045 AP2.0056 AP2.0060 AP2.0065 AP2.0069 AP2.0070 AP2.0071 AP2.0076 AP2.0078 AP2.0081 AP2.0087 AP2.0089 AP2.0102 AP2.0104 AP2.0105 AP2.0110 AP2.0111 AP2.0113 AP2.0120 AP2.0127 AP2.0137 AP2.0139 AP2.0146 AP2.0147 AP2.0149 AP2.0151 AP2.0155 AP2.0165 AP2.0170 AP2.0172 AP2.0184 AP2.0189 AP2.0201 AP2.0202 AP2.0203 AP2.0209 AP2.0211 AP2.0213 AP2.0217 AP2.0223 APT.0009 APT.0014 APT.0058 APT.0065 APT.0093 APT.0091 APT.0125 AP2.0225 AP2.0230 AP2.0232 AP2.0233 AP2.0240 AP2.0242 AP2.0253 AP2.0266 AP2.0265 AP2.0268 AP2.0269 AP2.0238 AP2.0248 AP2.0274 AP2.0283 AP2.0289 AP2.0285 AP2.0295 AP2.0296 AP2.0303 AP2.0308 AP2.0332 AP2.0350 AP2.0353 AP2.0352 AP2.0354 AP2.0346 AP2.0361 AP2.0366 AP2.0388 AP2.0393 AP2.0402 AP2.0403 AP2.0404 AP2.0405 AP2.0389 AP2.0390 AP2.0400 AP2.0401 AP2.0421 AP2.0420 AP2.0427 AP2.0412
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Inpeco S.A.
Reason for Recall:
When a Clot Detection Error is generated during the sample aspiration from the Primary Sample Tube the aspirated volume is dispensed into the first empty Secondary Sample Tube. This Secondary Sample Tube is flagged with error 2132 or 1442 and sent to IOM Priority Output Racks to be manually managed. The current error message associated to error 2132 or 1442 recommends that the operator manages these Secondary Sample Tubes according to Laboratory Practice, but it does not clarify that these Secondary Sample Tubes may be diluted with water from the hydraulic circuit of the Aliquoter Module this may lead to reporting of erroneous results if the diluted sample is used for testing.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Aptio Automation System with the Aliquoter Module (Inpeco P/N FLX-212) The FlexLab (aka Accelerator a3600 and Aptio Automation) is a modular system designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. The systems consolidate multiple Analytical instruments into a unified workstation.

Product Codes/Lot Numbers:

The impacted modules are the Aliquoter Modules (lnpeco Part Number FLX-212) with the firmware versions listed below or higher: - AQMb_3-3-0.H86 - AQMa_3-1-1-8.H86 and AQMb_3-1-1-8.H86 - xAQMb_1-1-0.elf; System Serial Numbers: AP2.0005 AP2.0008 AP2.0010 AP2.0014 AP2.0017 AP2.0018 AP2.0022 AP2.0023 AP2.0026 AP2.0027 AP2.0033 AP2.0036 AP2.0039 AP2.0043 AP2.0045 AP2.0056 AP2.0060 AP2.0065 AP2.0069 AP2.0070 AP2.0071 AP2.0076 AP2.0078 AP2.0081 AP2.0087 AP2.0089 AP2.0102 AP2.0104 AP2.0105 AP2.0110 AP2.0111 AP2.0113 AP2.0120 AP2.0127 AP2.0137 AP2.0139 AP2.0146 AP2.0147 AP2.0149 AP2.0151 AP2.0155 AP2.0165 AP2.0170 AP2.0172 AP2.0184 AP2.0189 AP2.0201 AP2.0202 AP2.0203 AP2.0209 AP2.0211 AP2.0213 AP2.0217 AP2.0223 APT.0009 APT.0014 APT.0058 APT.0065 APT.0093 APT.0091 APT.0125 AP2.0225 AP2.0230 AP2.0232 AP2.0233 AP2.0240 AP2.0242 AP2.0253 AP2.0266 AP2.0265 AP2.0268 AP2.0269 AP2.0238 AP2.0248 AP2.0274 AP2.0283 AP2.0289 AP2.0285 AP2.0295 AP2.0296 AP2.0303 AP2.0308 AP2.0332 AP2.0350 AP2.0353 AP2.0352 AP2.0354 AP2.0346 AP2.0361 AP2.0366 AP2.0388 AP2.0393 AP2.0402 AP2.0403 AP2.0404 AP2.0405 AP2.0389 AP2.0390 AP2.0400 AP2.0401 AP2.0421 AP2.0420 AP2.0427 AP2.0412

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0434-2021

Related Recalls

Class I - Dangerous

The Aliquoter Module (mALX) may generate secondary sample tubes without labels or with mismatched labels which may result in delayed results or erroneous results, this error occurs if the operators open the printer head of mALX to perform error recoveries or maintenance procedures and do not empty the printer tubes buffer to remove all the secondary sample tubes

Oct 30, 2025 Other Medical Devices View Details →
Class I - Dangerous

The Aliquoter Module (mALX) may generate secondary sample tubes without labels or with mismatched labels which may result in delayed results or erroneous results. This error occurs if the operators open the printer head of mALX to perform error recoveries or maintenance procedures and do not empty the printer tubes buffer to remove all the secondary sample tubes.

Oct 30, 2025 Other Medical Devices View Details →

The FlexLab X System Sample Integrity Module (SIM) includes function that determines the hemolysis, icterus, and lipemia (HIL) in blood samples. This function has not been evaluated for safety and effectiveness by the FDA, so its use may cause an increase in erroneous results or a delay in patient results.