FlexLab (FLX) Automation System with the Aliquoter Module (Inpeco P/N FLX-212) The FlexLab (aka Accelerator a3600 and Aptio Automation) is a modular system designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. The systems consolidate multiple Analytical instruments into a unified workstation.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    The impacted modules are the Aliquoter Modules (lnpeco Part Number FLX-212) with the firmware versions listed below or higher: - AQMb_3-3-0.H86 - AQMa_3-1-1-8.H86 and AQMb_3-1-1-8.H86 - xAQMb_1-1-0.elf; System Serial Numbers: FLX.008 FLX.0118 FLX.064 FLX.101 FLX.0116 FLX.0125 FLX.080 FLX.0206 FLX.0161 FLX.0152 FLX.012 FLX.0121 FLX.0155 FLX.033 FLX.047 FLX.0127 FLX.0128 FLX.0129 FLX.074 FLX.076 FLX.0113 FLX.0138 FLX.014 FLX.015 FLX.0164 FLX.051 FLX.0106 FLX.0158 FLX.0196 FLX.072 FLX.0107 FLX.0189 FLX.0126 FLX.0165 FLX.0135 FLX.017 FLX.0114 FLX.022 FLX.0119 FLX.0105 FLX.029 FLX.0115 FLX.031 FLX.032 FLX.0141 FLX.037 FLX.020 FLX.039 FLX.043 FLX.102 FLX.046 FLX.0143 FLX.050 FLX.0124 FLX.0131 FLX.056 FLX.060 FLX.0136 FLX.0140 FLX.018 FLX.021 FLX.023 FLX.030 FLX.044 FLX.052 FLX.082 FLX.083 FLX.093 FLX.071 FLX.077 TSH.003 FLX.0146 FLX.027 FLX.038 FLX.075 FLX.0160 FLX.0153 FLX.0154 FLX.0157 FLX.0166 FLX.0167 FLX.0170 FLX.0174 FLX.0179 FLX.0180 FLX.0182 FLX.0194 FLX.0183 FLX.0048 FLX.0049 FLX.0109 FLX.0145 FLX.0148
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Inpeco S.A.
Reason for Recall:
When a Clot Detection Error is generated during the sample aspiration from the Primary Sample Tube the aspirated volume is dispensed into the first empty Secondary Sample Tube. This Secondary Sample Tube is flagged with error 2132 or 1442 and sent to IOM Priority Output Racks to be manually managed. The current error message associated to error 2132 or 1442 recommends that the operator manages these Secondary Sample Tubes according to Laboratory Practice, but it does not clarify that these Secondary Sample Tubes may be diluted with water from the hydraulic circuit of the Aliquoter Module this may lead to reporting of erroneous results if the diluted sample is used for testing.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

FlexLab (FLX) Automation System with the Aliquoter Module (Inpeco P/N FLX-212) The FlexLab (aka Accelerator a3600 and Aptio Automation) is a modular system designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. The systems consolidate multiple Analytical instruments into a unified workstation.

Product Codes/Lot Numbers:

The impacted modules are the Aliquoter Modules (lnpeco Part Number FLX-212) with the firmware versions listed below or higher: - AQMb_3-3-0.H86 - AQMa_3-1-1-8.H86 and AQMb_3-1-1-8.H86 - xAQMb_1-1-0.elf; System Serial Numbers: FLX.008 FLX.0118 FLX.064 FLX.101 FLX.0116 FLX.0125 FLX.080 FLX.0206 FLX.0161 FLX.0152 FLX.012 FLX.0121 FLX.0155 FLX.033 FLX.047 FLX.0127 FLX.0128 FLX.0129 FLX.074 FLX.076 FLX.0113 FLX.0138 FLX.014 FLX.015 FLX.0164 FLX.051 FLX.0106 FLX.0158 FLX.0196 FLX.072 FLX.0107 FLX.0189 FLX.0126 FLX.0165 FLX.0135 FLX.017 FLX.0114 FLX.022 FLX.0119 FLX.0105 FLX.029 FLX.0115 FLX.031 FLX.032 FLX.0141 FLX.037 FLX.020 FLX.039 FLX.043 FLX.102 FLX.046 FLX.0143 FLX.050 FLX.0124 FLX.0131 FLX.056 FLX.060 FLX.0136 FLX.0140 FLX.018 FLX.021 FLX.023 FLX.030 FLX.044 FLX.052 FLX.082 FLX.083 FLX.093 FLX.071 FLX.077 TSH.003 FLX.0146 FLX.027 FLX.038 FLX.075 FLX.0160 FLX.0153 FLX.0154 FLX.0157 FLX.0166 FLX.0167 FLX.0170 FLX.0174 FLX.0179 FLX.0180 FLX.0182 FLX.0194 FLX.0183 FLX.0048 FLX.0049 FLX.0109 FLX.0145 FLX.0148

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0432-2021

Related Recalls

Class I - Dangerous

The Aliquoter Module (mALX) may generate secondary sample tubes without labels or with mismatched labels which may result in delayed results or erroneous results, this error occurs if the operators open the printer head of mALX to perform error recoveries or maintenance procedures and do not empty the printer tubes buffer to remove all the secondary sample tubes

Oct 30, 2025 Other Medical Devices View Details →
Class I - Dangerous

The Aliquoter Module (mALX) may generate secondary sample tubes without labels or with mismatched labels which may result in delayed results or erroneous results. This error occurs if the operators open the printer head of mALX to perform error recoveries or maintenance procedures and do not empty the printer tubes buffer to remove all the secondary sample tubes.

Oct 30, 2025 Other Medical Devices View Details →

The FlexLab X System Sample Integrity Module (SIM) includes function that determines the hemolysis, icterus, and lipemia (HIL) in blood samples. This function has not been evaluated for safety and effectiveness by the FDA, so its use may cause an increase in erroneous results or a delay in patient results.

Sep 3, 2025 Diagnostic Equipment View Details →