LineSider Spinal System 5.5/6.0mm, Non-mod Standard 7.1mm - Dual Lead, Cannulated. Thoracolumbosacral pedicle screw system; Spinal interlaminal fixation orthosis. Product sold as non-sterile and shipped in protective sterilization trays. Screws can also be shipped individually, as tray replenishments. Labeled as the following sizes: a. 5.5mm x 35mm; b. 5.5mm x 40mm; c. 5.5mm x 45mm; d. 5.5mm x 50mm; e. 5.5mm x 55mm; f. 6.5mm x 35mm; g. 6.5mm x 40mm; h. 6.5mm x 45mm; i. 6.5mm x 50mm; j. 6.5mm x 55mm; k. 6.5mm x 60mm; l. 7.5mm x 35mm; m. 7.5mm x 40mm; n. 7.5mm x 45mm; o. 7.5mm x 50mm; p. 7.5mm x 55mm; q. 7.5mm x 60mm; r. 8.5mm x 35mm; s. 8.5mm x 40mm; t. 8.5mm x 45mm; u. 8.5mm x 50mm; v.8.5mm x 55mm; w. 8.5mm x 60mm; Used in spinal fixation surgery.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    a. 5.5mm x 35mm, Part Number: LS-N6SG5535, UDI: (01)00818613022568(10)02210139, lot Number: 2210139. b. 5.5mm x 40mm, Part Number: LS-N6SG5540, UDI: (01)00818613022575(10)02210140, Lot Number: 2210140. c. 5.5mm x 45mm, Part Number: LS-N6SG5545, UDI: ((01)00818613022582(10)02210141, Lot Number: 2210141. d. 5.5mm x 50mm, Part Number: LS-N6SG5550, UDI: (01)00818613022599(10)02210142, Lot Number: 2210142. e. 5.5mm x 55mm, Part Number: LS-N6SG5555, UDI: (01)00818613022605(10)02210143, Lot Number: 2210143. f. 6.5mm x 35mm, Part Number: LS-N6SG6535, UDI: (01)00818613022643(10)02210144, Lot Number: 2210144. g. 6.5mm x 40mm, Part Number: LS-N6SG6540, UDI: (01)00818613022650(10)02210145, Lot Number: 2210145. h. 6.5mm x 45mm, Part Number: LS-N6SG6545, UDI: (01)00818613022667(10)02210146, Lot Number: 2210146, UDI: (01)00818613022667(10)02210146-1, Lot Number: 02210146-1. i. 6.5mm x 50mm, Part Number: LS-N6SG6550, UDI: (01)00818613022674(10)02210147, Lot Number: 2210147. j. 6.5mm x 55mm, Part Number: LS-N6SG6555, UDI: (01)00818613022681(10)02210148, Lot Number: 2210148. k. 6.5mm x 60mm, Part Number: LS-N6SG6560, UDI: (01)00818613022698(10)02210149, Lot Number: 2210149. l. 7.5mm x 35mm, Part Number: LS-N6SG7535, UDI: (01)00818613022711(10)02210150, Lot Number: 2210150. m. 7.5mm x 40mm, Part Number: LS-N6SG7540, UDI: (01)00818613022735(10)02210152, Lot Number: 2210151. n. 7.5mm x 45mm, Part Number: LS-N6SG7545, UDI: (01)00818613027891(10)02210131, Lot Number: 2210152. o. 7.5mm x 50mm, Part Number: LS-N6SG7550, UDI: (01)00818613022742(10)02210153, Lot Number: 2210153. p. 7.5mm x 55mm, Part Number: LS-N6SG7555, UDI: (01)00818613022759(10)02210154, Lot Number: 2210154. q. 7.5mm x 60mm, Part Number: LS-N6SG7560, UDI: (01)00818613022766(10)02210155, Lot Number: 2210155. r. 8.5mm x 35mm, Part Number: LS-N6SG8535, UDI: (01)00818613022780(10)02210156, Lot Number: 2210156. s. 8.5mm x 40mm, Part Number: LS-N6SG8540, UDI: (01)00818613022797(10)02210157, Lot Number: 2210157. t. 8.5mm x 45mm, Part Number: LS-N6SG8545, UDI: (01)00818613022803(10)02210158, Lot Number: 2210158. u. 8.5mm x 50mm, Part Number: LS-N6SG8550, UDI: (01)00818613022810(10)02210159, Lot Number: 2210159. v.8.5mm x 55mm, Part Number: LS-N6SG8555, UDI: (01)00818613022827(10)02210160, Lot Number: 2210160. w. 8.5mm x 60mm, Part Number: LS-N6SG8560, UDI: (01)00818613022834(10)02210161, Lot Number: 2210161.
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Integrity Implants Inc.
Reason for Recall:
Potential for weld separation of the Upper Tulip Head component from the Lower Tulip Head component, which would result in the separation of the Tulip Head from the Screw Shank.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

LineSider Spinal System 5.5/6.0mm, Non-mod Standard 7.1mm - Dual Lead, Cannulated. Thoracolumbosacral pedicle screw system; Spinal interlaminal fixation orthosis. Product sold as non-sterile and shipped in protective sterilization trays. Screws can also be shipped individually, as tray replenishments. Labeled as the following sizes: a. 5.5mm x 35mm; b. 5.5mm x 40mm; c. 5.5mm x 45mm; d. 5.5mm x 50mm; e. 5.5mm x 55mm; f. 6.5mm x 35mm; g. 6.5mm x 40mm; h. 6.5mm x 45mm; i. 6.5mm x 50mm; j. 6.5mm x 55mm; k. 6.5mm x 60mm; l. 7.5mm x 35mm; m. 7.5mm x 40mm; n. 7.5mm x 45mm; o. 7.5mm x 50mm; p. 7.5mm x 55mm; q. 7.5mm x 60mm; r. 8.5mm x 35mm; s. 8.5mm x 40mm; t. 8.5mm x 45mm; u. 8.5mm x 50mm; v.8.5mm x 55mm; w. 8.5mm x 60mm; Used in spinal fixation surgery.

Product Codes/Lot Numbers:

a. 5.5mm x 35mm, Part Number: LS-N6SG5535, UDI: (01)00818613022568(10)02210139, lot Number: 2210139. b. 5.5mm x 40mm, Part Number: LS-N6SG5540, UDI: (01)00818613022575(10)02210140, Lot Number: 2210140. c. 5.5mm x 45mm, Part Number: LS-N6SG5545, UDI: ((01)00818613022582(10)02210141, Lot Number: 2210141. d. 5.5mm x 50mm, Part Number: LS-N6SG5550, UDI: (01)00818613022599(10)02210142, Lot Number: 2210142. e. 5.5mm x 55mm, Part Number: LS-N6SG5555, UDI: (01)00818613022605(10)02210143, Lot Number: 2210143. f. 6.5mm x 35mm, Part Number: LS-N6SG6535, UDI: (01)00818613022643(10)02210144, Lot Number: 2210144. g. 6.5mm x 40mm, Part Number: LS-N6SG6540, UDI: (01)00818613022650(10)02210145, Lot Number: 2210145. h. 6.5mm x 45mm, Part Number: LS-N6SG6545, UDI: (01)00818613022667(10)02210146, Lot Number: 2210146, UDI: (01)00818613022667(10)02210146-1, Lot Number: 02210146-1. i. 6.5mm x 50mm, Part Number: LS-N6SG6550, UDI: (01)00818613022674(10)02210147, Lot Number: 2210147. j. 6.5mm x 55mm, Part Number: LS-N6SG6555, UDI: (01)00818613022681(10)02210148, Lot Number: 2210148. k. 6.5mm x 60mm, Part Number: LS-N6SG6560, UDI: (01)00818613022698(10)02210149, Lot Number: 2210149. l. 7.5mm x 35mm, Part Number: LS-N6SG7535, UDI: (01)00818613022711(10)02210150, Lot Number: 2210150. m. 7.5mm x 40mm, Part Number: LS-N6SG7540, UDI: (01)00818613022735(10)02210152, Lot Number: 2210151. n. 7.5mm x 45mm, Part Number: LS-N6SG7545, UDI: (01)00818613027891(10)02210131, Lot Number: 2210152. o. 7.5mm x 50mm, Part Number: LS-N6SG7550, UDI: (01)00818613022742(10)02210153, Lot Number: 2210153. p. 7.5mm x 55mm, Part Number: LS-N6SG7555, UDI: (01)00818613022759(10)02210154, Lot Number: 2210154. q. 7.5mm x 60mm, Part Number: LS-N6SG7560, UDI: (01)00818613022766(10)02210155, Lot Number: 2210155. r. 8.5mm x 35mm, Part Number: LS-N6SG8535, UDI: (01)00818613022780(10)02210156, Lot Number: 2210156. s. 8.5mm x 40mm, Part Number: LS-N6SG8540, UDI: (01)00818613022797(10)02210157, Lot Number: 2210157. t. 8.5mm x 45mm, Part Number: LS-N6SG8545, UDI: (01)00818613022803(10)02210158, Lot Number: 2210158. u. 8.5mm x 50mm, Part Number: LS-N6SG8550, UDI: (01)00818613022810(10)02210159, Lot Number: 2210159. v.8.5mm x 55mm, Part Number: LS-N6SG8555, UDI: (01)00818613022827(10)02210160, Lot Number: 2210160. w. 8.5mm x 60mm, Part Number: LS-N6SG8560, UDI: (01)00818613022834(10)02210161, Lot Number: 2210161.

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0419-2023

Related Recalls

Potential for weld separation of the Upper Tulip Head component from the Lower Tulip Head component, which would result in the separation of the Tulip Head from the Screw Shank.

Jan 27, 2022 Implants & Prosthetics Nationwide View Details →

Potential for intraoperative screw head separation at weld location.

Aug 3, 2022 Implants & Prosthetics View Details →