MA3320-PM Premium Adult - SYK Mistral-Air Premium Blankets. MODEL #: MA3302-PM. CATALOG #(s): MA3302-PM
Class I - DangerousWhat Should You Do?
- Check if you have this product: LOT #(s): 1640002, 1645003
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Stryker Medical Division of Stryker Corporation
- Reason for Recall:
- It was identified that the Mistral-Air Premium blankets may be capable of conducting electricity during a defibrillation or cardioversion therapy at higher energy levels (70 Joules or more) when the blanket is applied with the reflective side toward the patient (upside down).
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
MA3320-PM Premium Adult - SYK Mistral-Air Premium Blankets. MODEL #: MA3302-PM. CATALOG #(s): MA3302-PM
Product Codes/Lot Numbers:
LOT #(s): 1640002, 1645003
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0419-2018
Related Recalls
Cub Pediatric Crib, Model FL19H
Stryker Medical Division of Stryker
Cribs sold in the USA are missing two access door warning labels.
Stryker Arise 1000EX mattress, Part Number 2236000000
Stryker Medical Division of Stryker
Stryker has issued a voluntary recall of certain MV3 beds and Arise 1000EX mattresses because some of these products have been sold in close proximity, suggesting a potential intent for them to be used together. These products have not been tested for compatibility and thus are not designed to be used together. The combination of both is not recommended by Stryker.
Stryker MV3 bariatric bed, Part Number 5900000001
Stryker Medical Division of Stryker
Stryker has issued a voluntary recall of certain MV3 beds and Arise 1000EX mattresses because some of these products have been sold in close proximity, suggesting a potential intent for them to be used together. These products have not been tested for compatibility and thus are not designed to be used together. The combination of both is not recommended by Stryker.