Triage Drugs of Abuse Plus TCA 25 Test Kit, Model Numbers 92000, 92590, 92591; 25 individually pouched devices, 1 bottle of Wash Solution, 1 Pipette and 50 Pipette Tips in a labeled kit box. Product Usage: The Alere Triage Drugs of Abuse Panel plus Tricyclic Antidepressants uses eight discrete competitive immunoassay procedures for the simultaneous detection and identification of the major drugs and/or the urinary metabolites of eight different drug classes. The use of monoclonal antibodies that are specific for the drugs and/or metabolites of the eight drug classes ensures a high degree of sensitivity and specificity.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lots 403111, 406741, 396588, 398577, 396587
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Alere San Diego, Inc.
- Reason for Recall:
- The recalled lots have demonstrated unexpected false positive THC results.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- completed
Product Information
Full Description:
Triage Drugs of Abuse Plus TCA 25 Test Kit, Model Numbers 92000, 92590, 92591; 25 individually pouched devices, 1 bottle of Wash Solution, 1 Pipette and 50 Pipette Tips in a labeled kit box. Product Usage: The Alere Triage Drugs of Abuse Panel plus Tricyclic Antidepressants uses eight discrete competitive immunoassay procedures for the simultaneous detection and identification of the major drugs and/or the urinary metabolites of eight different drug classes. The use of monoclonal antibodies that are specific for the drugs and/or metabolites of the eight drug classes ensures a high degree of sensitivity and specificity.
Product Codes/Lot Numbers:
Lots 403111, 406741, 396588, 398577, 396587
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0418-2018
Related Recalls
There is a potential that the Cholestech LDX battery (an external battery pack available as an optional accessory to the Cholestech LDX Analyzer) may swell which could cause minor burns, cuts or abrasions, or low current electrical shock to the user.
Cholesterol test cassettes package insert is to be updated with new bilirubin interference information because it was found that samples containing bilirubin, at concentrations greater than 2.0 mg/dL (conjugated) or 1.6 mg/dL (unconjugated), can result in low recovery of Triglyceride (TRG) and High-Density Lipoproteins (HDL).
The polarity of the output of power supply was reversed. Due to polarity of the output of power supplies being reversed, it renders the power supplies nonfunctional and the analyzer inoperable.