Thermo scientific MAS Omni CARDIO, Liquid Assayed Integrated Cardiac Control, Model/Catalog Number: OCRD-L, OCRD-101, OCRD-202, OCRD-303, OCRD-SP, OCRD-MP, OCRD-UL; Product Description: In Vitro Diagnostic external control

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    OCRD-UL: UDI/DI 00884883011946, Lot Number OCRD2704U, Exp 2027-04-30; OCRD-L: UDI/DI 00884883011953, Lot Number OCRD2704L, Exp 2027-04-30; OCRD-101: UDI/DI 00884883011960, Lot Number OCRD27041, Exp 2027-04-30; OCRD-202: UDI/DI 00884883011977, Lot Number OCRD27042, Exp 2027-04-30; OCRD-303: UDI/DI 00884883011984, Lot Number OCRD27043, Exp 2027-04-30; OCRD-SP: UDI/DI 00884883011991, Lot Number OCRD2704S, Exp 2027-04-30; OCRD-MP: UDI/DI 00884883012004, Lot Number OCRD2704M, Exp 2027-04-30.
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Microgenics Corporation
Reason for Recall:
The firm received complaints from customers reporting vial-to-vial variability when using the control with high sensitivity Troponin I assays. This defect may lead to as delay in patient results for hs Troponin I assays.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Thermo scientific MAS Omni CARDIO, Liquid Assayed Integrated Cardiac Control, Model/Catalog Number: OCRD-L, OCRD-101, OCRD-202, OCRD-303, OCRD-SP, OCRD-MP, OCRD-UL; Product Description: In Vitro Diagnostic external control

Product Codes/Lot Numbers:

OCRD-UL: UDI/DI 00884883011946, Lot Number OCRD2704U, Exp 2027-04-30; OCRD-L: UDI/DI 00884883011953, Lot Number OCRD2704L, Exp 2027-04-30; OCRD-101: UDI/DI 00884883011960, Lot Number OCRD27041, Exp 2027-04-30; OCRD-202: UDI/DI 00884883011977, Lot Number OCRD27042, Exp 2027-04-30; OCRD-303: UDI/DI 00884883011984, Lot Number OCRD27043, Exp 2027-04-30; OCRD-SP: UDI/DI 00884883011991, Lot Number OCRD2704S, Exp 2027-04-30; OCRD-MP: UDI/DI 00884883012004, Lot Number OCRD2704M, Exp 2027-04-30.

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0415-2026

Related Recalls

Due to QMS Tacrolimus Calibrator recovering between 82 to 94% of their target values, resulting in over-quantitation in tacrolimus patient result values when used with the QMS Tacrolimus Immunoassay.

Jan 20, 2022 Other Medical Devices Nationwide View Details →

A reagent stability issue (still under investigation) causes less separation in the calibration curve, which affects results across the entire range. For controls, maximum bias of +/- 35% for low levels, +/- 33% for mid-range levels, and +/- 47% for high levels is observed. Bias in patient samples ranges from -15.7% to +5.2% with an average bias of -5.6% when comparing the QMS everolimus assay to an everolimus LC-MS reference method. Patient samples across the measuring range  sub-therapeutic, therapeutic, and supra-therapeutic  could be affected.

Aug 15, 2016 Diagnostic Equipment Nationwide View Details →