Good Neighbor Pharmacy, TRUE METRIX Self Monitoring Blood Glucose Meter. Model/Catalog Number: RE4011-01.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Model/Catalog Number: RE4011-01. UDI-DI (Meter Kit): 00087701426254. UDI-DI (Meter): 00021292007690. Serial Numbers: T13908301, T13908302, T13908303, T13908304, T13908305, T13908306, T13908307, T13908308, T13908309, T13908310, T13908311, T13908312, T13908313, T13908314, T13908315, T13908316, T13908317, T13908318, T13908319, T13908320, T13908321, T13908322, T13908323, T13908324, T13908325, T13908326, T13908327, T13908328, T13908329, T13908330, T13908331, T13908332, T13908333, T13908335, T13908336, T13908337, T13908338, T13908339, T13908340, T13908341, T13908342, T13908343, T13908344, T13908345, T13908346, T13908347, T13908348, T13908349, T13908350, T13908351, T13908352, T13908353, T13908354, T13908355, T13908356, T13908357, T13908358, T13908359, T13908360, T13908361, T13908362, T13908363, T13908364, T13908365, T13908366, T13908367, T13908368, T13908369, T13908370, T13908371, T13908372, T13908373, T13908374, T13908375, T13908376, T13908377, T13908378, T13908379, T13908380, T13908381, T13908382, T13908383, T13908384, T13908385, T13908386, T13908387, T13908388, T13908389, T13908390, T13908391, T13908392, T13908393, T13908394, T13908395, T13908396, T13908397, T13908398, T13908399, T13908400, T13909501, T13909502, T13909503, T13909504, T13909505, T13909506, T13909507, T13909508, T13909509, T13909510, T13909511, T13909512, T13909513, T13909514, T13909515, T13909516, T13909517, T13909518, T13909519, T13909520, T13909521, T13909522, T13909523, T13909524, T13909525, T13909526, T13909527, T13909528, T13909529, T13909530, T13909531, T13909532, T13909533, T13909534, T13909535, T13909536, T13909537, T13909538, T13909539, T13909540, T13909541, T13909542, T13909543, T13909544, T13909545, T13909546, T13909547, T13909548, T13909549, T13909550, T13909551, T13909552, T13909553, T13909554, T13909555, T13909556, T13909557, T13909558, T13909559, T13909560, T13909561, T13909562, T13909563, T13909564, T13909565, T13909566, T13909567, T13909568, T13909569, T13909570, T13909571, T13909572, T13909573, T13909574, T13909575, T13909576, T13909577, T13909578, T13909579, T13909580, T13909581, T13909582, T13909583, T13909584, T13909586, T13909587, T13909588, T13909589, T13909590, T13909591, T13909592, T13909593, T13909594, T13909595, T13909596, T13909597, T13909598, T13909599, T13909600, T13909981, T13909982, T13909983, T13909984, T13909985, T13909987, T13909988, T13909989, T13909990, T13911201, T13911203, T13911204, T13911205, T13911206, T13911207, T13911208, T13911209, T13911210, T13911211, T13911212, T13911213, T13911214, T13911215, T13911216, T13911217, T13911219, T13911220, T13911221, T13911222, T13911223, T13911224, T13911225, T13911226, T13911227, T13911228, T13911229, T13911230, T13911231, T13911232, T13911233, T13911236, T13911237, T13911238, T13911239, T13911240, T13911241, T13911242, T13911243, T13911244, T13911245, T13911246, T13911247, T13911248, T13911249, T13911250, T13911251, T13911252, T13911253, T13911254, T13911255, T13911256, T13911257, T13911258, T13911259, T13911260, T13911261, T13911262, T13911263, T13911264, T13911265, T13911266, T13911267, T13911268, T13911269, T13911270, T13911271, T13911272, T13911273, T13911274, T13911275, T13911276, T13911277, T13911278, T13911279, T13911280, T13911281, T13911282, T13911283, T13911284, T13911285, T13911286, T13911287, T13911288, T13911289, T13911290, T13911291, T13911292, T13911293, T13911294, T13911295, T13911296, T13911297, T13911298, T13911299, T13911300, T13911501, T13911502, T13911503, T13911504, T13911505, T13911506, T13911507, T13911508, T13911509, T13911510, T13911511, T13911512, T13911513, T13911514, T13911515, T13911516, T13911517, T13911518, T13911519, T13911520, T13911521, T13911522, T13911523, T13911524, T13911525, T13911526, T13911527, T13911528, T13911529, T13911530, T13911531, T13911532, T13911533, T13911534, T13911535, T13911536, T13911537, T13911538, T13911539, T13911540, T13911541, T13911542, T13911543, T13911544, T13911545, T13911546, T13911547, T13911548, T13911549, T13911550, T13911551, T13911552, T13911553, T13911554, T13911555, T13911556, T13911557, T13911558, T13911559, T13911560, T13911561, T13911562, T13911563, T13911564, T13911565, T13911566, T13911567, T13911568, T13911569, T13911570, T13911571, T13911572, T13911573, T13911574, T13911575, T13911576, T13911577, T13911578, T13911579, T13911580, T13911581, T13911582, T13911583, T13911584, T13911585, T13911586, T13911587, T13911588, T13911589, T13911590, T13911591, T13911592, T13911593, T13911594, T13911595, T13911596, T13911597, T13911598, T13911599, T13911600, T13912001, T13912002, T13912003, T13912004, T13912005, T13912006, T13912007, T13912008, T13912009, T13912010, T13912011, T13912012, T13912013, T13912014, T13912015, T13912016, T13912017, T13912018, T13912019, T13912020, T13912021, T13912022, T13912023, T13912024, T13912025, T13912026, T13912027, T13912028, T13912029, T13912030, T13912031, T13912032, T13912033, T13912034, T13912035, T13912036, T13912037, T13912038, T13912039, T13912040, T13912041, T13912042, T13912043, T13912044, T13912045, T13912046, T13912047, T13912048, T13912049, T13912050, T13912051, T13912052, T13912053, T13912054, T13912056, T13912057, T13912058, T13912059, T13912060, T13912061, T13912062, T13912063, T13912064, T13912065, T13912066, T13912067, T13912068, T13912069, T13912070, T13912071, T13912072, T13912073, T13912074, T13912075, T13912076, T13912077, T13912078, T13912079, T13912080, T13912081, T13912082, T13912083, T13912084, T13912085, T13912086, T13912087, T13912088, T13912089, T13912090, T13912091, T13912092, T13912093, T13912094, T13912095, T13912096, T13912097, T13912098, T13912099, T13912100, T13912201, T13912202, T13912203, T13912204, T13912205, T13912206, T13912207, T13912208, T13912209, T13912210, T13912211, T13912212, T13912213, T13912214, T13912215, T13912216, T13912217, T13912218, T13912219, T13912220, T13912221, T13912222, T13912223, T13912224, T13912225, T13912226, T13912227, T13912228, T13912229, T13912230, T13912231, T13912232, T13912233, T13912234, T13912235, T13912236, T13912237, T13912238, T13912239, T13912240, T13912241, T13912242, T13912244, T13912245, T13912246, T13912247, T13912248, T13912249, T13912250, T13912251, T13912252, T13912253, T13912254, T13912255, T13912256, T13912257, T13912258, T13912259, T13912260, T13912261, T13912262, T13912263, T13912264, T13912265, T13912266, T13912267, T13912268, T13912269, T13912270, T13912271, T13912272, T13912273, T13912274, T13912275, T13912276, T13912277, T13912278, T13912279, T13912280, T13912281, T13912282, T13912283, T13912284, T13912285, T13912286, T13912287, T13912288, T13912289, T13912290, T13912291, T13912292, T13912293, T13912294, T13912295, T13912296, T13912297, T13912298, T13912299, T13912300. Expiration Date: 02/21/2030.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Trividia Health, Inc.
- Reason for Recall:
- Manufacturing defect causing LCD display issues with missing or partial segments (ghosting) that can result in misreading glucose values, potentially leading to inappropriate treatment decisions for diabetes management.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Good Neighbor Pharmacy, TRUE METRIX Self Monitoring Blood Glucose Meter. Model/Catalog Number: RE4011-01.
Product Codes/Lot Numbers:
Model/Catalog Number: RE4011-01. UDI-DI (Meter Kit): 00087701426254. UDI-DI (Meter): 00021292007690. Serial Numbers: T13908301, T13908302, T13908303, T13908304, T13908305, T13908306, T13908307, T13908308, T13908309, T13908310, T13908311, T13908312, T13908313, T13908314, T13908315, T13908316, T13908317, T13908318, T13908319, T13908320, T13908321, T13908322, T13908323, T13908324, T13908325, T13908326, T13908327, T13908328, T13908329, T13908330, T13908331, T13908332, T13908333, T13908335, T13908336, T13908337, T13908338, T13908339, T13908340, T13908341, T13908342, T13908343, T13908344, T13908345, T13908346, T13908347, T13908348, T13908349, T13908350, T13908351, T13908352, T13908353, T13908354, T13908355, T13908356, T13908357, T13908358, T13908359, T13908360, T13908361, T13908362, T13908363, T13908364, T13908365, T13908366, T13908367, T13908368, T13908369, T13908370, T13908371, T13908372, T13908373, T13908374, T13908375, T13908376, T13908377, T13908378, T13908379, T13908380, T13908381, T13908382, T13908383, T13908384, T13908385, T13908386, T13908387, T13908388, T13908389, T13908390, T13908391, T13908392, T13908393, T13908394, T13908395, T13908396, T13908397, T13908398, T13908399, T13908400, T13909501, T13909502, T13909503, T13909504, T13909505, T13909506, T13909507, T13909508, T13909509, T13909510, T13909511, T13909512, T13909513, T13909514, T13909515, T13909516, T13909517, T13909518, T13909519, T13909520, T13909521, T13909522, T13909523, T13909524, T13909525, T13909526, T13909527, T13909528, T13909529, T13909530, T13909531, T13909532, T13909533, T13909534, T13909535, T13909536, T13909537, T13909538, T13909539, T13909540, T13909541, T13909542, T13909543, T13909544, T13909545, T13909546, T13909547, T13909548, T13909549, T13909550, T13909551, T13909552, T13909553, T13909554, T13909555, T13909556, T13909557, T13909558, T13909559, T13909560, T13909561, T13909562, T13909563, T13909564, T13909565, T13909566, T13909567, T13909568, T13909569, T13909570, T13909571, T13909572, T13909573, T13909574, T13909575, T13909576, T13909577, T13909578, T13909579, T13909580, T13909581, T13909582, T13909583, T13909584, T13909586, T13909587, T13909588, T13909589, T13909590, T13909591, T13909592, T13909593, T13909594, T13909595, T13909596, T13909597, T13909598, T13909599, T13909600, T13909981, T13909982, T13909983, T13909984, T13909985, T13909987, T13909988, T13909989, T13909990, T13911201, T13911203, T13911204, T13911205, T13911206, T13911207, T13911208, T13911209, T13911210, T13911211, T13911212, T13911213, T13911214, T13911215, T13911216, T13911217, T13911219, T13911220, T13911221, T13911222, T13911223, T13911224, T13911225, T13911226, T13911227, T13911228, T13911229, T13911230, T13911231, T13911232, T13911233, T13911236, T13911237, T13911238, T13911239, T13911240, T13911241, T13911242, T13911243, T13911244, T13911245, T13911246, T13911247, T13911248, T13911249, T13911250, T13911251, T13911252, T13911253, T13911254, T13911255, T13911256, T13911257, T13911258, T13911259, T13911260, T13911261, T13911262, T13911263, T13911264, T13911265, T13911266, T13911267, T13911268, T13911269, T13911270, T13911271, T13911272, T13911273, T13911274, T13911275, T13911276, T13911277, T13911278, T13911279, T13911280, T13911281, T13911282, T13911283, T13911284, T13911285, T13911286, T13911287, T13911288, T13911289, T13911290, T13911291, T13911292, T13911293, T13911294, T13911295, T13911296, T13911297, T13911298, T13911299, T13911300, T13911501, T13911502, T13911503, T13911504, T13911505, T13911506, T13911507, T13911508, T13911509, T13911510, T13911511, T13911512, T13911513, T13911514, T13911515, T13911516, T13911517, T13911518, T13911519, T13911520, T13911521, T13911522, T13911523, T13911524, T13911525, T13911526, T13911527, T13911528, T13911529, T13911530, T13911531, T13911532, T13911533, T13911534, T13911535, T13911536, T13911537, T13911538, T13911539, T13911540, T13911541, T13911542, T13911543, T13911544, T13911545, T13911546, T13911547, T13911548, T13911549, T13911550, T13911551, T13911552, T13911553, T13911554, T13911555, T13911556, T13911557, T13911558, T13911559, T13911560, T13911561, T13911562, T13911563, T13911564, T13911565, T13911566, T13911567, T13911568, T13911569, T13911570, T13911571, T13911572, T13911573, T13911574, T13911575, T13911576, T13911577, T13911578, T13911579, T13911580, T13911581, T13911582, T13911583, T13911584, T13911585, T13911586, T13911587, T13911588, T13911589, T13911590, T13911591, T13911592, T13911593, T13911594, T13911595, T13911596, T13911597, T13911598, T13911599, T13911600, T13912001, T13912002, T13912003, T13912004, T13912005, T13912006, T13912007, T13912008, T13912009, T13912010, T13912011, T13912012, T13912013, T13912014, T13912015, T13912016, T13912017, T13912018, T13912019, T13912020, T13912021, T13912022, T13912023, T13912024, T13912025, T13912026, T13912027, T13912028, T13912029, T13912030, T13912031, T13912032, T13912033, T13912034, T13912035, T13912036, T13912037, T13912038, T13912039, T13912040, T13912041, T13912042, T13912043, T13912044, T13912045, T13912046, T13912047, T13912048, T13912049, T13912050, T13912051, T13912052, T13912053, T13912054, T13912056, T13912057, T13912058, T13912059, T13912060, T13912061, T13912062, T13912063, T13912064, T13912065, T13912066, T13912067, T13912068, T13912069, T13912070, T13912071, T13912072, T13912073, T13912074, T13912075, T13912076, T13912077, T13912078, T13912079, T13912080, T13912081, T13912082, T13912083, T13912084, T13912085, T13912086, T13912087, T13912088, T13912089, T13912090, T13912091, T13912092, T13912093, T13912094, T13912095, T13912096, T13912097, T13912098, T13912099, T13912100, T13912201, T13912202, T13912203, T13912204, T13912205, T13912206, T13912207, T13912208, T13912209, T13912210, T13912211, T13912212, T13912213, T13912214, T13912215, T13912216, T13912217, T13912218, T13912219, T13912220, T13912221, T13912222, T13912223, T13912224, T13912225, T13912226, T13912227, T13912228, T13912229, T13912230, T13912231, T13912232, T13912233, T13912234, T13912235, T13912236, T13912237, T13912238, T13912239, T13912240, T13912241, T13912242, T13912244, T13912245, T13912246, T13912247, T13912248, T13912249, T13912250, T13912251, T13912252, T13912253, T13912254, T13912255, T13912256, T13912257, T13912258, T13912259, T13912260, T13912261, T13912262, T13912263, T13912264, T13912265, T13912266, T13912267, T13912268, T13912269, T13912270, T13912271, T13912272, T13912273, T13912274, T13912275, T13912276, T13912277, T13912278, T13912279, T13912280, T13912281, T13912282, T13912283, T13912284, T13912285, T13912286, T13912287, T13912288, T13912289, T13912290, T13912291, T13912292, T13912293, T13912294, T13912295, T13912296, T13912297, T13912298, T13912299, T13912300. Expiration Date: 02/21/2030.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0413-2026
Related Recalls
One (1) TRUE METRIX AIR blood glucose meter distributed in the United States was packaged into a TRUE METRIX blood glucose meter kit and has an incorrect factory-set unit of measure; the meter displays glucose results in mmol/L rather than mg/dL.
Product gives incorrect low blood glucose levels.
Product gives incorrect low blood glucose levels.