MEVION S250 Product Usage: Proton Radiation Therapy System
Class I - DangerousWhat Should You Do?
- Check if you have this product: Serial Number: S250-0001 through S250-0006
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Mevion Medical Systems, Inc.
- Reason for Recall:
- Recovery of Incorrect Isocenter -operational change
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
MEVION S250 Product Usage: Proton Radiation Therapy System
Product Codes/Lot Numbers:
Serial Number: S250-0001 through S250-0006
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0410-2017
Related Recalls
Mevion S250 and S250i Proton Therapy Systems hand pendant (P/N 10263 or P/N 37808)
Mevion Medical Systems
Malfunction of the hand pendant controlling movement of the therapy couch and nozzle may result in unintentional motion. Patient injury is possible if they come in contact with the nozzle or other objects in the treatment room.
Treatment beam information disappears on Treatment Console screen while beam delivery continues potential for harm to a patient could occur
MEVION S250 Proton Radiation Treatment System; intended to deliver proton radiation treatment to patients
Mevion Medical Systems
The perforated screen that divides the inside of the process water tank has a very sharp edge across the top of the Heat Exchanger leading to potential injury for servicer.