Multipolar¿ Bipolar Cup Liner 22 mm LD.; Item 00-5001-040-22 Product Usage: The liner snaps into the Multipolar¿ Bipolar Cup and serves as the articulation surface for the femoral head. The split poly locking ring helps contain the head within the liner and also provides a popping sound as conformation that the femoral head has completed seated, as noted in surgical technique 97-5003-003-00
Class I - DangerousWhat Should You Do?
- Check if you have this product: Item 00-5001-040- 22; Lot 61261192, 61261194, 61296091,
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Zimmer, Inc.
- Reason for Recall:
- Zimmer Inc. is initiating a removal of the Multipolar¿ Bipolar Cup Liner 22 mm J.D. (item 00-5001-040-22) due to reports that the split poly ring was incorrectly assembled.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Multipolar¿ Bipolar Cup Liner 22 mm LD.; Item 00-5001-040-22 Product Usage: The liner snaps into the Multipolar¿ Bipolar Cup and serves as the articulation surface for the femoral head. The split poly locking ring helps contain the head within the liner and also provides a popping sound as conformation that the femoral head has completed seated, as noted in surgical technique 97-5003-003-00
Product Codes/Lot Numbers:
Item 00-5001-040- 22; Lot 61261192, 61261194, 61296091,
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0410-2013
Related Recalls
Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.
Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.
Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.