Alinity m HCV AMP kit, List (REF) Numbers: a) 08N50-095 and b) 08N50-090. The Alinity m HCV assay is an in vitro reverse transcription-polymerase chain reaction (RT-PCR) assay for use with the automated Alinity m System to detect and quantitate hepatitis C virus (HCV) RNA in human serum or plasma.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    a) 08N50-095: UDI/DI 00884999047648, Lot Numbers: 381582, 382448, 383204; b) 08N50-090: UDI/DI 00884999047761, Lot Numbers: 383203, 385720, 381581, 382447
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Abbott Molecular, Inc.
Reason for Recall:
Abbott has received reports of an increase in reactive negative controls and false positive results in patient samples when using the Identified Lots of Alinity m HCV AMP Kit.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Alinity m HCV AMP kit, List (REF) Numbers: a) 08N50-095 and b) 08N50-090. The Alinity m HCV assay is an in vitro reverse transcription-polymerase chain reaction (RT-PCR) assay for use with the automated Alinity m System to detect and quantitate hepatitis C virus (HCV) RNA in human serum or plasma.

Product Codes/Lot Numbers:

a) 08N50-095: UDI/DI 00884999047648, Lot Numbers: 381582, 382448, 383204; b) 08N50-090: UDI/DI 00884999047761, Lot Numbers: 383203, 385720, 381581, 382447

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0407-2024

Related Recalls

There were reports of an increase in reactive negative controls and false positive results.

Apr 28, 2025 Implants & Prosthetics Nationwide View Details →

There were reports of an increase in reactive negative controls and false positive results.

Apr 28, 2025 Diagnostic Equipment Nationwide View Details →

The design of the Alinity m System Solutions drawer enclosure may allow liquid (liquid waste or system solutions) originating in the System Solutions drawer to flow beyond the footprint of the instrument and into the walking-path of the user.

Mar 20, 2025 Diagnostic Equipment Nationwide View Details →