Tina-Quant Hemoglobin A1c Gen. 2 Hemoglobin A1c test Catalog number 04528123190 / 05401640190

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Not applicable
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Roche Diagnostics Operations, Inc.
Reason for Recall:
Roche Diagnostics has confirmed elevated QC and patient sample recovery for individual Tina-quant¿ Hemoglobin A1c Gen.2 and Tina-quant Hemoglobin A1cDx Gen.2 cassettes/bottles. This issue occurred over multiple lots. Single cassettes/bottles of the reagents are affected. The issue can lead to erroneously high HbA1c results. If a patient with known diabetes is affected, an erroneously high HbA1c result might lead to therapeutic consequences, such as therapy escalation (introduction of another oral antidiabetic medication or insulin) or increase of the dosage. This can further lead to an increased risk of hypoglycemia. Considering the unreliable detectability of the issue, relevant medical risk for the patient cannot entirely be excluded.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Tina-Quant Hemoglobin A1c Gen. 2 Hemoglobin A1c test Catalog number 04528123190 / 05401640190

Product Codes/Lot Numbers:

Not applicable

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0395-2017

Related Recalls

Software defect, which allows the system to accept erroneous, non-monotonous calibrations for Spline-type assays. Cobas pro integrated solutions is an automated analyzer, intended for running qualitative, semi-quantitative and quantitative clinical chemistry and immunochemistry assays, as well as ion-selective measurements. When an erroneous calibration is active, the instrument fails to calculate new values and instead repeats the last successfully calculated result from any Spline-type assay for all subsequent measurements, leading to identical and erroneous patient and QC results. The defect could lead to patients receiving erroneous lab results, which could result in a remote risk of serious adverse health consequences. The following assays use spline type calibration and are affected by this issue: Cystatin C Gen.2, Ferritin Gen.4, Lipoprotein (a) Gen. 2, Lipoprotein (a) molarity, Vancomycin Gen.3, Kappa Free Light Chains Partner Channel, Lambda Free Light Chains Partner Channel, and fCAL turbo Partner Channel.

Jan 20, 2026 Other Medical Devices Nationwide View Details →