Carefusion V. Mueller GERALD BAYONET IRRIGATING BIPOLAR FORCEPS 1.0MM TIP OVERALL LENGTH 7-1/2" (190MM), REF F-1301
Class I - DangerousWhat Should You Do?
- Check if you have this product: UDI/DI 10885403040825, ALL LOTS
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Carefusion 2200 Inc
- Reason for Recall:
- BD has determined that the Instructions For Use (IFU) provided by BD is missing certain content that is contained in the manufacturer's IFU. This missing content is related to the interface with the device power supply, as well as cleaning and maintenance instructions. Refer to the attached V. Mueller" Bipolar Forcep IFU.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Carefusion V. Mueller GERALD BAYONET IRRIGATING BIPOLAR FORCEPS 1.0MM TIP OVERALL LENGTH 7-1/2" (190MM), REF F-1301
Product Codes/Lot Numbers:
UDI/DI 10885403040825, ALL LOTS
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0385-2023
Related Recalls
Snowden-Pencer MIS DIAMOND-TOUCH CLAMP BOWEL 50 MM JAW 5MM, 32CM, RATCHETED, REF SP90-7045; endoscopic surgical device
Carefusion 2200
Specific lots of the product have the potential to exhibit cracking or breaking of the rotation knob due to the incorrect use of an adhesive in the manufacturing process.
Snowden-Pencer MIS DIAMOND-TOUCH CLAMP DEBAKEY 35 MM JAW 5MM, 32CM, RATCHETED, REF SP90-7044; endoscopic surgical device
Carefusion 2200
Specific lots of the product have the potential to exhibit cracking or breaking of the rotation knob due to the incorrect use of an adhesive in the manufacturing process.
Snowden-Pencer MIS DIAMOND-TOUCH GRASPERS CROCODILE GRASPER, DOUBLE ACTI 5MM, 45CM, RATCHETED HANDLE, REF SP90-7266; endoscopic surgical device
Carefusion 2200
Specific lots of the product have the potential to exhibit cracking or breaking of the rotation knob due to the incorrect use of an adhesive in the manufacturing process.