Infinity M300 telemetry monitoring device Software versions VG2.3.1 and lower, Model Numbers MS25755, MS26076, MS26031, MS18623, MS22862, MS25301

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    All systems using Software version VG2.3.1 and lower
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Draegar Medical Systems, Inc.
Reason for Recall:
The devices have potential cybersecurity vulnerabilities, which can include Distributed Denial of Service (DDoS), Spoofing, and Tampering.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Infinity M300 telemetry monitoring device Software versions VG2.3.1 and lower, Model Numbers MS25755, MS26076, MS26031, MS18623, MS22862, MS25301

Product Codes/Lot Numbers:

All systems using Software version VG2.3.1 and lower

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0372-2020

Related Recalls

The Infinity M540 patient monitor may randomly reboot due to an error to correctly transmit and read the header data of files in the memory of the device. Under this situation, the device will try to reboot to mitigate the error. The device will be available again for use within 30 seconds. If this error continues and the M540 reboots three times in a time span of 10 minutes, it will enter a fail-state. A fail-state will annunciate itself with a continuous sound to alert the user. The M540 will reset to factory default and the user can manually configure and readmit the patient to continue patient monitoring.

Dec 23, 2020 Diagnostic Equipment Nationwide View Details →