FULL VISION INC. TRACKMASTER, TREADMILL GE T2100-ST2 220V, Model #317-07927GE CHINA

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    UDI/DI 00860176000675, Serial Numbers: GEDC-6765, GEDC-6766, GEDC-6767, GEDC-6768, GEDC-6849, GEDC-6850, GEDC-6851, GEDC-6852
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Full Vision Inc
Reason for Recall:
This has been identified a rare occurrence, under a specific scenario, where the circuit is found in a latch-up condition on the PCB that controls the safety function of the Magnetic Safety Tether, that when the magnet is pulled loose as during a fall, the treadmill may fail to stop.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

FULL VISION INC. TRACKMASTER, TREADMILL GE T2100-ST2 220V, Model #317-07927GE CHINA

Product Codes/Lot Numbers:

UDI/DI 00860176000675, Serial Numbers: GEDC-6765, GEDC-6766, GEDC-6767, GEDC-6768, GEDC-6849, GEDC-6850, GEDC-6851, GEDC-6852

Distribution:

Distributed in: GA, NY

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0370-2023