Trinity Biotech HbA1c (GHb) Contorls Kit, 500uL (Levels I & II), REF 01-04-0020, Part number 05-01-0040B, A1c analysis
Class I - DangerousWhat Should You Do?
- Check if you have this product: UDI/DI 0539151674808, Lot 12161
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Primus Corporation dba Trinity Biotech
- Reason for Recall:
- The Control Level I material value was showing results exceeding the upper limit of the acceptable range after a period of three (3) months. Additionally, per the firm, the unacceptable Control Level 1 value results in the Analyzer generating an error code which stops further testing and nullifies associated patient samples, resulting in a delay in patient results and a potential delay in therapeutic modifications.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Trinity Biotech HbA1c (GHb) Contorls Kit, 500uL (Levels I & II), REF 01-04-0020, Part number 05-01-0040B, A1c analysis
Product Codes/Lot Numbers:
UDI/DI 0539151674808, Lot 12161
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0350-2023
Related Recalls
A. MENARINI diagnostics Hb 9210 PREMIER, LEVEL1, LEVEL II, Hemoglobin A1c Controls for hemoglobin A1c Assay, REF 45052, part number 05-01-0926, A1c analysis
Primus Corporation
The Control Level I material value was showing results exceeding the upper limit of the acceptable range after a period of three (3) months. The unacceptable Control Level 1 value results in the Analyzer generating an error code which stops further testing and nullifies associated patient samples, resulting in a delay in patient results and a potential delay in therapeutic modifications.