The LivaNova VNS Therapy System, used for Vagus Nerve Stimulation (VNS), consists of an implantable VNS Therapy generator, lead, and external programming system used to change stimulation settings. The generator is an implantable, multi-programmable pulse generator that delivers electrical signals to the vagus nerve through the lead. The generator is housed in a hermetically sealed titanium case and is powered by a single battery. Electrical signals are transmitted from the generator to the vagus nerve by the lead. The lead and the generator make up the implantable portion of the VNS Therapy System. The VNS Therapy Programming System includes a computer pre-installed with VNS Therapy programming software and a programming wand. The physician uses the programming system to read and change generator settings. The Model 1000 Generator was commercially approved for distribution in the U.S. on October 3, 2017 via PMA P970003/S210.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Serial Numbers Less Than 100000
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- LivaNova USA Inc
- Reason for Recall:
- Lead impedance values reported by the affected VNS generator will be higher compared to those reported by previous models. This is due to a change in the timing of when affected VNS generator takes the lead impedance measurement during diagnostic testing. As a result, normal impedance ranges for the affected VNS generator have shifted relative to the existing thresholds of 600 - 5300 Ohms defined in labeling and as present in the programming software.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
The LivaNova VNS Therapy System, used for Vagus Nerve Stimulation (VNS), consists of an implantable VNS Therapy generator, lead, and external programming system used to change stimulation settings. The generator is an implantable, multi-programmable pulse generator that delivers electrical signals to the vagus nerve through the lead. The generator is housed in a hermetically sealed titanium case and is powered by a single battery. Electrical signals are transmitted from the generator to the vagus nerve by the lead. The lead and the generator make up the implantable portion of the VNS Therapy System. The VNS Therapy Programming System includes a computer pre-installed with VNS Therapy programming software and a programming wand. The physician uses the programming system to read and change generator settings. The Model 1000 Generator was commercially approved for distribution in the U.S. on October 3, 2017 via PMA P970003/S210.
Product Codes/Lot Numbers:
Serial Numbers Less Than 100000
Distribution:
Nationwide distribution
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0346-2020
Related Recalls
Vagus nerve stimulator generators may stop delivering stimulation due to an internal, mechanically activated component that may become stuck in a closed position, which may result in patients returning to baseline seizure frequency or depressive symptoms and in the longer term, patients may be required to undergo generator replacement surgery to replace their device.
Vagus nerve stimulator generators may stop delivering stimulation due to an internal, mechanically activated component that may become stuck in a closed position, which may result in patients returning to baseline seizure frequency or depressive symptoms and in the longer term, patients may be required to undergo generator replacement surgery to replace their device.
Their is a potential that implanted pulse generators may stop delivering therapy due to an internal component issue.