CAIRE Liberator 20, MODEL 13256195, Liquid Oxygen System Unit
Class I - DangerousWhat Should You Do?
- Check if you have this product: UDI/DI M766133374030M766132561950, Serial Numbers: CBB3022320073, CBB3022320077, CBB3022320074
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Caire, Inc.
- Reason for Recall:
- An audit discovered some inconsistencies in weld penetration on the longitudinal weld seam on the inner bottle of the device on some of the units.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
CAIRE Liberator 20, MODEL 13256195, Liquid Oxygen System Unit
Product Codes/Lot Numbers:
UDI/DI M766133374030M766132561950, Serial Numbers: CBB3022320073, CBB3022320077, CBB3022320074
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0339-2023
Related Recalls
Class I - Dangerous
Potential failure of affected units to recharge the battery of the oxygen concentrators.
Class I - Dangerous
Potential failure of affected units to recharge the battery of the oxygen concentrators.
Class I - Dangerous
Potential failure of affected units to recharge the battery of the oxygen concentrators.