ICU Medical 30" (76 cm) Appx 3.6 ml, 20 Drop Admin Set w/Integrated ChemoLock Port Drip Chamber, ChemoLock w/Red Cap, Bag Hanger REF: CL3511

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    UDI:10887709057026/Lot: 14400616
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
ICU Medical, Inc.
Reason for Recall:
Port weld of drug transfer device may separate or break during use and potentially result in a leak.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

ICU Medical 30" (76 cm) Appx 3.6 ml, 20 Drop Admin Set w/Integrated ChemoLock Port Drip Chamber, ChemoLock w/Red Cap, Bag Hanger REF: CL3511

Product Codes/Lot Numbers:

UDI:10887709057026/Lot: 14400616

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0333-2026

Related Recalls

Class I - Dangerous

Under certain conditions, the pump may not display a soft limit alert, and will not notify the clinician of a soft limit violation before confirming and starting delivery. This may potentially lead to under- or over-delivery.

Apr 29, 2026 Infusion Pumps Nationwide View Details →

Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is delivered from the primary line container at the piggyback therapy rate after the piggyback therapy is complete. In certain cases, as described below, the clinician may receive an Upper Hard Limit (UHL) or Lower Hard Limit (LHL) Violation message when programming a piggyback flush, which prevents the flush from being programmed.

Dec 15, 2025 Infusion Pumps Nationwide View Details →

Needle free access device intended as accessory to vascular access used in hemodialysis or accessory to intravascular administration sets has a silicone seal which may bulge at the top, separate from the body, tear; may result in fluid leaks, occluded fluid flow, inability to inject/withdraw blood product, delayed/interrupted therapy, exposure to biological contaminants, air infused into the body

Dec 19, 2025 Other Medical Devices Nationwide View Details →