EMPOWR Porous Knee System TIBIAL PUNCH, SMALL, REF 801-05-202, NON STERILE, Qty 1 For use in orthopedic surgery.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Lot 252564
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Encore Medical, Lp
Reason for Recall:
The Locking Punch Guide was not sufficiently guiding the large press fit punch, causing the implant to be placed 2-3mm more posterior than intended.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

EMPOWR Porous Knee System TIBIAL PUNCH, SMALL, REF 801-05-202, NON STERILE, Qty 1 For use in orthopedic surgery.

Product Codes/Lot Numbers:

Lot 252564

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0327-2018

Related Recalls

942-01-40G acetabular system package was found to contain 942-01-36H acetabular system; and several other acetabular system and knee insert devices may also have a packaging discrepancy, which may cause surgical delay to exchange devices or revision surgery.

Jan 5, 2026 Implants & Prosthetics View Details β†’