DJO EMPOWR KNEE LOCKING TIBIAL PUNCH GUIDE, REF 801-05-042, NON STERILE, Qty 1 For use in orthopedic surgery.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lot 252567
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
β οΈ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Encore Medical, Lp
- Reason for Recall:
- The Locking Punch Guide was not sufficiently guiding the large press fit punch, causing the implant to be placed 2-3mm more posterior than intended.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
DJO EMPOWR KNEE LOCKING TIBIAL PUNCH GUIDE, REF 801-05-042, NON STERILE, Qty 1 For use in orthopedic surgery.
Product Codes/Lot Numbers:
Lot 252567
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0326-2018
Related Recalls
Knee and Humeral socket implants contain incorrect labeling.
Knee and Humeral socket implants contain incorrect labeling.
EMPOWR 3D Knee Tibial Insert 5R 14MM, VE, REF: 342-14-705
Encore Medical