IRON assay, Reference Numbers 6K95-41 and 6K95-30
Class I - DangerousWhat Should You Do?
- Check if you have this product: Reference Number 6K95-41: UDI-DI: 00380740010928; Lot Number: 30610Y600; Reference Number 6K95-30: UDI-DI: 00380740010911; Lot Numbers: 21058Y600, 21101Y600, 30918Y600
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Sentinel CH SpA
- Reason for Recall:
- A high recovery of Iron assay has been observed at some customer sites. Internal studies confirmed the positive shift of Quality Control and patient samples results, up to 15% for Iron Products (Alinity c Iron Reagent REF. 08P3920) to be used on Abbott Alinity Instrument and to 30% for Iron Products (IRON REF. 6K95-41 and IRON REF. 6K95- 30) to be used on Abbott Architect Instrument. When using the affected lots, there is a potential for delay in reporting patient results due to QC failure, also a potential for falsely elevated results.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
IRON assay, Reference Numbers 6K95-41 and 6K95-30
Product Codes/Lot Numbers:
Reference Number 6K95-41: UDI-DI: 00380740010928; Lot Number: 30610Y600; Reference Number 6K95-30: UDI-DI: 00380740010911; Lot Numbers: 21058Y600, 21101Y600, 30918Y600
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0325-2024
Related Recalls
Alinity c Iron Reagent, Reference Number 08P3920
Sentinel CH SpA
A high recovery of Iron assay has been observed at some customer sites. Internal studies confirmed the positive shift of Quality Control and patient samples results, up to 15% for Iron Products (Alinity c Iron Reagent REF. 08P3920) to be used on Abbott Alinity Instrument and to 30% for Iron Products (IRON REF. 6K95-41 and IRON REF. 6K95- 30) to be used on Abbott Architect Instrument. When using the affected lots, there is a potential for delay in reporting patient results due to QC failure, also a potential for falsely elevated results.
Calibrator vials were incorrectly labeled. Use of incorrectly labeled vials will cause the instrument to generate a warning that calibrators are incorrectly loaded and calibration will not be completed successfully.
ACE Control Set - IVD Control set for the quantitative determination of Angiotensin Converting Enzyme (ACE) REF 1667001
Sentinel CH SpA
Claim for stability after reconstitution from lyophilized to liquid form does not meet the requirements. New IFU instructions for handling the product provided to end users.