HURRICANE RX DILATION BALLOON 6MM 4CM- -for endoscopic dilatation of strictures of the biliary tree and the Sphincter of Oddi. UPN: M00545920
Class I - DangerousWhat Should You Do?
- Check if you have this product: GTIN: 08714729283836 Batches: 25576770, 25576881, 25576882, 25576883, 25576884, 25576885, 25608762, 25623386, 25623387, 25623388, 25623389, 25623390, 25623391, 25623392, 25623393, 25623394, 25623395, 25623396, 25623397, 25623398, 25649814, 25649817, 25649820, 25649821, 25649825, 25649826, 25649830, 25649831, 25678823, 25678827, 25678832, 25678837, 25680097, 25704217, 25704218, 25704219, 25704220, 25704221, 25704222, 25704223, 25739152, 25739153, 25739154, 25739155, 25739157, 25739158, 25771773, 25771774, 25771775, 25771776, 25771777, 25771778, 25836083, 25836084, 25836085, 25836086, 25836087, 25836088, 25836089, 25836090, 25836091, 25836092, 25878557, 25878558, 25878559, 25878980, 25878981, 25878982, 25878983, 25878984, 25878985, 25940124, 25940125, 25942723, 25947907, 25947909, 25947911, 25947913, 25947915, 25975775, 25975776, 25975777, 25975778, 25995786, 25996198, 26000471, 26000476, 26000477, 26000478, 26000479, 26000580, 26000700, 26010393, 26010394, 26010395, 26010396, 26058288, 26058291, 26058293, 26058294, 26058295, 26058296, 26058297, 26087734, 26087735, 26087736, 26087737, 26087738, 26087739, 26087740, 26087741, 26123187, 26123188, 26123189, 26123190, 26123191, 26123247, 26123248, 26123249, 26175470, 26175471, 26175472, 26175473, 26207983, 26207984, 26207985, 26207986, 26207987, 26207988, 26207989, 26207990, 26207991, 26238132, 26238133, 26238134, 26240368, 26240370, 26240371, 26240373, 26247248, 26247249, 26247250, 26247251, 26247252, 26247253, 26247254, 26247255, 26247256, 26247257, 26283632, 26283633, 26283634, 26283635, 26283636, 26283637, 26283638, 26283639, 26284140, 26284141, 26284142, 26284143, 26324749, 26324751, 26324752, 26324754, 26350404, 26350405, 26350406, 26350407, 26356444, 26366753, 26366754, 26366755, 26366756, 26366757, 26406522, 26406523, 26406524, 26406525, 26406526, 26406527, 26406528, 26406529, 26406530, 26406531, 26426921, 26426922, 26426923, 26428970, 26429811, 26429812, 26435537, 26435538, 26435539, 26435580, 26452047, 26456916, 26456917, 26462399, 26473523, 26473769, 26473770, 26473771, 26473772, 26473773, 26473774, 26495372, 26495950, 26500637, 26500638, 26512452, 26512453, 26512454, 26512455, 26512456, 26512457, 26512458, 26512459, 26512480, 26512481, 26512483, 26512485, 26549972, 26552065, 26556788, 26556789, 26559542, 26559543, 26559544, 26559548, 26581407, 26581408, 26582103, 26582104, 26582105, 26582106, 26582107, 26588072, 26588073, 26588074, 26588075, 26588076, 26628443, 26628444, 26638830, 26638831, 26651713, 26651714, 26651715, 26651716, 26665561, 26665567, 26665571, 26665572, 26665573, 26685875, 26685876, 26685877, 26685878, 26685879, 26699192, 26699193, 26699194, 26699195, 26699196, 26699197, 26728782, 26728784, 26728785, 26759345, 26760530, 26778894, 26778895, 26778896, 26778897, 26778898, 26779080, 26779081, 26779082, 26779083, 26793782, 26793783, 26793784, 26808197, 26808198, 26808199, 26808300, 26808301, 26808302, 26808303, 26848720, 26848721, 26848722, 26848723, 26848724, 26848725, 26871168, 26871169, 26871170, 26871171, 26871172, 26871173, 26871174, 26910997, 26910998, 26910999, 26911840, 26911841, 26911842, 26911845, 26919040, 26919041, 26919042, 26960166, 26960167, 26960168, 26960169, 26960170, 26960171, 26960172, 26960173, 26960174, 26960175, 27009668, 27027289, 27027290, 27027291, 27027292, 27062054, 27062055, 27062056, 27062057, 27065830, 27065831, 27065832, 27065833, 27065834, 27065835, 27100017, 27100018, 27100019, 27100240, 27100241, 27100242, 27110096, 27110097, 27121582, 27121583, 27121584, 27121585, 27121620, 27121621, 27121625, 27148636, 27148637, 27148638, 27162326, 27162327, 27162328, 27162329, 27182896, 27182897, 27182898, 27182899, 27203768, 27203769, 27203770, 27203771, 27203772, 27203773, 27255217, 27255219, 27255220, 27255221, 27255222, 27255223, 27308422, 27308423, 27308424, 27308425, 27335091, 27335092, 27335093, 27335094, 27340629, 27340630, 27340632, 27340633, 27340634, 27340638, 27347592, 27347593, 27384705, 27384706, 27384713, 27384715. Exp. 05/10/2022 to 05/26/2023
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Boston Scientific Corporation
- Reason for Recall:
- Balloon pinhole results in losss pressure or fails to gain or maintain pressure, causing prolonged procedure
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
HURRICANE RX DILATION BALLOON 6MM 4CM- -for endoscopic dilatation of strictures of the biliary tree and the Sphincter of Oddi. UPN: M00545920
Product Codes/Lot Numbers:
GTIN: 08714729283836 Batches: 25576770, 25576881, 25576882, 25576883, 25576884, 25576885, 25608762, 25623386, 25623387, 25623388, 25623389, 25623390, 25623391, 25623392, 25623393, 25623394, 25623395, 25623396, 25623397, 25623398, 25649814, 25649817, 25649820, 25649821, 25649825, 25649826, 25649830, 25649831, 25678823, 25678827, 25678832, 25678837, 25680097, 25704217, 25704218, 25704219, 25704220, 25704221, 25704222, 25704223, 25739152, 25739153, 25739154, 25739155, 25739157, 25739158, 25771773, 25771774, 25771775, 25771776, 25771777, 25771778, 25836083, 25836084, 25836085, 25836086, 25836087, 25836088, 25836089, 25836090, 25836091, 25836092, 25878557, 25878558, 25878559, 25878980, 25878981, 25878982, 25878983, 25878984, 25878985, 25940124, 25940125, 25942723, 25947907, 25947909, 25947911, 25947913, 25947915, 25975775, 25975776, 25975777, 25975778, 25995786, 25996198, 26000471, 26000476, 26000477, 26000478, 26000479, 26000580, 26000700, 26010393, 26010394, 26010395, 26010396, 26058288, 26058291, 26058293, 26058294, 26058295, 26058296, 26058297, 26087734, 26087735, 26087736, 26087737, 26087738, 26087739, 26087740, 26087741, 26123187, 26123188, 26123189, 26123190, 26123191, 26123247, 26123248, 26123249, 26175470, 26175471, 26175472, 26175473, 26207983, 26207984, 26207985, 26207986, 26207987, 26207988, 26207989, 26207990, 26207991, 26238132, 26238133, 26238134, 26240368, 26240370, 26240371, 26240373, 26247248, 26247249, 26247250, 26247251, 26247252, 26247253, 26247254, 26247255, 26247256, 26247257, 26283632, 26283633, 26283634, 26283635, 26283636, 26283637, 26283638, 26283639, 26284140, 26284141, 26284142, 26284143, 26324749, 26324751, 26324752, 26324754, 26350404, 26350405, 26350406, 26350407, 26356444, 26366753, 26366754, 26366755, 26366756, 26366757, 26406522, 26406523, 26406524, 26406525, 26406526, 26406527, 26406528, 26406529, 26406530, 26406531, 26426921, 26426922, 26426923, 26428970, 26429811, 26429812, 26435537, 26435538, 26435539, 26435580, 26452047, 26456916, 26456917, 26462399, 26473523, 26473769, 26473770, 26473771, 26473772, 26473773, 26473774, 26495372, 26495950, 26500637, 26500638, 26512452, 26512453, 26512454, 26512455, 26512456, 26512457, 26512458, 26512459, 26512480, 26512481, 26512483, 26512485, 26549972, 26552065, 26556788, 26556789, 26559542, 26559543, 26559544, 26559548, 26581407, 26581408, 26582103, 26582104, 26582105, 26582106, 26582107, 26588072, 26588073, 26588074, 26588075, 26588076, 26628443, 26628444, 26638830, 26638831, 26651713, 26651714, 26651715, 26651716, 26665561, 26665567, 26665571, 26665572, 26665573, 26685875, 26685876, 26685877, 26685878, 26685879, 26699192, 26699193, 26699194, 26699195, 26699196, 26699197, 26728782, 26728784, 26728785, 26759345, 26760530, 26778894, 26778895, 26778896, 26778897, 26778898, 26779080, 26779081, 26779082, 26779083, 26793782, 26793783, 26793784, 26808197, 26808198, 26808199, 26808300, 26808301, 26808302, 26808303, 26848720, 26848721, 26848722, 26848723, 26848724, 26848725, 26871168, 26871169, 26871170, 26871171, 26871172, 26871173, 26871174, 26910997, 26910998, 26910999, 26911840, 26911841, 26911842, 26911845, 26919040, 26919041, 26919042, 26960166, 26960167, 26960168, 26960169, 26960170, 26960171, 26960172, 26960173, 26960174, 26960175, 27009668, 27027289, 27027290, 27027291, 27027292, 27062054, 27062055, 27062056, 27062057, 27065830, 27065831, 27065832, 27065833, 27065834, 27065835, 27100017, 27100018, 27100019, 27100240, 27100241, 27100242, 27110096, 27110097, 27121582, 27121583, 27121584, 27121585, 27121620, 27121621, 27121625, 27148636, 27148637, 27148638, 27162326, 27162327, 27162328, 27162329, 27182896, 27182897, 27182898, 27182899, 27203768, 27203769, 27203770, 27203771, 27203772, 27203773, 27255217, 27255219, 27255220, 27255221, 27255222, 27255223, 27308422, 27308423, 27308424, 27308425, 27335091, 27335092, 27335093, 27335094, 27340629, 27340630, 27340632, 27340633, 27340634, 27340638, 27347592, 27347593, 27384705, 27384706, 27384713, 27384715. Exp. 05/10/2022 to 05/26/2023
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0323-2022
Related Recalls
LUX-DX II, Arrythmia detector and alarm, Model M302, with SERVER SW LATITUDE DRAGON US, Model 6460, Version 7.5
Boston Scientific
For some patients upgraded from a LUX-Dx M301 device to a new Model M302 or M312, the new ICM device is not collecting PVC Burden data (both M302/M312) or monitoring for Bradyarrhythmia and Pause (M312 only). In addition, for these devices that are not being monitored, the LATITUDE Clarity programming screens have discrepant information, indicating that the monitoring is enabled for these features in one location and not enabled in another location.
For some patients upgraded from a LUX-Dx M301 device to a new Model M302 or M312, the new ICM device is not collecting PVC Burden data (both M302/M312) or monitoring for Bradyarrhythmia and Pause (M312 only). In addition, for these devices that are not being monitored, the LATITUDE Clarity programming screens have discrepant information, indicating that the monitoring is enabled for these features in one location and not enabled in another location.
Updated software, Brady software maintenance release 6 (SMR6), is now available for the ACCOLADE" family of pacemakers and cardiac resynchronization therapy pacemakers (CRT-Ps)1 and the advisory population is expanding to include all CRT-P and dual-chamber extended life (DR-EL) devices. Dual chamber modes are specifically indicated for treatment of the following: " Conduction disorders that require restoration of AV synchrony, including varying degrees of AV block " VVI intolerance (i.e., pacemaker syndrome) in the presence of persistent sinus rhythm " Low cardiac output or congestive heart failure secondary to bradycardia