Company:
Zimmer Biomet, Inc.
Reason for Recall:
Complaints have been reported that the handpiece is inoperable. The investigation found that the complaints were related to the lack of preventative maintenance contrary to what is recommended in the instructions for use included with the handpiece. An inoperable device may lead to a delay in surgery
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Zimmer¿ Gender Solutions" Patellofemoral Joint Prosthesis Milling Handpiece

Product Codes/Lot Numbers:

Item Number 00-5927-040-00 Serial Number All

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0317-2017

Related Recalls

Zimmer Biomet is conducting a medical device correction to update the Instructions for Use (IFU) and the Surgical Techniques for the Polaris Spinal System Translation Screw and Cypher MIS Screw System. Users should be aware that changes were made to the contraindications section indicating that the Translation Screws should not be used with a direct current stimulation, as a high probability of construct failure exists when using Translation Screws containing CoCr with direct current simulation system.

Nov 17, 2020 Other Medical Devices Nationwide View Details →