Elscint SPX4; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    all serial numbers in distribution
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING
Reason for Recall:
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Elscint SPX4; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging

Product Codes/Lot Numbers:

all serial numbers in distribution

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0292-2026

Related Recalls

There is a potential intermittent issue on certain Omni Legend systems that can result in a streaking artifact in the PET clinical scan images. This streaking artifact is most easily identified in transaxial slices on the acquisition console (both corrected and non-corrected for attenuation).

Dec 24, 2025 Other Medical Devices Nationwide View Details →

SMV DSI, System, Tomography, Computed, Emission

GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING

Class I - Dangerous

GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service (herein referred to as Impacted Systems and listed in the Affected Product Details section of this letter) could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury. There have been no reports of detector falls or injuries as a result of this potential issue. The Impacted Systems affected by this notice are those that have not been manufactured for several years. GE HealthCare has previously communicated End of Guaranteed Service for these Impacted Systems.

Dec 10, 2025 Other Medical Devices View Details →
Class I - Dangerous

GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.

Sep 12, 2025 Diagnostic Equipment View Details →