PT Link Waterbath, Model#PT100. Lab Vision PT Module Plus provides an automated standarized alternative to labor intensive and time consuming dewaxing and epitote retrieval procedures.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Catalog #: PT-Module
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Dako North America Inc.
Reason for Recall:
Dako is recalling the PT Link Waterbath because of a possible electrical hazard. Specifically, if the heater pad under the reagent tank becomes wet, it may present an electrical shock hazard when the lid is open.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

PT Link Waterbath, Model#PT100. Lab Vision PT Module Plus provides an automated standarized alternative to labor intensive and time consuming dewaxing and epitote retrieval procedures.

Product Codes/Lot Numbers:

Catalog #: PT-Module

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0287-2013

Related Recalls

An incorrect HER2 probe concentration was used when producing the probe component (Vial 3) in this kit lot. This has resulted in the HER2 probe concentration being too low.

Jun 10, 2016 Diagnostic Equipment Nationwide View Details →

There is a defect in one lot of a buffer, non-specific background staining that has, in some instances, interfered with interpretation of the staining results.

Sep 4, 2015 Diagnostic Equipment Nationwide View Details →