epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System, SMN 10736515
Class I - DangerousWhat Should You Do?
- Check if you have this product: UDI-DI: 00809708121860; Lot Numbers: 08-24092-60 08-24093-60 08-24094-60 08-24095-60 08-24096-60
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Siemens Healthcare Diagnostics Inc
- Reason for Recall:
- Siemens Healthcare Diagnostics Inc. has confirmed that a negative sodium bias may exist in one lot of epoc BGEM BUN Test Cards approximately 2 months prior to expiration. The observed maximum negative bias from internal testing is -10 mmol/L. No positive bias has been observed. Quality Control (QC) fluid may or may not detect the issue depending on the bias magnitude and the frequency of QC testing. One customer complaint has been associated with this issue. No adverse events have been reported. All affected lots have recently expired before the issuance of the recall.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System, SMN 10736515
Product Codes/Lot Numbers:
UDI-DI: 00809708121860; Lot Numbers: 08-24092-60 08-24093-60 08-24094-60 08-24095-60 08-24096-60
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0286-2025
Related Recalls
Siemens Dimension Tacrolimus Flex Reagent Cartridge (TAC), REF DF207, SMN 10700795; enzyme immunoassay
Siemens Healthcare Diagnostics
A potential positive bias in quality control (QC) and patient sample results in three lots of the Dimension Tacrolimus (TAC) Flex reagent cartridge.
Dimension Creatinine Flex reagent cartridge. Material Number: 10872079.
Siemens Healthcare Diagnostics
Siemens Healthcare Diagnostics is recalling certain lots of Dimension Creatinine Flex reagent cartridge (Dimension CRE2) because a potential imprecision in the Dimension CRE2 quality control (QC) and patient sample results using lot numbers GA6307 and BA7005 on the Dimension System. This imprecision may lead to erroneously decreased or increased creatinine patient results
Atellica CH Enzymatic Hemoglobin A1c (A1c_E); Siemens Material Number (SMN): 11097536;
Siemens Healthcare Diagnostics
A potential for depressed results for Atellica CH A1c_E when processing Atellica CH RCRP on the same analyzer. Customers processing A1c_E and RCRP on separate analyzers are not impacted by the recall. The frequency and magnitude of the negative bias and imprecision increase over time and are not consistent with the assay Instructions For Use (IFU). When this issue is present, Quality Control (QC) results demonstrate a negative trend and increased imprecision. The clinical impact of a negative (A1c_E) bias may lead to a delay in diagnosis for diabetes or lead to less-than-optimal treatment for existing diabetes.