epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System, SMN 10736515
Class I - DangerousWhat Should You Do?
- Check if you have this product: UDI-DI: 00809708121860; Lot Numbers: 08-24092-60 08-24093-60 08-24094-60 08-24095-60 08-24096-60
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Siemens Healthcare Diagnostics Inc
- Reason for Recall:
- Siemens Healthcare Diagnostics Inc. has confirmed that a negative sodium bias may exist in one lot of epoc BGEM BUN Test Cards approximately 2 months prior to expiration. The observed maximum negative bias from internal testing is -10 mmol/L. No positive bias has been observed. Quality Control (QC) fluid may or may not detect the issue depending on the bias magnitude and the frequency of QC testing. One customer complaint has been associated with this issue. No adverse events have been reported. All affected lots have recently expired before the issuance of the recall.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System, SMN 10736515
Product Codes/Lot Numbers:
UDI-DI: 00809708121860; Lot Numbers: 08-24092-60 08-24093-60 08-24094-60 08-24095-60 08-24096-60
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0286-2025
Related Recalls
Food Panel 3 Specific Allergen; Siemens Material Number (SMN): 10385891; Catalog Number: FP3L4;
Siemens Healthcare Diagnostics
The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.
3gAllergy Specific IgE Universal Kit; Kit Component: 3gAllergy Specific IgE Adjustor Antibody (L2UNS1), 3gAllergy Specific IgE Control Antibody (L2UNS2); Siemens Material Number (SMN): 10380875; Catalog Number: L2KUN6;
Siemens Healthcare Diagnostics
The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.
IMMULITE 2000 Oak Mix Specific Allergen; Siemens Material Number (SMN): 10386105; Catalog Number: L2KUN6;
Siemens Healthcare Diagnostics
The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.