Philips Sterilizable Defibrillator Paddles, Switched Internal Paddles, Model Numbers M4741A, M4742A, M4743A, M4744A - Product Usage: used in pairs to transmit a controlled electrical shock from an external defibrillator directly to the exposed heart muscle of a post-pubescent patient in order to intentionally stop start the heartbeat during cardiopulmonary surgery.
Class I - DangerousWhat Should You Do?
- Check if you have this product: All units manufactured and distributed January 2015 to August 2020.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Philips North America, LLC
- Reason for Recall:
- The periodic Paddle Checks recommended in the Instructions for Use for Sterilizable Defibrillator Paddles may not detect one failure mode for the Switched Internal Defibrillator Paddles. Philips has created an addendum to the IFU, that includes a test for the switched paddles only, which will detect this failure mode.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Philips Sterilizable Defibrillator Paddles, Switched Internal Paddles, Model Numbers M4741A, M4742A, M4743A, M4744A - Product Usage: used in pairs to transmit a controlled electrical shock from an external defibrillator directly to the exposed heart muscle of a post-pubescent patient in order to intentionally stop start the heartbeat during cardiopulmonary surgery.
Product Codes/Lot Numbers:
All units manufactured and distributed January 2015 to August 2020.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0284-2021
Related Recalls
Philips SmartPath to dStream for 3.0T with MR Elastography (MRE). 1. Model Number (REF): 782145.
Philips North America
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
Philips SmartPath to Ingenia Elition X with MR Elastography (MRE). 1. Model Number (REF): 782118. 2. Model Number (REF): 782144. 3. Model Number (REF): 782163.
Philips North America
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.