O-arm 1000 2nd Edition Imaging System, OARM ASSY Bl70000027GERR SYS PRODUCT REFURB, Model Number Bl70000027GERR Product Usage: The O-arm Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The 0-arm Imaging System is compatible with certain Image Guided Surgery Systems.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    All Serial Numbers
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Medtronic Navigation, Inc.-Littleton
Reason for Recall:
Software anomalies were reported to affect the following system functionalities: startup and shutdown, system and network communication, motion control, image acquisition and output, dose reporting, and logs generation. Software version 3.1.7 will be installed on all 2nd edition imaging systems.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

O-arm 1000 2nd Edition Imaging System, OARM ASSY Bl70000027GERR SYS PRODUCT REFURB, Model Number Bl70000027GERR Product Usage: The O-arm Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The 0-arm Imaging System is compatible with certain Image Guided Surgery Systems.

Product Codes/Lot Numbers:

All Serial Numbers

Distribution:

Nationwide distribution

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0280-2019

Related Recalls

Potential for an electrical component of the O-arm" O2 Imaging System to intermittently shut off,failure can cause the potential for surgical delay, additional imaging, cancellation and rescheduling of surgery, or completion of surgery without the benefit of image guidance.

Apr 26, 2024 Diagnostic Equipment Nationwide View Details →

Ground cable installed incorrectly.

Jul 25, 2023 Diagnostic Equipment Nationwide View Details →
Class I - Dangerous

Due to a manufacturing issue affecting particular O-arm" O2 Imaging Systems, Medtronic has determined that the gantry tractor motor drive belt may loosen over an extended number of uses at a higher frequency in these systems. If the belt loosens, it may result in rotation of 3D images about the gantry isocenter. While the 3D image is anatomically accurate within the image itself and may be used to confirm therapy, its electronic registered location may be rotated relative to the actual physical position of the patient.

Nov 15, 2019 Diagnostic Equipment Nationwide View Details →