AVOXimeter 1000E, Product Number: AVOX1000E. The AVOXimeter 1000E is a battery-operated bedside whole blood oximeter.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    UDI-DI: 10711234530023. Serial Number: 6824
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Accriva Diagnostics, Inc.
Reason for Recall:
Values used to calibrate whole blood oximeter instruments were observed to be higher than normal, which may result in incorrect patient results where oxygen levels are higher than actual results, which may lead to patient care mismanagement requiring medical intervention.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

AVOXimeter 1000E, Product Number: AVOX1000E. The AVOXimeter 1000E is a battery-operated bedside whole blood oximeter.

Product Codes/Lot Numbers:

UDI-DI: 10711234530023. Serial Number: 6824

Distribution:

Distributed in: US, AZ

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0279-2025

Related Recalls

Class I - Dangerous

Platelet Reactivity Test distributed without appropriate US-FDA market clearance; if used healthcare personnel could use the baseline PRUTest measurement prior to initiating drug therapy, and then use the percent inhibition formula in the notice to manually calculate the percentage of platelet aggregation inhibition for a given patient receiving drug treatment with P2Y12 receptor inhibitor.

Dec 2, 2024 Diagnostic Equipment Nationwide View Details →