Prismaflex Control Unit, software versions below 7.21
Class I - DangerousWhat Should You Do?
- Check if you have this product: Product Code: 115269 (GTIN: 07332414117702), 955542 GTIN: 07332414124236); All Lot numbers
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Baxter Healthcare Corporation
- Reason for Recall:
- Communication error alarms may result in interruption of therapy, delay in therapy, or blood loss due to non-restitution of blood in the extracorporeal circuit.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Prismaflex Control Unit, software versions below 7.21
Product Codes/Lot Numbers:
Product Code: 115269 (GTIN: 07332414117702), 955542 GTIN: 07332414124236); All Lot numbers
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0279-2020
Related Recalls
Pumps were released without full testing being performed, including occlusion alarm testing.
Sterile Water for Irrigation, 250 mL, Single-Dose Plastic Bottle, Rx Only, Baxter Healthcare Corporation Deerfield, IL 60015 NDC: 0338-0004-02.
Baxter Healthcare
Presence of particulate matter: plastic particles from the bottle rim were observed floating into the solution
SIGMA Spectrum Infusion Pump, Model/Catalog Number: 35700BAX2
Baxter Healthcare
Certain pumps have potentially been released from service with defective grease applied to the cam and motor gears. Due to its low viscosity, the defective grease may lead to the device having insufficient lubrication on the cam, which could lead to premature wear of the mechanism assembly, resulting in over-infusion or a free-flow situation.